3. Disassembly and re-assembly

3.1 Detector

In principle, the detector should be returned to Canon Inc. for repairs.
However, the connector pads of the battery holder may be replaced with commercially available
repair parts should they peel off or become worn.

3.1.1 Replacing the battery connector pads

If the battery is attached after the connector pads have peeled off, the connector pins could be
damaged or foreign matter could get into the gap, causing a short circuit. Follow the procedure
below to replace the connector pads should they peel off or become worn.
1)
Remove the battery pack from the detector.
Slide the two lock switches toward the outside, allowing the battery pack to lift up slightly.
2)
Peel the connector pads (together with the double-sided adhesive tape on the back of the
pads) off of the battery holder.
4. Repair Guide
- 2 -
Connector portion
of battery holder