Alarms

Parameter alarm levels

Arrhythmia alarm levels

5-14
To make a change in the telemetry unit default alarm level for a parameter, first access
the Telemetry Alarm Control Defaults tab as described in the Telemetry Alarm
Control Defaults section in this chapter. Then follow the procedure below.
1.
In Telemetry Alarm Control Defaults tab, use the mouse to click in the Level
field of the parameter for which the alarm level is to be changed. A down arrow
button appears in the field.
2.
Click on the down arrow button. A popup list of alarm level selections appears.
3.
Click on the desired alarm level to select it.
4.
Once you have set the desired level, click the Apply button for the changes to
take effect.
5.
If you are finished making changes to the Telemetry Alarm Control Defaults tab
sheet, click the OK button.
To make a change in the telemetry unit default alarm level for arrhythmia alarms, first
access the Telemetry Alarm Control Defaults tab as described in the Telemetry
Alarm Control Defaults section in this chapter. Then follow the procedure below.
1.
In the Telemetry Alarm Control Defaults tab, the Arrhythmia Alarm Levels for
which unit defaults can be set appears on the right side of the window:
2.
Click in the Level field of the arrhythmia alarm you wish to modify. A down
arrow button appears in the field.
NOTE
The Arrhythmia Alarm Levels for ASYSTOLE and VFIB/VTAC cannot be
changed. Therefore, the text in the Level field for these alarms always
appears dimmed.