Page 3
• VERSUCHEN SIE NICHT, DAS GERÄT ZU REPARIEREN, BEVOR DIESES WARNUNG KUNDENDIENST-HANDBUCH NICHT ZU RATE GEZOGEN UND VERSTANDEN (DE) WURDE. • WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN DURCH ELEKTRISCHE SCHLÄGE, MECHANISCHE ODER SONSTIGE GEFAHREN KOMMEN.
Page 4
LOGIQ P9/P7 5604324, R IRECTION EVISION ERVICE ANUAL ESTE MANUAL DE SERVICIO SÓLO EXISTE EN INGLÉS. • SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEHC SOLICITA UN IDIOMA QUE NO SEA EL INGLÉS, ES RESPONSABILIDAD DEL CLIENTE OFRECER UN SERVICIO DE TRADUCCIÓN.
Page 5
ΣΤΟΝ ΑΣΘΕΝΗ ΑΠΟ ΗΛΕΚΤΡΟΠΛΗΞΙΑ, ΜΗΧΑΝΙΚΟΥΣ Ή ΑΛΛΟΥΣ ΚΙΝ∆ΥΝΟΥΣ. EZEN KARBANTARTÁSI KÉZIKÖNYV KIZÁRÓLAG ANGOL NYELVEN ÉRHETŐ EL. • HA A VEVŐ SZOLGÁLTATÓJA ANGOLTÓL ELTÉRŐ NYELVRE TART IGÉNYT, AKKOR A VEVŐ FELELŐSSÉGE A FORDÍTÁS ELKÉSZÍTTETÉSE. • NE PRÓBÁLJA ELKEZDENI HASZNÁLNI A BERENDEZÉST, AMÍG A FIGYELMEZTETÉS KARBANTARTÁSI KÉZIKÖNYVBEN LEÍRTAKAT NEM ÉRTELMEZTÉK.
Page 6
LOGIQ P9/P7 5604324, R IRECTION EVISION ERVICE ANUAL ÞESSI ÞJÓNUSTUHANDBÓK ER EINGÖNGU FÁANLEG Á ENSKU. • EF ÞJÓNUSTUAÐILI VIÐSKIPTAMANNS ÞARFNAST ANNARS TUNGUMÁLS EN ENSKU, ER ÞAÐ Á ÁBYRGÐ VIÐSKIPTAMANNS AÐ ÚTVEGA ÞÝÐINGU. • REYNIÐ EKKI AÐ ÞJÓNUSTA TÆKIÐ NEMA EFTIR AÐ HAFA SKOÐAÐ OG VIÐVÖRUN...
Page 7
LOGIQ P9/P7 5604324, R IRECTION EVISION ERVICE ANUAL ŠĪ APKALPES ROKASGRĀMATA IR PIEEJAMA TIKAI ANGĻU VALODĀ. • JA KLIENTA APKALPES SNIEDZĒJAM NEPIECIEŠAMA INFORMĀCIJA CITĀ VALODĀ, NEVIS ANGĻU, KLIENTA PIENĀKUMS IR NODROŠINĀT TULKOŠANU. • NEVEICIET APRĪKOJUMA APKALPI BEZ APKALPES ROKASGRĀMATAS BRĪDINĀJUMS IZLASĪŠANAS UN SAPRAŠANAS.
Page 8
LOGIQ P9/P7 5604324, R IRECTION EVISION ERVICE ANUAL ACEST MANUAL DE SERVICE ESTE DISPONIBIL NUMAI ÎN LIMBA ENGLEZĂ. • DACĂ UN FURNIZOR DE SERVICII PENTRU CLIENŢI NECESITĂ O ALTĂ LIMBĂ DECÂT CEA ENGLEZĂ, ESTE DE DATORIA CLIENTULUI SĂ FURNIZEZE O TRADUCERE.
Page 9
LOGIQ P9/P7 5604324, R IRECTION EVISION ERVICE ANUAL TA SERVISNI PRIROČNIK JE NA VOLJO SAMO V ANGLEŠČINI. • ČE PONUDNIK SERVISNIH STORITEV ZA STRANKO POTREBUJE NAVODILA V DRUGEM JEZIKU, JE ZA PREVOD ODGOVORNA STRANKA SAMA. • NE POSKUŠAJTE SERVISIRATI OPREME, NE DA BI PREJ PREBRALI IN OPOZORILO RAZUMELI SERVISNI PRIROČNIK.
Page 10
LOGIQ P9/P7 5604324, R IRECTION EVISION ERVICE ANUAL BU SERVİS KILAVUZU YALNIZCA İNGİLİZCE OLARAK SAĞLANMIŞTIR. • EĞER MÜŞTERİ TEKNİSYENİ KILAVUZUN İNGİLİZCE DIŞINDAKİ BİR DİLDE OLMASINI İSTERSE, KILAVUZU TERCÜME ETTİRMEK MÜŞTERİNİN SORUMLULUĞUNDADIR. DİKKAT • SERVİS KILAVUZUNU OKUYUP ANLAMADAN EKİPMANLARA MÜDAHALE (TR) ETMEYİNİZ.
Page 12
GE Healthcare personnel. In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical work on these products will comply with the requirements of the applicable electrical codes.
Page 13
The contents of this publication may not be copied or duplicated in any form, in whole or in part, without prior written permission of GE Healthcare. GE Healthcare may revise this publication from time to time without written notice. TRADEMARKS All products and their name brands are trademarks of their respective holders.
Page 14
LOGIQ P9/P7 5604324, R IRECTION EVISION ERVICE ANUAL Revision History Revision Date Reason for change October 2014 Initial release December 2014 Enhancement feedback May 2015 Enhancement feedback March 2016 Initial release for R2 August 2016 Correction November 2016 Initial release for R2.5...
LOGIQ P9/P7 5604324, R IRECTION EVISION ERVICE ANUAL Table of Contents CHAPTER 1 Introduction Overview ............1 - 1 Purpose of Chapter 1 .
Page 16
Electrical Requirements ........2 - 2 LOGIQ P9/P7 Power Requirements ......2 - 3 Inrush Current .
Page 17
LOGIQ P9/P7 5604324, R IRECTION EVISION ERVICE ANUAL CHAPTER 3 Setup Instructions Overview............3 - 1 The Purpose of Chapter 3 .
Page 18
Dataflow Examples ........3 - 55 Stand-alone LOGIQ P9/P7 ....... . 3 - 56 LOGIQ? + PC within a “Sneaker Net”...
Page 19
ANUAL You will need one network cable to connect the LOGIQ P9/P7 to a wall outlet on the hospital’s network..... 3 - 57 Connection using Wireless Option .
Overview 1-1-1 Purpose of Chapter 1 This chapter describes important issues related to safely servicing the LOGIQ P9/P7 ultrasound system. The service provider must read and understand all the information presented in this manual before installing or servicing a unit.
5641296 LOGIQ P7 with High Cabinet 5641037 LOGIQ P9 with Articulating Monitor Arm 5724885 LOGIQ P7 R2 with High Cabinet 5724886 LOGIQ P9 R2 with Articulating Monitor Arm 5759916 LOGIQ P7 R2.5 5759915 LOGIQ P9 R2.5 5794264 LOGIQ P9 R3...
ANUARY ERVICE ANUAL 1-1-5 Purpose of Operator Manual(s) The Operator Manual(s) should be fully read and understood before operating the LOGIQ P9/P7 and also kept near the unit for quick reference. Section 1-2 Important Conventions 1-2-1 Conventions Used in this Manual...
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 1-2-2 Standard Hazard Icons Important information will always be preceded by the exclamation point contained within a triangle, as seen throughout this chapter. In addition to text, several different graphical icons (symbols) may be used to make you aware of specific types of hazards that could cause harm.
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 1-2-3 Product Icons The following table describes the purpose and location of safety labels and other important information provided on the equipment. Table 1-5 Product Icons...
Page 46
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Table 1-5 Product Icons (Continued) LABEL/SYMBOL PURPOSE/MEANING LOCATION "CAUTION - Dangerous voltage" (the lightning flash with arrowhead in equilateral triangle) is used to indicate electric various shock hazards.
Page 47
Unique Device Identification (UDI) Label. The UDI label consists of a series of alpha-numeric characters and barcode, which uniquely identify the LOGIQ P9/P7 system as a medical device manufactured by General Electric. Scan or enter the UDI information into the patient health record as required by country-specific laws.
Servicing should be performed by authorized personnel only. Only personnel who have participated in a LOGIQ P9/P7 Training are authorized to service the equipment. 1-3-3 Mechanical Safety The LOGIQ P9/P7 weighs 60 kg or more, depending on installed peripherals, (133 lbs or more) CAUTION when ready for use. !! CAUTION: Care must be used when moving it or replacing its parts.
• Never immerse the probe connector or adapter into any liquid. NOTE: For detailed information on handling probes, refer to the LOGIQ P9/P7 Basic User Manual and the care card supplied with the probe. Chapter 1 - Introduction...
The ultrasound system is an extremely sensitive and complex medical system. Any unauthorized peripherals may cause system failure or damage! The LOGIQ P9/P7 may be used with an isolation transformer to provide the required separation from mains for both, the system and the auxiliary devices.
Page 51
After each installation, the leakage currents have to be measured according to WARNING WARNING IEC 60601-1 respectively ANSI/AAMI ES60601-1. !! WARNING: All peripherals mounted on the LOGIQ P9/P7 system chassis must be firmly secured in position. NOTICE !! NOTICE: Chapter 1 - Introduction 1-11...
1-3-6 Battery Safety NOTE: The LOGIQ P9/P7 ultrasound system is supplied with a lithium ion battery in the power supply module, as option. The lithium ion technology used in the system’s battery is significantly less hazardous to the environment than the lithium metal technology used in some other batteries (such as watch batteries).
ANUAL 1-3-7 Labels Locations The LOGIQ P9/P7 ultrasound system comes equipped with product labels and icons. These labels and icons represent pertinent information regarding the operation of the unit. Figure 1-1 Label Location (R1, R2 and R2.5) * Required for Asia.
AC Outlet. 1-3-10 Returning/Shipping System, Probes and Repair Parts When returning or shipping the LOGIQ P9/P7 system in the original packaging: • system must be lowered to its minimum height with monitor flapped down (see Figure on...
In addition to electromagnetic energy, EMC also includes possible effects from electrical fields, magnetic fields, electrostatic discharge and disturbances in the electrical power supply. For applicable standards please refer to Chapter 2 in the Basic User Manual of the LOGIQ P9/P7 ultrasound system. 1-4-2...
2-1-1 Purpose of Chapter 2 This chapter provides the information required to plan and prepare for the installation of a LOGIQ P9/ P7 ultrasound unit. Included are descriptions of the facility and electrical needs to be met by the purchaser.
!! CAUTION: 2-2-1-1 Cooling The cooling requirement for the LOGIQ P9/P7 is 680 BTU/hr. This figure does not include cooling needed for lights, people, or other equipment in the room. NOTE: Each person in the room places an additional 300 BTU/hr. demand on the cooling system.
2-2-2-5 Main Power Plug The LOGIQ P9/P7 ultrasound system is supplied with a main power plug, as standard. In the event that the unit arrives without a power plug, or with the wrong plug, contact your GE dealer. When necessary, the installation engineer will supply the appropriate power plug to meet the applicable local regulations.
Ultrasound machines also generate EMI. The LOGIQ P9/P7 complies with limits as stated on the EMC label. However, there is no guarantee that interference will not occur in a particular installation.
Site preparation takes time. Begin Pre-installation checks as soon as possible. If possible, allow six weeks before delivery, for enough time to make necessary changes. Have two people available to deliver and unpack the LOGIQ P9/P7 ultrasound system. CAUTION Attempts to move the unit considerable distances (or on an incline) by one person alone, could !! CAUTION: result in personal injury and/or damage to the system.
