System Compatibility With Cxdi-70C Wireless; Restrictions On Connections - Canon CXDI-80C Service Manual

Table of Contents

Advertisement

3.2 System compatibility with CXDI-70C Wireless

The CXDI-80C Wireless uses the same system components as the CXDI-70C Wireless, as shown
below.
Battery Pack
70C Detector Unit
80C Detector Unit

3.3 Restrictions on connections

The control software that can be connected to the product is Ver.1.40 and up. The number of units
which can be connected is restricted as below by the software requirements.
Registered units: 10, simultaneously connected units: 4
The hardware requirements do not impose restrictions, but communication performance may be
degraded by interference when wireless channels are shared.
No such restrictions apply to wired connections.
◆ Handling of ferrite cores
(1) [70C Wireless (2.4 GHz compatible)]
The system conforms to IEC 60601-1-2 when ferrite cores are attached to the detector cable and to the
LAN cable used to connect to the X-ray interface box.
It is therefore necessary to attach ferrite cores to these two cables in order to use the system.
(2) [70C Wireless (2.4 GHz/5 GHz compatible)]
The system conforms to IEC 60601-1-2 when no ferrite cores are attached to the sensor cable and to the
LAN cable used to connect to the X-ray interface box.
It is therefore not necessary to attach ferrite cores to these two cables in order to use the system.
Note: Caution on mixed usage of different detector models
If any of the detectors used is the 70C Wireless (2.4 GHz compatible) model, ferrite cores must be
attached to the two cables.
1
.
O
v
e
1
.
O
v
e
Sensor
Communication Unit
- 4 -
r
v
i
e
w
r
v
i
e
w
X-ray Generator
X-ray I/F Box
IF Cable (20m)
Access Point
(2.3m)
IR Data
USB
(< 5m)
Operation Unit
(1.8m)
HUB
LAN
LAN (20m)
LAN
(1.8m)

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents