Philips IntelliVue TRx Instructions For Use Manual page 176

For the its4840a/its4850a intellivue telemetry system
Hide thumbs Also See for IntelliVue TRx:
Table of Contents

Advertisement

Alarm Behavior with Telemetry Overview
Alarm
Suspend/
Pause
Alarm Silence
10-4
Pairing with IntelliVue Patient Monitors
When alarms are suspended/paused, the messages and types of alarms affected
depend on where the suspend/pause was initiated.
If alarms are
these alarms are
suspended/
suspended/
paused
paused
from....
Information
both bedside and
Center
telemetry
measurements
IntelliVue
bedside
Patient Monitor
measurements only
When an active alarm is silenced, the types of alarms that are silenced depend on
the alarm source and where the silence was initiated.
and this message appears
Information Center: ALARMS
SUSPENDED or ALARMS
PAUSED
Bedside: ALARMS OFF in
Overview window and ALARMS
OFF or ALARMS PAUSED on the
monitor (depending on
configuration)
Information Center: BED
ALARMS SUSPENDED or BED
ALARMS PAUSED (depending on
configuration)
Bedside: ALARMS OFF or
ALARMS PAUSED