2-2-6 System Specifications 2-2-6-1 Physical Dimensions of LOGIQ P9/P7 The physical dimensions and weight (without Peripherals) of the LOGIQ P9/P7 unit are summarized in Table 2-6. NOTE: Physical dimensions (especially height and depth) depend on control console and monitor positioning.
The products involved (and the accompanying electrical installations) are highly sophisticated and special engineering competence is required. All electrical work on these products must comply with the requirements of applicable electrical codes. The purchaser of GE equipment must only utilize qualified personnel to perform electrical servicing on the equipment.
NOTE: All relevant preliminary network port installations at the prepared site must be performed by authorized contractors. The purchaser of GE equipment must utilize only qualified personnel to perform servicing of the equipment. 2-3-3 Site Recommendations The following are (optional) site recommendations.
With DICOM, images can be archived, stored, and retrieved faster, easier, and at a lower cost. 2-3-4-4 DICOM Option Pre-installation Requirements To configure the LOGIQ P9/P7 ultrasound unit to work with other network connections, the network administrator must provide some necessary information. Use the...
Setup Instructions Section 3-1 Overview 3-1-1 The Purpose of Chapter 3 This chapter contains information needed to setup the LOGIQ P9/P7 ultrasound system. Included are procedures to receive, unpack and configure the equipment. A worksheet is provided (see: page 3-65 page 3-71) to help ensure that all the required information is available, prior to setup the system.
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 3-2 Set Up Reminders 3-2-1 Average Installation Time Once the site has been prepared, the average installation time required is shown in Table 3-2 below.
ANUAL 3-2-2 Installation Warnings 1) Since the LOGIQ P9/P7 weighs approximately 90 kg (200 lbs.) without peripherals, two people are required to unpack it. 2) There are no operator serviceable components. To prevent shock, do not remove any covers or panels.
!! CAUTION: (When covers are removed, EMI may be present). OPERATOR MANUAL(S) CAUTION The User Manual(s) should be fully read and understood before operating the LOGIQ P9/P7. !! CAUTION: Keep manuals near the unit for reference. ACOUSTIC OUTPUT HAZARD CAUTION...
ERVICE ANUAL Section 3-3 Receiving and Unpacking the Equipment Please read this section carefully before unpacking the LOGIQ P9/P7 ultrasound system and its CAUTION (optional) peripherals. !! CAUTION: The LOGIQ P9/P7 ultrasound system, together with peripherals, probes and accessories are shipped from the factory in a single durable shipping crate which is mounted on a raised wooden platform base.
Page 74
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 3-3 Receiving and Unpacking the Equipment (cont?) The envelope with delivery address, packing list and invoice is located on the front panel of the crate.
Page 75
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Table 3-5 Unpacking Procedure Step Task Remove the monitor pads from the monitor and remove the 4 pads of the box bottom assy. monitor pad Carefully remove foam packing material and plastic bag from the ultrasound unit and monitor.
EVISION ANUARY ERVICE ANUAL Section 3-4 Packing materials - recycling information The packing materials for LOGIQ P9/P7 are recyclable: • The Transportation Box is made of cardboard. • The inner reinforcements are made of Ethafoam (Polyethylene foam). • The plastic foil is made of LDPE (Low Density Polyethylene).
Check the power cord for cuts, loose hardware, tire marks, exposed insulation, or any deterioration. Power Cord Verify continuity. Replace the power cord, as required. NOTE: Report any items that are missing, back-ordered, or damaged, to your GE sales representative. The contact address is shown in Contact Information on page 1-17.
Verify that all EMI rules listed in the following table are followed: The LOGIQ P9/P7 ultrasound unit is approved for use in hospitals, clinics and other environmentally qualified facilities, in terms of the prevention of radio wave interference. Operation of the ultrasound unit in an inappropriate environment can cause electronic interference to radios and television sets situated near the medical equipment.
1,5 m 2,5 m 1,5 m 1,5 m NOTE: For more detailed Safety Considerations when connecting auxiliary devices to the LOGIQ P9/P7 system, please review: Chapter 1 - Auxiliary Devices Safety, on page 1-10. Chapter 3 - Setup Instructions 3-11...
There are three USB ports (stack) on rear panel. The BW printer may be connected to any of the 3x USB stack connector. 3) When all the cables are connected, press the Power ON switch on the Black & White printer. 4) Power ON/Boot up the LOGIQ P9/P7 system as described in Section 3-7-2 on page 3-21. All software drivers are pre-installed for the designated Black &...
Page 83
ANUARY ERVICE ANUAL 3-6-3 Connecting the Secondary “Patient” LCD Monitor Secondary LCD “Patient” Monitor MUST NEVER be connected to the LOGIQ P9/P7 ultrasound CAUTION systems mains supply directly! !! CAUTION: Always connect it to the supplied Isolation Transformer! The Secondary Monitor is the only item to be connected to the Transformer.
Composite (BNC) Cable S-Video Cable CAUTION !! CAUTION: A Secondary “Patient” Monitor MUST NEVER be connected to the LOGIQ P9/P7 ultrasound systems mains supply directly! Always connect it to an appropriate Isolation Transformer Figure 3-7 Connection Scheme - Secondary LCD Monitor 3-16...
ERVICE ANUAL 3-6-4 Connecting the Footswitch The footswitch should be directly connected to any accessible USB-port on the LOGIQ P9/P7 (e.g., on rear of the system). NOTE: Connection of the Footswitch is always the same (no differences between PC-Motherboard version of the system).
Devices on page 3-19. !! NOTICE: The USB Flash Memory Stick may be connected to an accessible USB port of the LOGIQ P9/P7 system. Refer to Section 3-9-2 "External I/O Connectors" An external USB Flash Memory Stick can be connected once the system is powered ON, or after shutdown.
When an external USB-storage device (such as a USB-memory stick or an external hard disk) is connected to the LOGIQ P9/P7, Windows detects the device and automatically installs a driver. During this process, several dialogs may pop up, starting with the “Found New Hardware“ dialog.
3-7-1 Connecting the Unit to a Power Source The connection of the LOGIQ P9/P7 ultrasound unit to a power source should be performed by a qualified person who has completed basic LOGIQ P9/P7 System User Training. Use only the power cords, cables and plugs provided by or designated by GE to connect the unit to the power source.
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 3-7-2 Power On / Boot Up 3-7-2-1 Scanner Power On 1) Ensure that the wall outlet is of appropriate type, and that the Circuit Breaker is turned off.
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 3-7-2-2 Back End Processor Boot Up When AC power is applied to the scanner, the ON/OFF button on the control panel illuminates blue, NOTICE indicating that the System (including the Back-end Processor) is in Standby mode.
4) Start-up progress bars indicating software loading procedures, are displayed on the monitor, as shown in Figure 3-11 below.Start-up progress bars indicating software loading procedures, Figure 3-11 GE Healthcare wallpaper with progress bar (R1.x.x) Figure 3-12 GE Healthcare wallpaper with progress bar (R2.x.x and later) Chapter 3 - Setup Instructions 3-23...
Page 92
5) Startup GE NOTE: Startup GE Healthcare wallpaper with progress bar may differ depending on product type and release. 6) The software initiates and sets up the Front-End electronics and the rest of the scanner (incl. the clicking sound of the relays on the PIL board).
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 3-7-3 Power Off / Shutdown When you power off the unit, the system performs an automatic shutdown sequence Figure 3-13 System Exit Window (R2 and R2.5)
Page 94
Use this button to exit from the System-Exit menu and return to the previous operation. • Exit (Only available when logged in as GE Service with Service Dongle) Select this button when you want to exit to the Windows Desktop.
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 3-7-3-1 Scanner Shutdown 1) If not already in read mode, freeze the image. 2) Press the ON/OFF button (see: Figure 3-10) on the control panel. Following dialog appears.
Page 96
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 3) Select the SHUTDOWN button. The system performs an automatic full shutdown sequence. 4) Switch OFF the Circuit at the rear of the system. NOTE: The mains outlet of the system for peripheral auxiliary equipment are commonly switched with the ON/OFF button.
Page 97
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 5) After complete power down, disconnect the main power cable from the system or unplug it from the AC wall outlet socket. Circuit Breaker Main Power Figure 3-17 Circuit and main power at rear of system 6) Press on the brakes to block the front caster wheels.
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 3-7-4 Transducer Connection 3-7-4-1 Connecting the Probe Inspect the probe before and after each use for damage or degradation to the housing, strain CAUTION relief, lens, seal, cable and connector. DO NOT use a transducer which appears damaged until !! CAUTION: functional and safe performance is verified.
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 3-7-4-2 Connecting the CW Pencil Probe Insert the probe connector into the probe port all the way seated in. Carefully position the probe cord so it is free to move and is not resting on the floor.
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 3-8 Printer Installation NOTE: For Connection schemes refer to Section 3-6 "Connection of Auxiliary Devices" on page 3-11. For further installation instructions see: • Section 3-8-1 "Installing Digital Black & White Printer Sony UP-D898" on page 3-32 •...
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 3-8-2 Digital Color Printer Sony UP-D25MD 1) Power off/Shutdown the system as described in: Section 3-7-3 on page 3-25. 2) Connect USB cable into the available USB port on the rear of the system. There are three available USB ports on rear of the system.
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 3-8-3 Installing the Bluetooth Deskjet Printer 1) Power off/Shutdown the system as described in: Section 3-7-3 on page 3-25. 2) Physically connect the Bluetooth dongle as described on Section 3-8-3-1 on page 3-34.
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 3-8-3-2 Verifying Bluetooth device driver 1) Turn on the system. 2) Enter maintenance mode by using service dongle. 3) Go to Device Manager by pressing shortcut on the taskbar 4) Check that there is “Bluetooth Radios”...
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 3-8-3-3 Prepare the Printer and connect the Bluetooth Printer Adapter 3-8-3-3-1 HP Officejet 100 1) Connect the power cord to a wall outlet socket. 2) Turn on the printer.
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 3-8-4 Adding Printer to the system 1) Select Utility -> Connectivity -> Service. 2) Add the Standard Print. 3) Highlight Standard Print in the Service list.
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 3-8-5-2 UP-D25MD - Printer Settings 1) Call up the ‘Printer Preferences’; operation see: Section 3-8-5 "Adjustment of Printer Settings" Settings for Paper Size MUST match with the used Paper (large/small) and also the right color ink NOTICE cartridge has to be used.
Page 108
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL b) Color Correction: set Printer Hardware Color Correction1 Figure 3-27 Graphics page (Color Balance + Color Correction) c) Lightness: Sharpness = 7; Dark = 0; Gamma = 14; Light = 0...
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 3-8-6 Setting Printer to Print Reports 3-8-6-1 Setting up the Printer to Print Reports To set up the Off-Line Printer to print reports, 1) Press Utility -> System -> Peripherals.
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 3-9 System Configuration 3-9-1 System/General Preset Menu The System/General screen allows you to specify hospital name and system date and time. Figure 3-30 System/General Preset Menu (R1.x.x) Figure 3-31 System/General Preset Menu (R2.x.x)
Page 111
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Figure 3-32 System/General Preset Menu (R3.x.x) Table 3-8 Location Preset Parameter Description Hospital Type the institution’s name. Department Type the institution's department name. Preset Region (restart needed) Select region (None, Americas, Asia, Europe or Japan).
Page 112
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Table 3-10 General User Interface Preset Parameter Description Color Level (restart needed) Select System Color according to the condition of the room. Table 3-11 Title Bar...
Page 113
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Table 3-15 Scan Assistant Preset Parameter Description Auto Selection Off, Category, or Description. • Off. The Scan Assistant selection on the Patient screen is completely manual. It will say “None”...
Section 3-9-1 "System/General Preset Menu" on page 3-42 More detailed information pertaining System Setup and Measure Setup adjustments is found in the NOTICE LOGIQ P9/P7 Basic User Manual, which is available in different languages. !! NOTICE: Chapter 3 - Setup Instructions...
