Telemetry Functions at the Information Center
Unit Settings - Telemetry Setup
Control
Function
Suppress
Enable/disable the SpO
SpO
Inops
algorithm to detect NBP running
2
with NBP
and suppress sending technical
alarms from the transceiver for
60 seconds.
Warning
Warning
If you measure SpO
that has an inflated NBP cuff,
a non-pulsatile SpO
alarm can occur. If the
monitor is configured to
suppress this alarm, there can
be a delay of up to 60 seconds
in indicating critical patient
status, such as sudden pulse
loss or hypoxia.
Pleth Wave
Enable/disable the transmission
of the Pleth wave and its
subsequent display to the
Information Center. For
Continuous mode only.
Pulse
Enable/disable the transmission
of the Pulse rate and its
subsequent display to the
Information Center. For
Continuous mode only.
9-12
Telemetry Functions at the Information Center & TeleMon
Settings
enable
2
disable
on a limb
2
technical
2
enable
disable
Note—
6-lead monitoring, the Pleth
wave will replace the Vb
wave in the Patient Window.
enable
disable
When enabled, during
Factory
Default
enable
disable
(Pleth wave
is not
displayed.)
disable
(Pulse rate is
not
displayed.)