Canon CXDI-80C Service Manual page 64

Table of Contents

Advertisement

3) Open the external connector cover of the detector, and connect the sensor cable to the external
connector.
* When connecting the detector cable, do not insert the connector the wrong way round.
* Since the external connector cover may become dislodged from the detector, take care not to
open the cover too far. Also, under no circumstances must the external connector cover be
removed from the detector.
* When the sensor cable is not going to be connected to the detector, the connector area must be
covered using the external connector cover.
* Care is required in handling the sensor cable because the sensor cable connector can be
broken if the cable is pulled from the detector without the lock released.
Detector
External connector
When disconnecting the detector cable:
Before disconnecting the sensor cable, push in the two lock releases to release the lock properly.
2. Installation procedures
- 31 -
External connector cover
Detector cable