By utilizing data flows, the user can configure the LOGIQ P9/P7 ultrasound unit to optimally meet the needs of the facility, while keeping the user interface unchanged. Once the dataflow is selected, the actual location of the database is entirely transparent.
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 3-12-1-2 Dataflow Examples The local database is used for patient archiving. Images are stored to internal hard drive. The local database is used for patient archiving.
3-12-1-4 LOGIQ? + PC within a “Sneaker Net” A PC is used for review and work on studies acquired on one or more LOGIQ P9/P7 scanners without being connected in a network. The images are first stored on the LOGIQ P9/P7 scanner’s hard drive (Archive) and then exported from the scanner’s hard drive to a sneaker device (e.g., DVD/CD), and finally imported from the sneaker...
You will need a network switch and one network cable for each unit connected to the switch. 3-12-2-3 Connection via Hospital Network 3-12-2-4 You will need one network cable to connect the LOGIQ P9/P7 to a wall outlet on the hospital’s network. 3-12-3 Connection using Wireless Option Refer to: Chapter 5 for theory.
This greatly improves query capability. DICOM Structured Reporting creates coded clinical data that can be used for clinical research, outcomes analysis, and disease management. DICOM Structured Reporting is a standardized format for medical results. LOGIQ P9/P7 supports the following templates: •...
Page 127
The ultrasound system is pre-configured for many services, with default settings selected. You can change these services and settings as needed. You must restart the LOGIQ P9/P7 (shutdown) after making any changes to connectivity CAUTION settings in the Utility menus. This includes any changes on the TCPIP or dataflow setup !! CAUTION: screens.
3) Select Save settings. 4) Re-boot the ultrasound system. NOTE: TCPIP settings do not get restored when restoring backups. This is per system design. The LOGIQ P9/P7 IP address MUST BE unique. Figure 3-36 Connectivity TCPIP Preset Menu Table 3-22 Computer Name...
Page 129
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Table 3-23 Wired Network Configuration Preset Parameter Description Type the IP Address of the Ultrasound system. NOTE: IP stands for Internet Protocol. Every IP-Address device on the network has a unique IP address.
Properties: IP Address Type the device's IP address. Properties: AE Title AE Title of the LOGIQ P9/P7. NOTE: Only available for MyComputer. Properties: Port Number IP Port Number Used for DICOM, set by default to 104. NOTE: Only available for MyComputer.
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 3-13-6 Service For each Device that you added to the system, you need to set up the service(s) that device supports (you must be an administrator to update these screens).
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 3-13-6-3 Changing parameters for a service There are certain parameters that may need to be set up for each service: Table 3-26 Service Parameters: Common Service Parameters...
Agent Configuration (R3 only) Overview System may be configured to connect to the back office server (FFA Site), which enables GE Service Personnel to connect to the system to monitor health of the system, perform diagnostics, exchange files, or view console desktop.
Page 134
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Use Agent Configuration to: • Edit and configure the following: - Enterprise server, Select “Production” for install base. Select “Staging”for test purpose - Enterprise host name in the agent - Enterprise port number in the agent - Proxy server, Proxy port in the agent, Consult institutions’...
Verifying Back Office Connection - CSD 1) Login to FFA Site from Remote PC using Chrome web browser As of 2018, FFA Site address is https://ffa.am.health.ge.com 2) In System ID field, type the CRM Number of the machine and select [Get Started] Figure 3-41 FFA Site 3) At CONNECT mode, Select [Connect] of the “Service Desktop”...
ANUAL 3-14-1-3 Verifying Back Office Connection – VOC 1) On Console, invoke “Disruptive Mode” and “VCO” by left click “GE ICON” and select “Connect Clinical Lifeline Figure 3-43 Connecting Clinical Lifeline 2) Login to FFA Site from Remote PC using Chrome web browser As of 2018, FFA Site address is https://ffa.am.health.ge.com...
Page 137
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 4) At CONNECT mode, Select [Connect] of the “Virtual Console” on Remote Console Figure 3-45 Remote Console 5) After selecting [Connect] button, operator may need to wait long time to launch the remote service...
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 3-15 Connectivity Setup Worksheet Site System Information Comments: Floor: Site: Dept: Room: LOGIQ P9/P7 Type: REV: CONTACT INFORMATION Name Title E-Mail Address Phone TCP/IP Settings...
Page 139
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL LOGIQ P9/P7 IP Address Local Port Host Name AE Title Net Mask ROUTING INFORMATION GATEWAY IP Addresses Destination IP Addresses Default ROUTER1 ROUTER2 ROUTER3 DICOM APPLICATION INFORMATION...
Check that the correct User Manual(s) for the system and software revision, is included with the installation. Specific language versions of the User Manual may also be available. Check with your GE Sales Representative for availability. 3-72 Section 3-16 - Paperwork...
OPIO Interface Check (R3 only) 4-3-10 4-39 Touch Panel Calibration (R3 only) 4-42 Board Diagnostics Most of the information pertaining to this Functional Checks chapter is found in the LOGIQ P9/P7 Basic NOTICE User Manual; !! NOTICE: Section 4-2 Required Equipment •...
Page 142
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL • At least one transducer. See Chapter 9 - Probes, on page 9-30 for an overview. (normally you should check all the transducers used on the system)
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 4-3 General Procedure SYSTEM REQUIRES ALL COVERS CAUTION Operate this unit only when all board covers and frame panels are securely in place. !! CAUTION: The covers are required for safe operation, good system performance and cooling purposes.
Rating Plate design slightly differ depending on Console destination. At minimum simply check Console number, manufacturing date and serial Number. 5) Check presence of UL Label. (But R3 system have UL symbol in LOGIQ P9/P7 RATING LABEL) Chapter 4 - Functional Checks...
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 4-3-2 Mechanical Parts Functional Check Pay attention to moving parts and pinch point when performing mechanical parts functional CAUTION check. !! CAUTION: Figure 4-4 System Mechanical Checkpoint...
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 4-3-2-4 Swivel/Brake Lock Caster System weighs approximately 60kg. Pay attention for system stability when working with the CAUTION casters !! CAUTION: NOTE: This procedure applies to RIGHT REAR Caster only.
Page 152
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 4-3-2-5 Swivel/Brake Lock Caster System weighs approximately 90kg. Pay attention for system stability when working with the CAUTION casters. !! CAUTION: NOTE: This procedure applies to casters other than RIGHT REAR Caster.
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 4-3-3 Power On/Off After turning off a system, wait at least 10 seconds before turning it on again. NOTICE The system may not be able to boot if power is recycled too quickly.
Page 154
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL OFF (STAND-BY) (R3 only) OFF (Stand-by) ON (R3 only) Figure 4-12 On/Off Button on Control Panel and indicator color 6) Verify On/Off button color changes to GREEN.
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 4-3-3-2 Sleep Mode Check NOTE: This procedure assumes systems is already up and running. 1)To going into the sleep mode, enter the scan screen and lightly press the Power On/Off switch (see: Figure 4-13) once.
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 4-3-3-3 Complete power down After turning off a system, wait at least 10 seconds before turning it on again. The system may not be NOTICE able to boot if power is recycled too quickly.
Use this button to exit from the System-Exit menu and return to the previous operation. • Exit (Only available when logged in as GE Service with Service Dongle) Select this button when you want to exit to the Windows Desktop.
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 4-3-4 System Information 4-3-4-1 Software Version 1) From the touch panel, select Utility -> System -> About. 2) Note “Software Version” in Software Group. 3) Note "Software Revision" in Software Group.
4-3-4-2 Service Platform Confirmation (R1, R2, R2.5 only) 1) From the touch panel, select Utility -> Service. 2) On Service Browser, select GE Service, and enter the password. 3) Note “SvcPform” version. Figure 4-17 Software Platform version Chapter 4 - Functional Checks...
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 4-3-4-3 USB Port Test Required Equipment : Service Dongle External Port Test 1) Plug service dongle to the following locations. Refer to Section 3-9-2 for locations.
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 4-3-5 System Integration Checks 4-3-5-1 OPIO Test Figure 4-18 OPIO 1) Power ON the system. 2) Check backlight on AN Keyboard, B Buttons, Freeze Key. 3) Select UTILITY in touch panel.
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Figure 4-20 6) Connect any probe. (if not connected) 7) Go to Live CF Mode. (Mode Rotary Line Area in touch panel, Press “CF” Mode button, unfreezes.) 8) Turn “GAIN”...
4) Start Recording. Perform image scanning including Doppler sound. 5) Stop Recording. 6) Exit from stand-by. 7) Play. 8) Verify that Recorded image and sound can be played back. NOTE: LOGIQ P9/P7 support USBHDD file capture. and this menu can select in " Utility " " System "...
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 4-3-6-6 Gel Warmer Test 1) Insert Gel Bottle into Gel Warmer. 2) Power ON the warmer. 3) Verify Gel gets warm. NOTE: Depending on environment, Gel requres 30 minute or more to warm up.
Page 165
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 9) Press "Save" and restart the system into Question window by pressing "OK" Button. Figure 4-21 User Configurable Key Page (R1.x.x) Figure 4-22 User Configurable Key Page (R2.x.x or later)
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 4-3-7-6 M-Mode 1) Press B Mode Button. 2) Press M Mode Button. (Active M Mode button on "User Configurable Mode Rotary".) 3) Verify that system scan B/M with the activated probe.
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 4-3-7-9 WLAN (Optional) NOTE: This test is only applied if the WLAN option has been installed 1) Connect a service dongle to USB port and press a power button on console panel.
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 4-3-8 OPIO Interface Check (R1, R2, R2.5 only) NOTE: In above example, key “Alt” is pressed. NOTE: Alternatively, OP Panel Interface test may be executed from Common Service Desktop.
Page 174
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 3) Click the "YES" on Diagnostic Mode page Figure 4-34 Diagnostic Mode page screen after enable the Disruptive mode 4) Select Diagnostics -> OP Panel Interface...
Page 175
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 5) Click the "Start" on OP Panel Interface page Figure 4-36 OP Panel Interface page screen 6) Select "Op-Panel Open" 7) Verify OP Panel button operation is reflected on diag window.
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 4-3-9 OPIO Interface Check (R3 only) NOTE: Alternatively, OP Panel Interface test may be executed from Common Service Desktop. However, performance / response of the diagnostics may be degraded when executed with scanner software still active.
Page 177
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 4) Select Diags -> Run Diags -> Service Menu -> OpPanel -> OpPanel Test Figure 4-39 OP Panel Interface page screen 5) Click the "Start" on OP Panel Interface page Figure 4-40 OP Panel Interface page screen 6) Select "Op-Panel Open"...
Page 178
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL NOTE: If four touch points are out of possible error range, repeat these 8-9 steps three times. Chapter 4 - Functional Checks 4-38...
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 4-3-10 Touch Panel Calibration (R3 only) 1) Power ON the scanner, and wait until the main screen comes up. 2) Insert the service dong to the front USB port.
Page 180
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 9) Press the Reset button in Display tab -> Display options -> Details Figure 4-41 Tablet PC Settings dialog 10) Press the OK when pop-up the window like section 4-42 Figure 4-42 Digitizer Calibration Tool 11) Press “Calibrate...”...
Page 181
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 12) Calibrated carefully 16 calibration points in touch panel. Figure 4-43 Calibration Layout NOTE: Figure 4-35 is intended to be shown upside-down. 13) Press the Yes when pop-up the window like section 4-44...
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 4-4 Board Diagnostics Please see Section 7-6 for the details. NOTICE !! NOTICE: Battery Category Item DC4D Other Module MCB Swept Demodulator Test MCB Front End Interface Test...
Page 183
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Essential Test COMEXP HD Test Long HD Test Short Memory COMEXP COMEXP/ Network Interface EthernetPort COMEXP/ Network Interface Reliability Test EthernetPort RTC Clock COMEXP Video cards...
Components and Functions (Theory) Section 5-1 Overview 5-1-1 Purpose of Chapter 5 This chapter explains LOGIQ P9/P7’s system concepts, component arrangement, and subsystem function. It also describes the Power Distribution scheme and probes. Table 5-1 Contents in Chapter 5 Section Description...
ANUARY ERVICE ANUAL Section 5-2 General Information LOGIQ P9/P7 is a digital beamforming curved-, linear- and phased array ultrasound imaging system. The system supports various operating mode such as • B-Mode, M-Mode, Color Flow (CFM) • Power Doppler Imaging (PDI), PW Doppler, CW Doppler •...
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 5-2-4 System Ports USB Ports Rear Panel Figure 5-5 System Ports System is equipped with 4 USB ports in front. These are customer use and Service port connected to USB controller via USB Hub.
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 5-2-4-1 Front USB Customer use and Service port connected to USB controller via USB Hub. NOTE: Front USBs are non-isolated port and not fit to connect external powered USB Devices.
Page 192
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL LAN : 1000Base-T, 100Base-T, 10Base-T USB : USB 2.0, 480Mbps, 5V 500mA. NOTE: Rear USBs are non-isolated port and not fit to connect external powered USB Devices.
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 5-2-4-3 HDMI NOTE: HDMI for external monitor output is DVI-D. However, connector shape is HDMI connector. Table 5-6 Pin Out Pin number Signal name TMDS Data 2+...
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 5-3 Power ON Sequence Sequence 1) Circuit breaker ON 2) AC/DC in IOBOX generates standby 5V 3) Standby 5V is supplied to OPIO, COM Express and each board...
% stenosis, volume, trace, circumference, and enclosed area. Coded Contrast Imaging LOGIQ P9 only : Provides contrast imaging capability CW Doppler Allows examination of blood flow data all along the Doppler Mode cursor rather than from any specific depth.
Page 198
The velocity of the shear wave is used to generate an elasticity image. HRes Contrast Upgrade LOGIQ P9 only : Provides additional high resolution contrast imaging modes. Tablet Tools LOGIQApps are available for Android devices via Bluetooth: Remote Control App and Photo/Barcode Reader App.
Page 199
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL NOTE: In case the user scans any other barcodes in the manual below apart from “USB Serial” then the barcode reader might be unable to scan the QR Code on the Pinpoint Needle or Patient or worklist. In this case, please re-scan the “USB Serial”...
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 5-4-2 Smart Devices Applications (LOGIQ Apps) The Smart Devices Applications are available for Android via Bluetooth: • Remote Control App • Camera App To link to the device: 1.) Insert Bluetooth dongle into USB slot...
Printers Description LOGIQ P9/P7 Sony UP-D898 Digital B/W Printer Kit B/W Printer Kit LOGIQ P9/P7 Sony UP-D25MD Digital Color A6 Printer Kit Digital Color A6 Printer Kit LOGIQ P9/P7 Color Inkjet Printer (Off-board) Inkjet Printer Chapter 5 - Components and Functions (Theory)
• Depth: 740mm • Weight: 60kg (133 lbs) 5-7-2 LCD Monitor The LOGIQ P9/P7 has a free adjustable LCD monitor in relation to the user interface. • position up/down: +/- 7.5 cm • position left/right: +/- 18cm • rotation up/down: +90°/-15°...
Clean the Air Filter periodically. Through the filter grid on the front of the system, air flow into the LOGIQ P9/P7 scanner. By means of the 2 FAN, air is blown through the nest-box, and the warm air exits the scanner through holes in the left side panel and rear of the system.
ANUAL Section 5-8 Exiting to Windows Desktop ® LOGIQ P9/P7 backend software operates on embedded version of Windows 7 operating system (R2 ® and R2.5) or ersion of Windows 10 Enterprise operating system (R3). To access some of the functions, service personnel occasionally require to enter Windows desktop to access files, folders or applications.
Page 207
5) On “Maintenance Access” window, type in service password (user name not required). NOTE: Service Password is rotating password that changes on 6 months period. Only the GE service personnel or GE authorized dealers have access to maintenance password. NOTE: Maintenance Password is function of date/time.
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 5-9 Service Platform 5-9-1 Introduction The Service Platform will increase service productivity and reduce training and service costs. 5-9-2 Access / Security The Service Platform has different access and security user levels.
Page 209
5-9-2-2 How the Customer enables/disables Disruptive Mode and VCO 1) If not already in read mode, FREEZE the image. 2) Move the cursor to the GE icon and press the (= right-click). right trackball key 3) Select Connect Clinical Lifeline (see: Figure 5-15 on page 5-25 (R1, R2, R2.5 only) or...
Page 210
5-10-1 Internationalization The user interface provided by the service platform is designed for GE personnel and as such is in English only. There is no multi-lingual capability built into the Service Interface. There are different possibilities to access the Common Service Desktop and its available features:...
Page 211
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 5-10-2 CSD Top Page 5-10-2-1 General Layout Table 5-12 System Information System ID, Serial, IP Address, InSite StatusÅc.etc Connected Probes Connected and System recognized probe listed...
Page 212
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 5-10-2-2 Decoding System Temperature and Voltage NOTE: These are reference only values and therefore, parameters are not constant. Some parameters may be added/deleted depending on the system configuration, type, and variation.
Page 214
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 5-10-3 Error Logs When the Error Logs page is selected, different log viewing options are available. Log Viewer is displayed in a separate window. 5-10-4 Diagnostics...
Page 215
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 5-10-5 Image Quality In the Image Quality page, you can verify image quality. Figure 5-20 Common Service Desktop - Image Quality NOTE: This page is not populated in this version.
Page 216
Figure 5-21 Common Service Desktop - Configuration Remote access is ONLY possible if the service platform is properly configured (either by the user NOTICE or a GE technician at site). Operation see: Section 7-5-6 "CSD: Configuration" on page 7-18. !! NOTICE: 5-32 Section 5-10 - Common Service Desktop (CSD) (R1, R2, R2.5 only)
Page 217
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 5-10-7 Utilities The Utilities page contains a variety of Windows utility tools to indicate the status of the system, in addition to various other tools. Figure 5-22 Note on Utilities: ®...
Page 218
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 3) Select “Disk Defragmenter” from left pane Figure 5-24 4) Select Disk Drive to defragment on right (upper) pane 5) Press “Defragment” to execute defragmentation of the selected disk NOTE: Press “Analyze”...
Page 219
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 5-10-8 Replacement There is no accessible parameters in Replacement page. 5-10-9 In the PM page, information about planned, proactive and preventive maintenance is displayed, as described in Chapter 10 - Care &...
Page 220
• Local user – a user locally logged into the machine will be able to set the LOGIQ P9/P7 to Disruptive mode or allow a Disruptive mode request from a remote user through the Service desktop. The local user must have Authorize Remote Service Access to allow Disruptive mode.
Page 221
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 5-11-3 Color statuses Throughout the Service desktop, colors indicate the following: • Green – Status is normal • Orange – Status is a warning • Red – Status is an error...
Page 222
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL With Service Advanced (Class C), these are the available options: • HOME • Diags - Run Diags - Diags History • DICOM - DICOM Service Status - DICOM Spooler •...
Page 223
LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL With GE Service access (Class M) and an SSA key, these are the available options: • HOME • Diags - Run Diags - Diags History • DICOM...
Page 224
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL With Remote access (Back office server), these are the available options: • HOME • Diags - Run Diags - Diags History • DICOM - DICOM Service Status - DICOM Spooler •...
Page 225
ANUAL 5-11-5 Advanced Information - Service Desktop and SSA (Secure Service Access) SSA Key The SSA key provides secure access for GE service personnel to advanced tools to service the system • Access to all service features • Access to Windows Desktop •...
Page 226
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 5-11-6 Home Home configurations vary depending upon the purchased service level. Figure 5-27 Home with Class C and Class M Access Figure 5-28 Home with Class A Access For more information, see: •...
Page 227
EVISION ANUARY ERVICE ANUAL System Information System Information displays general information about the LOGIQ P9/P7. When the LOGIQ P9/P7 has been successfully configured with the back office, these elements will have the corresponding values: • Agent Registered will be Yes •...
Page 228
• No – The agent is not verified in the back office. • Not Available – The agent is not running or has not been configured. GE part number for the LOGIQ P9/P7. The same number as listed on the rating Model Number plate.
Page 229
ANUAL Software Status Use Software Status to view general information about the software installed on the LOGIQ P9/P7. The information on Software Status is available to all service class licenses. To access Software Status, navigate to Utility > Service > Home.
Page 230
ANUARY ERVICE ANUAL System Status Use System Status to view status information on the LOGIQ P9/P7. Specifically, do the following: • View the hardware configuration file • View temperatures for FRUs, fans, and a graphical representation of the monitored temperatures •...
Page 231
30 days or last status checkpoint whichever is sooner. Based on the latest occurrence, displays all abnormal boots of the LOGIQ P9/P7 for Bootup Issues either the past 30 days or the latest check point.
Page 232
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Hardware Configuration Hardware Configuration displays the contents of the hardware configuration file which includes all the vital product data (VPD). To access Hardware Configuration, navigate to Utility > Service > Home, and then under System Status, select Hardware Configuration.
Page 233
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Temperature Use Temperature to troubleshoot problems with monitored FRUs and the fans. Specifically, use these pages: • FRU Status • Graphs To access Temperature, navigate to Utility > Service > Home, and then under System Status, select Temperature.
Page 234
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL FRU Status FRU Status displays a summary of the FRUs being monitored and the details for each FRU. To access this page, under System Status, select Temperature and then select FRU Status.
Page 235
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Graphs Graphs displays trend graphs for selected elements which have been selected as key indicators of overall temperature status inside the card rack. To access this page, under System Status, select Temperature and then select Graphs.
Page 236
EVISION ANUARY ERVICE ANUAL Voltage Use Voltage to troubleshoot problems with monitored FRUs of the LOGIQ P9/P7. Specifically, use these pages: • FRU Status To access Voltage, navigate to Utility > Service > Home, and then under System Status, select Voltage.
Page 237
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL This table shows all the elements available on FRU Status with descriptions. Table 5-19 FRU Status Element DESCRIPTION Name Name of the FRU. Status of the voltage for the FRU. Valid values are: •...
Page 238
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Bootup Issues Bootup Issues displays a summary of bootup issues detected by the system that prevented normal scanning, shutdowns, or abnormal statuses. To access this page, under System Status, select Bootup Issues.
Page 239
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL This table shows all the elements available on Bootup Issues with descriptions. Table 5-20 Bootup Issues Element DESCRIPTION Identifier for the system software boot. For example, 10 for the tenth boot and so on.
Page 240
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Alert Logs Alert Logs displays the details for each alert. Alerts are status logs typically reported as a result of an error. Alerts are sent to the back office to be managed and reported to service.
Page 241
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL USB Devices USB Devices displays the details for each USB device. To access this page, under System Status, select USB Devices. Figure 5-39 USB Devices This table shows all the elements available on USB Devices with descriptions.
Page 242
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Disk Drive Space Use Disk Drive Space to view the hard drive partitions including total size and available free space in GB. Includes the overall health of the drive in one of these colors: •...
Page 244
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL This table shows all the elements available on Disk Drive Space with descriptions. Table 5-23 Disk Drive Space Element DESCRIPTION C: partition shows the graphical representation of the following: C: Drive •...
Page 245
Network Configuration Use Network Configuration to view network (wired and wireless) information (the full ipconfig details) for the network configured with the LOGIQ P9/P7. The information on Network Configuration is available to Class C and Class M licenses. To access Network Configuration, navigate to Utility > Service > Home.
Page 246
This table shows all the elements available on Network Configuration with descriptions. Table 5-24 Network Configuration Element DESCRIPTION Host Name Name of the local host for the LOGIQ P9/P7. More Details Displays the ipconfig details. Wired Connection lpV4 Address Local IP address for the wired network connection.
Page 247
ANUAL Connected Probes Connected Probes shows probes connected to the LOGIQ P9/P7. The order on the user interface is top down matching the left-to-right order on the LOGIQ P9/P7. The information on Connected Probes is available to all service class licenses.
Page 248
Run Diags – Run diagnostics on hardware components. • Diag History – View a history of the diagnostics you have run and the results. NOTE: Reboot the system after performing any diagnostics before returning the LOGIQ P9/P7 to customer use. For more information, see: •...
Page 249
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL This table shows all the elements available on Run Diags with descriptions. Table 5-26 Run Diags Element DESCRIPTION Service Menu Available service diagnostics in a diagnostics tree.
Page 250
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL To run a diagnostic: 1.) Navigate to Utility (second page) > Service > Utilities > Disruptive Mode Utility and set Disruptive mode to On. 2.) Navigate to Utility (second page) > Service > Diags > Run Diags.
Page 251
ANUAL Diag History Use Diag History to view a history of the diagnostics run on the LOGIQ P9/P7 and the results. The information on Diag History is available to Class C and Class M licenses. To access Diag History, navigate to Utility > Service > Diags > Diag History.
Page 252
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL To view a history of diagnostics: 1.) Navigate to Utility (second page) > Service > Diags > Diag History. 2.) Select the options that you want to view.
Page 253
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 5-11-8 DICOM Use DICOM to troubleshoot DICOM components and connectivity. Specifically, use these pages: • DICOM Service Status • DICOM Spooler For more information, see: •...
Page 254
Application entity title of the LOGIQ P9/P7. Port No of Console Port number of the LOGIQ P9/P7. DICOM Connectivity Status Service Name Name of the DICOM services configured for the LOGIQ P9/P7. Status of the DICOM service. Valid values are: • OK • Not Ok Status •...
Page 255
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL DICOM Spooler DICOM Spooler provides a summary of the current and recent outgoing jobs. The information on DICOM Spooler is available to Class C and Class M licenses.
Page 256
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL This table shows all the elements available on DICOM Spooler with descriptions. Table 5-29 DICOM Spooler Element DESCRIPTION Job Status Displays the number of jobs for each type by status. For example, Pending.
Page 257
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 5-11-9 Utilities Utilities configurations vary depending upon the service class. For more information, see: • Change Password on page 5-79 • Checkpoints on page 5-91 •...
Page 258
Disk fragmentation can reduce the amount of disk space available, and slow computing speed. Use the disk defragmenter to restore optimum disk space and speed performance. If LOGIQ P9/ P7 performance would be significantly reduced when the disk is being defragmented, a warning message displays.
Page 259
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Gather Logs Gather Logs provides a way to collect system logs and place the log files in the D:\Service directory for retrieval by the online center. These log files do not include protected data such as crash dumps and keyboard shadow logs.
Page 260
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Delete Files Delete Files displays all the files and folders present in the D:\Service folder and allows for their deletion. Deleting unneeded files improves performance and reduces the need to defragment the disk drive.
Page 261
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Disable Remote Connectivity This utility will uninstall remote connectivity items like Agent Configuration, SSH and Virtual Console Observation (VCO) from Service Desktop. If remote connectivity is disabled, then the Submit Changes button on Agent Configuration and the Start/Stop buttons for SSH and VCO will all be disabled.
Page 262
3.) Select Start. When VCO starts, the Windows task bar and Windows desktop display. 4.) Modify settings and programs as needed. 5.) To stop virtual console observation, click Stop. When VCO is stopped, the Windows desktop is hidden. 6.) Restart the LOGIQ P9/P7. For more information, see: • Utilities on page 5-73...
Page 263
RISK OF LOSING DATA. CAUTION IF THE PASSWORD IS LOST, GE WILL NOT BE ABLE TO RECOVER OR RESET IT. LOSS OF A PASSWORD MAY RESULT IN THE LOSS OF PATIENT DATA. The information on Change Password is available to all service class licenses. Change Password is not available through a remote connection.
Page 264
GE WILL NOT BE ABLE TO RECOVER OR RESET CHANGED PASSWORDS. SECURELY RECORD THE NEW PASSWORD. 4.) Click Update Password. 5.) When a SVCService user password has been changed, reboot the LOGIQ P9/P7 to reflect the password change. For more information, see: •...
Page 265
ANUAL Software Reload Software Reload displays versions of the base image, application software, base patch, service platform, and documentation packages currently installed on the LOGIQ P9/P7. Software Reload reimages the C:\ drive and installs the base image. • If the drives are encrypted, Software Reload will have a catastrophic effect if the customer does not have the recovery key.
Page 266
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL To reload software: 1.) If the drives are encrypted, obtain the recovery key. 2.) Navigate to Utility (second page) > Service > Utilities > Disruptive Mode Utility and set Disruptive mode to On.
Page 267
ANUAL Reset Patient Database Reset Patient Database resets the LOGIQ P9/P7 patient database. This operation automatically shuts down the application before cleaning the patient database. Once the operation is complete, the LOGIQ P9/P7 reboots with an empty database. Disruptive mode must be On to reset the patient database. If Disruptive mode is Off, a warning message displays asking to turn on Disruptive mode.
Page 268
On to clean user-defined data. If Disruptive mode is Off, a warning message displays asking to turn on Disruptive mode. Before cleaning user-defined data, the LOGIQ P9/P7 application software automatically shuts down. Once the operation is complete, the system reboots.
Page 269
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL To clean user-defined data: 1.) If you want to use the current settings later, back them up now. 2.) Navigate to Utility (second page) > Service > Utilities > Disruptive Mode Utility and set Disruptive mode to On.
Page 270
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Third Party Software Licenses Third Party Software Licenses displays the third-party software licenses used as part of the service platform. The information on Third Party Software Licenses is available to all service class licenses.
Page 271
Class M Key Status • Not Plugged In • Plugged In Drive Letter Drive where the SSA key is plugged into the LOGIQ P9/P7. Expire Date Date the SSA key is set to expire. SSO ID Identifier for the user assigned to the SSA key.
Page 272
2.) If open, close the Service desktop. 3.) Restart the LOGIQ P9/P7. 4.) Once the LOGIQ P9/P7 has restarted, plug in the SSA key. 5.) Navigate to Utility (second page) > Service > Utilities > SSA License. 6.) Click Restore SSA.
Page 273
System Shutdown With remote access, System Shutdown provides a way to shut down or restart the LOGIQ P9/P7. Disruptive mode must be On to shut down or restart the LOGIQ P9/P7. If Disruptive mode is Off, a warning message displays.
Page 274
ANUARY ERVICE ANUAL To remotely start, restart, or shut down the LOGIQ P9/P7: 1.) Navigate to Utility (second page) > Service > Utilities > Disruptive Mode Utility and set Disruptive mode to On. 2.) Navigate to Utility (second page) > Service > Utilities > System Shutdown and then select an operation: To shut down the LOGIQ P9/P7, select Shutdown System.
Page 275
Checkpoints allows for the creation of checkpoints and displays the history of those checkpoints. Checkpoints are used to indicate when the LOGIQ P9/P7 has been serviced, status has been reset, and information related to the service performed has been added.
Page 276
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL To create a check point: 1.) Navigate to Utility (second page) > Service > Utilities > Checkpoints. 2.) Under Create Checkpoints, from Action, select the action.
Page 277
If No is selected, then Disruptive mode is not turned on and an error message displays saying that permission to change the mode was denied. When Disruptive mode is On, the GE icon in the status bar changes to red and displays this message: "Due to Service testing restart needed."...
Page 278
• View information about past transfers of (APM) information. • Set up automatic/scheduled transfer of allowed data files from the LOGIQ P9/P7 to the server. • Manually transfer data files allowed data files from the LOGIQ P9/P7 to the server.
Page 279
Transfer Machine Logs to GE – Default is checked Transfer Operator Information to GE Transfer Images (captured without patient information) to GE – Default is checked. 3.) Click Save. 4.) Navigate to Utility (second page) > Service > Utilities > Data Transfer.
Page 280
This protocol is used for remote login to the operational system. SSH allows remote login to the LOGIQ P9/P7 and remote file transfer from the LOGIQ P9/P7. Remotely log into the LOGIQ P9/P7 using the username “GEService” and its password.
Page 281
ANUAL Network Capture Network Capture displays network traffic between the LOGIQ P9/P7 and configured devices. A network capture outputs two log files: one for main logging with no protected information and another including protected information. These log files are useful when debugging connectivity issues.
Page 282
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL To perform a network capture: 1.) Navigate to Utility > Service > Utilities > Network Capture. 2.) From Network Capture, do the following: Under Maximum Size, select the allowed size of the generated log file.
Page 283
Software Options Option Software options on the LOGIQ P9/P7. Status Status of the options on the LOGIQ P9/P7. Software Option Details Product Name of the product. Number for the hardware. The hardware number is the hash of the serial number Hardware Number that is used to generate the option key.
Page 284
5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Table 5-43 Options Element DESCRIPTION Software Option Keys Available Keys List of the option keys installed on the LOGIQ P9/P7. 5-100 Section 5-11 - Common Service Desktop (R3 only)
Page 285
Element DESCRIPTION Agent Configuration Phone number for the person at the customer site a GE remote service engineer Contact Details would contact. The phone number is entered during installation and reviewed at every service call to make sure the information is correct.
Page 286
• No – The agent is not verified in the back office. • Not Available – The agent is not configured or running. GE part number for the LOGIQ P9/P7. The same number as listed on the rating Agent Model Number plate.
Page 287
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Chapter 6 Service Adjustments Section 6-1 Overview 6-1-1 Purpose of Chapter 6 This chapter describes how to test and adjust the mechanical capabilities of a scanner that may be out of specification.
Page 288
Section 6-3 LCD Monitor Adjustment The LOGIQ P9/P7 has a free adjustable LCD monitor in relation to the user interface. The digital control panel is located at the front of the color monitor. It is NOT recommended to change the pre-adjusted settings. However, if you are not satisfied with the factory settings, use these controls to program those you prefer in each resolution.
Page 289
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Table 6-2 Setting for LOGIQ Main Menu Sub Menu Range P9/P7 Others H position 0~100% V position 0~100% Reset Return English/German/ French/Spanish/ Language English Italian/Swedish/ Chinese/Japanese...
Page 290
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 6-3-1 Brightness/Contrast (R1.x.x) 6-3-1-1 Brightness Adjusting the monitor’s contrast and brightness is one of the most important factors for proper image quality. If these controls are set incorrectly, the Gain, TGC, Dynamic Range and even Acoustic Output may have to be chaged more often than necessary to compensate.
Page 291
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 6-3-2 OSD Control on the Utility mode (R1.x.x) Figure 6-3 Monitor 1) Go to Utility -> System -> General. 2) See Monitor category. 3) There are preset parameters below for monitor setting.
Page 292
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 6-3-3 21.5 inch Wide Monitor Adjustment (R2.x.x and later) All Monitor settings have been optimized at the factory, so normally, there should be no need for any further adjustments.
Page 293
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Display scan screen while the Utility is ac- Display Test patterns tive while the Utility is active Select Color Profile Select Gamma Select Room Profile Adjust Brightness and Constrast for Room profile “User Defined”...
Page 294
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 6-3-3-1 Scan Screen In order to view the monitor while adjusting the Contrast/Brightness and Color Profile, press the Scan Screen on the second Utility Page. 1) Press Scan Screen. The scan screen displays on the main display while the Utility touch panel is active.
Page 295
7-16 7-5-2 How the Customer enables/disables Disruptive Mode and VCO 7-16 7-5-3 GE Icon and Remote Connection 7-16 7-5-4 Customer Granting Full Remote Access Permission to GE Service Technician 7-18 7-5-5 Common Service Desktop (CSD) 7-18 7-5-6 CSD: Configuration 7-19...
Page 296
NOTE: This information is normally collected with the Alt+D or Gather Logs utility. - Product Name = LOGIQ P9/P7 1) From the touch panel, select Utility -> System-> About. Figure 7-1 About (R1, R2, R2.5 only)
Page 297
There are two possibilities to contact GE: • by means of the GE “Remote Status Icon” that is displayed on the bottom of the screen. Move the cursor to the GE icon and press the (= left-click) to display the left trackball key “Request For Service”...
Page 298
If the service platform is not configured an Error message is displayed. The request is NOT sent! The request is saved in QUEUE for reviewing and/or resending. Figure 7-5 Contact GE - Queue Section 7-2 - Collect Vital System Information...
Page 299
Request for Service (RFS) (R3 only) 7-2-3-1 Request for Service 1) Move the cursor to the GE icon and press the right-click, select “Contact GE” Figure 7-6 Contact GE 2) The Request for Service (RFS) page is displayed Figure 7-7 Contact GE – Request for Service...
Page 300
EVISION ANUARY ERVICE ANUAL 3) Click the “Sibmit” button to send the problem description to GE Service/Application representatives If Agent configuration is not configured correctly, the "Submit" button is disabled. Refer to the below. NOTICE !! NOTICE: Figure 7-8 Contact GE – Not configured correctly of the Agent configuration 4) Click the “RFS History”...
Page 301
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 7-3 Check Points Voltages 7-3-1 LED for Power Status LEDs near the front cooling fan can provide information pertaining to the DC power distribution status.
Page 302
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 7-4 Screen Captures and Logs Sometimes, the customer or field engineer wants to capture a presentation on the screen to identify the issue may be recovered by the Online Center (OLC). This is accomplished by saving the log files.
Page 303
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 7-4-3 System Warning/Errors and Logs 7-4-3-1 Temperature warning 7-4-3-1-1 Cause CPU Temperature exceeded threshold (104°C). 7-4-3-1-2 What to do 1) Shut down the system and Clean the filter.
Page 304
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 7-4-3-3 System voltage failure 7-4-3-3-1 Cause System detected unexpected voltage level from voltage level monitoring circuits. 7-4-3-3-2 What to Do 1) Press OK and reboot the system.
Page 305
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 7-4-3-4 Hardware configuration error 7-4-3-4-1 Cause Hardware(s) not detected as valid system. System will boot up but will not allow scanning. 7-4-3-4-2 What to do 1) Press OK, and shutdown the system and reboot.
Page 306
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 7-4-3-5 System Error 7-4-3-5-1 Cause Unrecoverable error occurred and unable to continue scan. Error dialog does not indicate the root cause of system error. To identify the root cause clearly, it is necessary to investigate the log files.
Page 309
LOGIQ P9/P7 scanner requires permission and customer input before a GE service technician or OLC can access the customer’s ultrasound scanner remotely. “Disruptive Mode” can be selected by the customer directly on the LOGIQ P9/P7 ultrasound system (see: Section 7-5-2 on page 7-16), or remotely by the service technician or OLC.
Page 310
7-5-2 How the Customer enables/disables Disruptive Mode and VCO 1) If not already in read mode, FREEZE the image. 2) Move the cursor to the GE icon and press the (= right-click). right trackball key 3) Select Connect Clinical Lifeline (see: Figure 7-12 on page 7-16 (R1, R2, R2.5 only) or...
Page 311
Customer Granting Full Remote Access Permission to GE Service Technician 7-5-4-1 If GE Service Technician requests Remote Access Permission If a GE Service technician requests remote access to your LOGIQ P9/P7 scanner, following “InSite Notification” appears on the systems screen. select [Yes] to enable “Disruptive Mode”...
Page 312
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 7-5-5 Common Service Desktop (CSD) Refer to Section 3-13 for accessing Common Service Desktop remotely. 7-5-6 CSD: Configuration Refer to Section 3-13 for accessing Common Service Desktop remotely.
Page 313
DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 7-6 Minimum Configuration to Boot/Scan LOGIQ P9/P7 will require at minimum MPI, MCB (with COMExpress),MDC, IO Box, HDD(Mounted on MBP) and OPIO to boot up. Chapter 7 - Diagnostics/Troubleshooting 7-19...
Page 314
IRECTION EVISION ANUARY ERVICE ANUAL Section 7-7 Troubleshooting Trees, Instructions and Tech Tips Table 7-2 below outlines LOGIQ P9/P7 troubleshooting trees and instructions shown in the sub-sections. Table 7-2 Troubleshooting Trees and Instructions Sub-section Description Page Number 7-7-1 7-21 System does not boot up...
Page 315
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 7-7-1 System does not boot up System doesn’t boot Verify that Is Standby Switch - Power Cord is connected to Mains illuminated ? - Power Switch is turned on...
Page 316
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 7-7-2 Noise disturbs the image Noise is shown in Image. Is there any electrical devices that affect the Scanner? Keep the Scanner away from those electrical devices.
Page 317
ANUARY ERVICE ANUAL 7-7-3 Trackball - Impaired Sensitivity The trackball has low sensitivity! Contact GE service representative Is it difficult to roll the to replace the trackball module. trackball? Set up the appropriate trackball response time (via [Trackball Speed] button in the System Setup - User Settings page).
Page 318
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 7-7-4 Printer Malfunction Printer malfunction Is printer Check all signal and power supply cable properly connections between Printer and Scanner connected? Load paper by using the correct format.
Page 319
How to import patient data base from older consoles Customers should back up their dataset through DICOM Export from their older scanner to import back onto the LOGIQ P9/P7. The LOGIQ P9/P7 can only understand DICOM media and cannot understand GE-specific database or private xml.
Page 320
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 7-7-8 Use ScLogWindow The ScLogWindow application helps identify errors while the system is running. Access this application from the Windows desktop or through Task Manager. From ScLogWindow, you can: •...
Page 321
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL To open ScLogWindow from Task Manager: 1.) Insert and activate the SSA key. 2.) Press Ctrl+Alt+Del and select Task Manager. 3.) In Task Manager, select File > Run New Task.
Page 322
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL To view logs in the ScLogWindow: 1.) With the ScLogWindow open, double-click the EchoLoader icon. 2.) Use Alt+Tab to put the ScLogWindow in front. 3.) While the application is booting up, observe logging in ScLogWindow.
Page 323
Expiration of an SSA key When you plug in an SSA key to the LOGIQ P9/P7 and enter the code to unlock it, a message should display on the screen acknowledging the connection. If the message does not appear, the SSA key may have expired.
Page 324
Service alert information is contained in /Log/SvcAlert*.txt. Use this log to troubleshoot issues with alerts generated by the LOGIQ P9/P7. Alerts can also be viewed on the Service desktop. Review log file content to determine the errors the LOGIQ P9/P7 has generated in the past.
Page 325
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL System configuration logs Hardware configuration information is contained in /Log/SysCfg.csv. Use this file to troubleshoot issues with hardware configuration (also known as VPD). This log helps to determine the part numbers of boards installed along with versions and serial numbers.
Page 327
Attempting do-it-yourself repairs invalidate warranty and are an infringement to regulations and are inadmissible acc. to IEC 60601-1. Energy Control and Power Lockout for LOGIQ P9/P7. NOTICE When servicing parts of the system where there is exposure to voltage greater than 30 Volts: 1.
Page 328
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Table 8-1 Chapter 8 Contents Section Description Page Number 8-10 8-71 Replacement of Side Tray & Foot Rest Cover 8-11 8-73 Replacement of Monitor and LCD Arm Plastic Covers...
Page 329
Replacement of MBOB 8-48 8-189 Replacement of MIO-B 8-49 8-192 Replacement of BBC 8-50 8-198 Replacement of LOGIQ P9 ROTATION MODULE KIT 8-51 8-204 Replacement of ODD Dummy Cover R3 8-52 8-205 Replacement of Probe Light assy Chapter 8 - Replacement Procedures...
Page 330
ANUARY ERVICE ANUAL 8-1-2 Returning/Shipping System, Probes and Repair Parts When returning or shipping the LOGIQ P9/P7 system in the original packaging: • system must be lowered to its minimum height with monitor flapped down (see: Figure on page 3-8) •...
Page 331
SEPARATELY. PLEASE CONTACT THE MANUFACTURER OR OTHER AUTHORIZED DISPOSAL COMPANY TO DECOMMISSION YOUR EQUIPMENT. Energy Control and Power Lockout for LOGIQ P9/P7. NOTICE When servicing parts of the system where there is exposure to voltage greater than 30 Volts: 1. Turn off the scanner.
Page 332
Equipment being returned must be clean and free of blood and other infectious substances. GE Healthcare policy states that body fluids must be properly removed from any part or equipment prior to shipment. GE Healthcare employees, as well as customers, are responsible for ensuring that parts/ equipment have been properly decontaminated prior to shipment.
Page 333
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-3 System Software - Installation/Upgrade Procedure 8-3-1 Introduction This section describes procedure to install/upgrade software to system. 8-3-2 Manpower 1 person, 1 hour - Depends on contents of System DVD, peripherals, etc. (R1.x.x or R2.x.x) - Depends on contents of USB media, peripherals, etc.
Page 334
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL • The Hard Disk Drive (HDD) on LOGIQ P7/P9 is divided into four visible (R1.x.x.) or five (R2.x.x or later) visible partitions. Table 8-2 Use of the Partitions (R1.x.x)
Page 335
Peripherals (see: “System - Recording Peripheral settings”). About (see: “System - Recording software and hardware versions”). General, Record preset Region, Language and Regional Options. (See the latest revision ofthe LOGIQ P9/P7 Basic User Manual)) 8-3-5-2-1-2 Connectivity - Recording the TCP/IP settings 1) From the touch panel, select Utility ->...
Page 336
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Figure 8-1 Connectivity - Recording the TCP/IP settings (R1.x.x) Figure 8-2 Connectivity - Recording the TCP/IP settings (R2.x.x and later) 2) Record all settings in Table 8-4.
Page 337
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 2) Select the Properties Tab. 3) Record all settings in Table 8-5. Table 8-5 Record settings from WLAN screen PARAMETER VALUE NETWORK NAME (SSID) NETWORK AUTHENTICATION...
Page 338
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-3-5-2-1-4 Connectivity - Recording the AE Title and Port settings 1) From the touch panel, select Utility -> Connectivity -> Device. 2) Select My Computer in the List of Devices, if not already selected.
Page 339
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 4) Record all settings in Table 8-7. Table 8-7 Record settings for each device listed in Device table Name IP Address 8-3-5-2-1-5 Connectivity - Recording Device Destination settings 1) From the touch panel, select Utility ->...
Page 340
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-3-5-2-1-6 Connectivity - Recording Dataflow settings 1) From the touch panel, select Utility -> Connectivity -> Dataflow. Figure 8-6 Connectivity - Recording Dataflow settings 2) Record all settings in Table 8-9 for each Destination Device and Service.
Page 341
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-3-5-2-1-7 Connectivity - Recording the Print Key Assignments 1) From the touch panel, select Utility -> Connectivity -> Buttons. Figure 8-7 Connectivity - Recording the Print Key Assignments...
Page 342
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-3-5-2-1-8 Connectivity - Recording other settings for Print 1) From the Touch Panel, select Utility -> Connectivity -> Buttons. Figure 8-8 Button settings 2) Record all settings in...
Page 343
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Figure 8-9 Connectivity - Service settings 2) Record all settings in Table 8-13. Table 8-13 Record settings from Service screen Destination Services (Types) Properties Name Device Name...
Page 344
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Figure 8-10 Connectivity - Recording Miscellaneous settings 2) Record all settings in Table 8-14. Table 8-14 Record settings from Miscellaneous screen Patient/Exam Menu Options ON/OFF or Value...
Page 345
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Table 8-14 Record settings from Miscellaneous screen Warn video titles exist in the internal storage Columns in examination listings ON/OFF or Value (Date | Category | Exam Description | Img. Size | M&A | Disk...
Page 346
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Make sure that the Software Option Keys (alphanumeric passwords) have been WARNING WARNING recorded correctly. If the key is incorrect, you will not be able to log on after the !! WARNING: software installation has been completed.
Page 347
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 2) Record all settings in Table 8-16. Table 8-16 Record settings from Users screen Group User ID Prefix Last Name First Name Op Rights Membership 8-3-5-2-1-13 System - Data Store Management 1) From the touch panel, select Utility ->...
Page 348
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-3-5-2-1-14 System - Recording Peripheral settings 1) From the touch panel, select Utility -> System-> Peripherals. Figure 8-14 System - Recording Peripheral settings (R1.x.x.) Figure 8-15 System - Recording Peripheral settings (R2.x.x. and later)
Page 349
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Table 8-19 Standard printer and Default printer (R2.x.x. and later) Parameter Value Standard Printer Default Printer Table 8-20 DVR (R2.x.x. and later) Parameter Value Media ____ DVR ____ USB...
Page 350
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-3-5-2-1-16 Location, Region and Language 1) From the touch panel, select Utility -> System-> General. Figure 8-17 Location, Region and Language (R1.x.x) Figure 8-18 Location, Region and Language (R2.x.x and later)
Page 351
ANUARY ERVICE ANUAL 8-3-6 System Software - Installation Procedure Two separate methods of software installation is offered in LOGIQ P9/P7. A) Software Load from DVD(R1/R2) or USB media (R3) B) Software Re-Load from Repository Situation • Harddrive is new • Re-installation for application or operating system issues •...
Page 352
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-3-7 Software Installation from DVD 8-3-7-1 System Software - Installation Procedure (R1.x.x) 1) Prepare the Base System Software DVD. 2) Power ON the scanner and press “DEL”(If AN key is installed) or press left bottom key of trackball (If there’s no AN Key) to enter BIOS...
Page 353
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 4) Go to “Boot” in BIOS menu and change the setting value from “SATA 1 Drive” to “SATA 2 Drive” in ‘1st Boot Device’, change the setting value from “SATA 2 Drive” to “SATA 1 Drive” in ‘2nd Boot Device’...
Page 354
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 9) The notification screen appears, and then hit the Enter key. NOTE: Use simulator A/N keyboard if customer’s system does not have external A/N keyboard 10) Select either [1] “Load the complete disk”, or...
Page 355
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 11) A confirmation window appears, and then hit the Enter key. 12) Approx. 40 minutes later, loading the software will be completed. 13) Remove Base System Software DVD, and type "exit" to reboot the system.
Page 356
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 17) Go to “Save & Exit” in BIOS menu and select “Restore Defaults”. 18) Go to “Boot” in BIOS menu and check that ‘1st Boot Device’ is set to “SATA 1” and ‘2nd Boot Device’...
Page 357
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 19) Go to “Save & Exit” in BIOS menu and select “Save Changes and Exit”. Then, the system will re- boot. Chapter 8 - Replacement Procedures...
Page 358
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-3-7-2 System Software - Installation Procedure (R2.x.x) 1) Prepare the Base System Software DVD. 2) Power ON the scanner and press “DEL”(If AN key is installed) or press left bottom key of trackball (If there’s no AN Key) to enter BIOS...
Page 359
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 4) Go to “Boot” in BIOS menu and change the setting value from “SATA 1 Drive” to “SATA 2 Drive” in ‘1st Boot Device’, change the setting value from “SATA 2 Drive” to “SATA 1 Drive” in ‘2nd Boot Device’.
Page 360
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 9) The notification screen appears, and then hit the Enter key. NOTE: Use the trackball and left set key if the customer’s system does not have an external A/N keyboard.
Page 361
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 11) A confirmation window appears, and then hit the Enter key. 12) Approx. 30~40 minutes later, loading the software will be completed. 13) Remove Base System Software DVD, and hit the Enter key to reboot the system.
Page 362
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 17) Go to “Save & Exit” in BIOS menu and select “Restore Defaults”. 18) Go to “Boot” in BIOS menu and check that ‘1st Boot Device’ is set to “SATA 1” and ‘2nd Boot Device’...
Page 363
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 19) Go to “Save & Exit” in BIOS menu and select “Save Changes and Exit”. Then, the system will re- boot. Chapter 8 - Replacement Procedures...
Page 364
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-3-7-3 System Software - Installation Procedure (R3.x.x) 8-3-7-3-1 Set the system boot order (BIOS setting) 1.) Prepare the THUMB DRIVER R3.x.x (USB media) including Base Image Load Windows 10.
Page 365
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 6.) In “Boot” page, Change “Boot Option #1” to USB memory which is inserted. NOTE: The name of USB memory can be displayed different depend on the inserted USB memory.
Page 366
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-3-7-3-2 Base and Application Software installation • Base Image Software Load Procedure 1.) Power ON the scanner. 2.) Wait few moments for USB drive to read USB contents.
Page 367
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 6.) The Installation progress screen appears like below. Figure 8-25 The notification screen 3 Do not interrupt the software loading at any time. WARNING WARNING 7.) Approx.20~ 30 minutes later, loading the software will be completed.
Page 368
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-3-7-3-3 Application Software Load Procedure 1.) After the reboot again, the application software installation will start automatically. 2.) About few minutes, the following options will appear. Select your product. Click “Start Installation”...
Page 369
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-3-7-4 Application Software Installation Preparation (R1.x.x or R2.x.x) 1) Verify the system boot up to Windows Screen 2) Prepare the Application Software DVD. 3) Press DVD-Drive eject button, and place the Application Software DVD on the tray.
Page 370
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-3-7-5 Application Software Installation (R1.x.x or R2.x.x) 1) Application Software Installation process will take approximately 20 minutes 2) Select your product. Click “Start Installation” button. 3) After the installation completes, the system will be rebooted automatically.
Page 371
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 5) If SW License window appears, then enter correct Option String and click “OK”. Chapter 8 - Replacement Procedures 8-45...
Page 372
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-3-7-6 Software DVD / USB Storage Remove the rear cover and store the software DVD / USB in the system. 1) Remove the rear cover. Figure 8-29 Removing the rear cover 2) Store the Software DVD in the pocket.
Page 373
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 3) Store the Software USB STICK in the pocket as below picture. (R3 only) USB STICK POCKET Figure 8-31 USB stick pocket Chapter 8 - Replacement Procedures...
Page 374
8-3-8 Software Installation from Common Service Desktop (R1.x.x or R2.x.x) NOTE: This procedure requires PC with LAN connection capable of reaching to to GE Network 8-3-8-1 Involking Software Reload 1) Select [SERVICE] from [UTILITY] Menu (R1.x.x) or Select [SERVICE] from [UTILITY] Menu (R2.x.x)
Page 375
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 3) Verify entering to Common Service Desktop (CSD) 4) Select “Disruptive Mode” under Common Utility, in [Utilities] Tab 5) Acknowledge “Disruptive Mode” on Console Screen 6) Verify Disruptive Mode...
Page 376
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8) Click Setup software reload. 9) Select [SYSTEM SHUTDOWN] from Common Utility Menu in CSD 10) Select Restart System Retain Disruptive Mode 11) Press [SUBMIT] 12) On console side, wait until application software loading completes.
Page 377
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-3-9 Software Installation from Common Service Desktop (R3.x.x) For detail information, please refer to 8-3-9 Software Reload on page 5-81 Chapter 8 - Replacement Procedures 8-51...
Page 378
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-3-10 System Setup Procedure to display the messages in Japanese/Chinese To setup the system to display the messages in Japanese/ Chinese: NOTE: You cannot enter text in Japanese/Chinese.
Page 379
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 3) Under "Current system locale:", select Japanese (Japan) or Chinese (Simplified PRC). Press OK. Figure 8-35 Set Language - example 4) The "Change System Locale" pop-up prompt appears. Select Cancel to not reboot the system yet.
Page 380
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 7) Open the Administrative tab and check Japanese or Chinese displays under Current language for non-Unicode programs. Figure 8-37 Administrative tab - example 8-54 Section 8-3 - System Software - Installation/Upgrade Procedure...
Page 381
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8) Under "Format:", select "Japanese (Japan)"/"Chinese (Simplified PRC)". Press Apply, Press OK. Figure 8-38 Formats 9) In Utility--> System--> General, set the Language as Japanese or Chinese. Press Save and reboot the system using Power on/off button.
Page 382
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-4 Software and Functional Checks after Installation Procedure 8-4-1 Functional Check Table 8-23 Item Test Procedure Reference Note Power ON 4-3-3 Power On/Off [___] Passed [___] Failed...
Page 383
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-5 Settings - Backup and Restore This section briefly describes Image Setting backup and restore function. For detailed documentations and warnings, refer to Chapter.16, Section “System/Backup and Restore Preset Menu”...
Page 384
For the sake of simplicity, we have used the CD-R in the examples in this section. 8-5-1-2 General Notes NOTE: When you restore backup data from the Utility menu, the LOGIQ P9/P7 application usually restarts automatically when the restoring is complete. NOTE: To perform backup and restore procedures, you must login with administrator privileges.
Page 385
6) Select Restore. The system performs the restore. As it proceeds, status information is displayed on the Backup/ Restore screen. 7) The LOGIQ P9/P7 restarts automatically when Restore is done. 8-5-4 EZBackup and EZMove EZBackup or move do not backup patients without images, or reports. If the entire...
Page 386
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL BEFORE starting the EZBackup, select Unlock All in Utility -> Admin -> Logon. 2) Specify the EZBackup/EZMove setup on the Utility -> System -> Backup/Restore page.
Page 387
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-5-5 Option Keys 1) From the touch panel, select Utility -> System-> Backup/Restore. 2) ADD or REMOVE new option key at “Enter New Option Key” field.
Page 388
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-6 Replacement of Front Cover 8-6-1 Manpower 1 person, 1 min. 8-6-2 Tools None required. 8-6-3 Removal Procedure 1) Simply pull out Front Cover. Front Cover Figure 8-41 Front Cover 2) Front cover is attached using four “latches”...
Page 389
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 3) Pull out the Air filter from Front Cover. AIR FILTER FRONT Figure 8-43 Removal of Air filter 8-6-4 Installation Procedure 1) Front cover and Air filter to be installed in reverse order of removal.
Page 390
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-7 Replacement of Top Cover 8-7-1 Manpower 1 person, 5 min. 8-7-2 Tools None required. 8-7-3 Removal Procedure 1) Top Cover shapes differ for each configuration, but removal/installation procedure are the same.
Page 391
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-7-4 Installation Procedure 1) Cover to be installed in reverse order of removal. 2) Make sure “latches” on both sides are engaged. NOTE: For Mid and High Cabinet type, there is “Guide Rail” for installation.
Page 392
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-8 Replacement of Side Cover 8-8-1 Manpower 1 person, 15 min. 8-8-2 Tools Standard Phillips Screwdriver. 8-8-3 Removal Procedure 8-8-3-1 Removal Procedure - Side Cover (L) 1) Remove Top Cover.
Page 393
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 3) Pull out Side Cover (L). There is mechanical hook on bottom of the part. Figure 8-50 Pull out Side Cover(L) Chapter 8 - Replacement Procedures...
Page 394
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-8-3-2 Removal Procedure - Side Cover (R) 1) Remove Top Cover. Figure 8-51 Removal of Top cover 2) Remove knobs Figure 8-52 Removing RS probe knobs...
Page 395
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 3) Remove 3 screws for Side Cover (R). Figure 8-53 Side Cover (R) screw. 4) Pull out Side Cover(R). There is mechanical hook on bottom of the part.
Page 396
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-9 Replacement of Rear Cover 8-9-1 Manpower 1 person, 1 min. 8-9-2 Tools No tool. 8-9-3 Removal Procedure 1) Pull the rear cover and remove it as following figure Figure 8-55 Removal of Rear Cover.
Page 397
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-10 Replacement of Side Tray & Foot Rest Cover 8-10-1 Manpower 1 person, 15 min. 8-10-2 Tools Standard Phillips Screwdriver. NOTE: “Short” type driver preferable for Foot Rest Cover removal/installation.
Page 398
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-10-3-2 Removal Procedure - Foot Rest Cover 1) Remove Front Cover (refer to section 8-6) 2) Remove Top Cover (refer to section 8-7) 3) Remove Side Cover (refer to section 8-8) 4) Remove 3 screws.
Page 399
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-11 Replacement of Monitor and LCD Arm Plastic Covers Be care that the monitor does not fall during the replacement process. WARNING WARNING 8-11-1 Manpower 2 person, 15 minutes.
Page 400
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 2) Unscrew 5 screws, then remove cable clamp. NOTE: There is cable position marker at red circle. Caution - This Clamp was only applied by opposite direction Figure 8-61 Unscrew cable clamp screws 3) Disconnect three Monitor cables.
Page 401
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 4) Unscrew 2 screws, then separate the Monitor from the Monitor Stand Arm. Refer to the figure below. Figure 8-63 Unscrewing 2 screws & separating the Monitor...
Page 402
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-11-6 LCD Arm Plastic Covers Remove Procedure 1) Unscrew 1 screw, than remove 3rd Arm Cover. Refer to the figure below. Figure 8-64 Removing the 3rd Arm Cover 2) Remove 2nd Axis Cover.
Page 403
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 4) Unscrew 1 screw and Remove the 1st Arm Cover. Refer to the figure below. Figure 8-67 Removing the 1st Arm Cover 8-11-7 LCD Arm Plastic Covers Installation Procedure Parts to be installed in reverse order of removal.
Page 404
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-12 Replacement of Monitor Covers 8-12-1 Manpower 2 persons, 10 minutes. 8-12-2 Tools Standard Phillips Screwdriver, Hex Wrench 8-12-3 Pre-Work 1) LCD Monitor and Monitor Cable Cover removed (refer to section 8-11)
Page 405
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-12-5 Installation Procedure 1) Parts to be installed in reverse order of removal. Make sure not to remain any dust on Glass during Front Cover and Glass Filter Assy...
Page 406
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-13 Replacement of Fixed Arm Cover and Fixed Arm Holder Grip 8-13-1 Manpower 2 person, 5 minutes. 8-13-2 Tools Standard Phillips Screwdriver, Hex Wrench. 8-13-3...
Page 407
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-13-4 Installation Procedure Parts to be installed in reverse order of removal. 8-13-5 Functional Check Visual Check only. Chapter 8 - Replacement Procedures 8-81...
Page 408
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-14 Replacement of OPIO and Related Parts NOTE: This section describes removal and installation of OPIO and related parts. OPIO Assy itself, Main PWA Assy, LCD Touch module Assy, HUB Board, Encoder, Freeze key Assy, OPIO AN Key &...
Page 409
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-14-3-1 Removal Procedure - Trackball Ring NOTE: Removal of Trackball Ring and Cleaning does not require tools. 1) Turn the ring counter clock-wise 2) Lift trackball from OP Panel...
Page 410
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 2) Remove 2 screws on OPIO covers. Figure 8-76 Rear of OPIO Base 3) Remove USB cables on Touch panel and unscrew 2 screws. Refer to following figure.
Page 411
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 4) Lift OP Panel. Figure 8-78 Lift OP Panel 5) Remove USB locking type Cable 6) Remove OPIO Power cable. 7) Remove USB locking type connector.
Page 412
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-14-3-3 Removal Procedure - AN Keyboard (Option) 1) Remove 4 screws on AN keyboard Brkt. ( 2 screws on right side and 2 screws on left side) Figure 8-80 Removal of screws 2) Unscrew 2 screws on AN keyboard Brkt only half .
Page 413
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 4) Disconnect USB cables as following figure. Figure 8-83 Disconnecting USB cable 8-14-3-4 Removal Procedure - Trackball 1) Remove a connector from back of the LCD Touch Panel Assy.
Page 414
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-14-3-5 Removal Procedure - LCD Touch Panel Assy 1) Disconnect connectors as following figure. Figure 8-85 Disconnection of LCD Touch panel Assy 2) Remove 4 screws as following figure.
Page 415
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-14-3-6 Removal Procedure - Freeze key assy 1) Remove connector as following figure Figure 8-87 Disconnection of freeze key Assy connector 2) Remove 3 screws from Freeze key Assy Figure 8-88 Screw points of freeze key assy .
Page 416
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-14-3-7 Removal Procedure - OP Panel Main PWA Assy 1) Carefully lift up knobs as following figures 2) Place the new knobs. Figure 8-89 Position of knobs...
Page 417
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 3) Disconnect freeze key assy connector, trackball connector and Touch panel connectors. Refer to following figure. Touch panel Freeze key connector connector Trackball connector Figure 8-91 Removal of connectors on Main PWA assy...
Page 418
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 4) Remove 23 screws around PWA. Refer to following figure. Figure 8-92 Screw points on Main PWA Assy 8-92 Section 8-14 - Replacement of OPIO and Related Parts...
Page 419
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-14-3-8 Removal Procedure - Encoder on PWA 1) Remove OP Panel Main PWA Assy. 2) Loosen nut. 3) Remove part Figure 8-93 Removal of Encoder 8-14-3-9 Removal Procedure - Touch panel 1) Remove OP Panel Assy.
Page 420
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 4) Remove 10 screws on BRKT and remove 1 BRKT Refer to following figure. Figure 8-95 Screw points to remove 3 BRKT 5) Remove flexible connectors and carefully lift up Touch panel.
Page 421
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL NOTE: Make sure to connect Touch panel connector to hub board as blow figure. Figure 8-98 Installation of Touch panel connector Make sure not to remain any dust on Glass during OPIO Upper Cover and Touch Screen Assy...
Page 422
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-15 Replacement of the Caps for Hardkeys 8-15-1 Manpower 1 person, approx. 1 minute/cap. 8-15-2 Tools Small-sized slotted screwdriver or tweezers. 8-15-3 Removal Procedure 1) By means of a small slotted screwdriver, carefully push against the hardkey cap.
Page 423
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-15-5 Functional Check Test Refer to Note System Exterior Visual Check Control panel visual check only section 4-3-1 Power On/Off section 4-3-3 System Integration Checks section 4-3-5-1 OPIO Test only...
Page 424
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-16 Replacement of Key Caps (by special native language keys) 8-16-1 Manpower 1 person, 30 minutes. 8-16-2 Tools • Regular Standard Phillips Screwdriver • Screwdriver (Diameter : 3mm)
Page 425
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 3) Carefully pull out key caps. Refer to below Figure 8-77 Figure 8-103 Detaching key caps. 8-16-5 Installation Procedure 1) Parts to be installed in reverse order of removal.
Page 426
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 4) Run the OPIO Test tool (ScFpApiTest.exe) in the desktop screen and then click the “Op-panel Open” button of test tool. (Program path: C:\scanner\target\bin\ScFpApiTest.exe) Figure 8-104 5) Check on the “Keyboard Test on”...
Page 427
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-17 Replacement around Nest Box NOTE: This section covers replacement of components related to Nest Box. : MPI,MCB,MDC,MIO,CWD,DCWD,EEPROM,DC4D,Battery,HDD and SOM 8-17-1 Manpower 1 person, 15 minutes for each part.
Page 429
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-17-5 Removal Procedure - Nest Box NOTE: Before working with Nest box, be sure to remove probes. 1) Remove front cover. (refer to section 8-6) 2) Disconnect 1 fan connector of bottom fan.
Page 431
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-17-7 Removal Procedure - Board Assy in Nest Box 1) Get Next Box access. 2) Remove 7 screws on EMI bracket. Figure 8-113 Screw points on Nest Box...
Page 432
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 3) Draw MPI, MCB and IO Box out of the Nest Frame. Figure 8-114 Separating Board Assy MPI board assy MCB board assy IO BOX assy...
Page 433
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-17-8 Removal Procedure - Separating MPI and MCB 1) Unscrew 6 screws. Refer to following figure. Figure 8-116 Screw points on MPI assy 2) Carefully separate MPI assy from MCB assy. Refer to following figure.
Page 434
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-17-9 Removal Procedure - Separating DCWD Board(Option) from MPI Philips screw-driver required to add this part. 1) Remove 3 screws on MPI assy. Figure 8-118 DCWD Board screw points 2) Carefully separate DCWD board from MPI assy not to damage the DCWD connector.
Page 435
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-17-10 Separating MDC assy from MCB assy 1) Remove 7 screws around MDC assy. Refer to following figure. Figure 8-120 Screw points around MDC assy 2) Carefully separate MDC assy from MCB assy. Refer to following figure.
Page 436
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-17-11 Handling EEPROM on MCB Board EEPROM on MCB board contains vital information. A) System Serial Number that is matched against stored software option key. B) ComExpress Functional REV that is monitored for sub-system compatibility.
Page 437
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 3) Remove EEPROM from replacement MCB board if present. 4) Verify direction of the EEPROM Slot by aligning EEPROM “notch” mark to that of MCB board. print mark.
Page 438
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 2) Remove MCWD board from MCB assy. Refer to following figure. NOTE: Be careful not to damage CWD connector. MCWD board assy Figure 8-126 Removal of MCWD board.
Page 439
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-17-13 Separating DC4D Unit from IO BOX 1) Unscrew 12 screws to remove the IO BOX Cover. Figure 8-128 Screw points of IO BOX cover 2) Unscrew five screws to remove the DC4D from the MIO Assy. Refer to following figure.
Page 440
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-17-14 Separating Battery Unit from IO BOX 1) Unscrew 12 screws to remove the MIO Cover. (refer to section 8-17-13) 2) Unscrew 8 screws and 2 Battery connectors. Refer to following figure.
Page 441
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-17-15 Separating HDD 1) Open the front cover. (refer to section 8-6) 2) Remove 4 screws around HDD Bracket Refer to following figure. Figure 8-131 Screw points around HDD Bracket.
Page 442
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-17-16 Separating COM Express 1) Unscrew 5 screws on bottom side of MCB board assy. Refer to following figure. Figure 8-133 Screw points for removal of COM Express.
Page 443
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-17-17 Installation Procedure - MPI, MCB, MDC, DC4D, HDD, Battery and COM Express 1) Parts to be installed in reverse order of removal. 8-17-18 Functional Check - MPI, MCB, MDC, IO Box...
Page 444
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-18 Replacement of Front fan 8-18-1 Manpower 1 Person, 10 minutes. 8-18-2 Tools Standard Phillips Screwdriver. 8-18-3 Removal Procedure 1) Remove the Front Cover from system. (refer to section 8-6) 2) Remove 4 screws and disconnect fan connector.
Page 445
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-19 Replacement of Bottom fan 8-19-1 Manpower 1 Person, 10 minutes. 8-19-2 Tools Hex Wrench. 8-19-3 Removal Procedure 1) Remove the Front Cover from system. (refer to section 8-6) 2) Remove the nest Box from system.
Page 446
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 4) Place bottom fan cable not to be entangled on system frame as following figure Figure 8-140 Placing bottom fan connector 5) Pull out bottom fan assy as following figure...
Page 447
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 6) Remove 4 screws on the fan. Figure 8-142 Screw points of bottom fan. 8-19-4 Installation Procedure Parts to be installed in reverse order of removal.
Page 448
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-20 Replacement of Rear Handle 8-20-1 Manpower 1 Person, 10 minutes. 8-20-2 Tools Standard Phillips Screwdriver and 6mm Hex Wrench. 8-20-3 Removal Procedure 1) Remove 1 screw and remove the Link Arm Top Cover as following figure.
Page 449
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-21 Replacement of Link Arm Cable Hook 8-21-1 Manpower 1 Person, 10 minutes. 8-21-2 Tools Standard Phillips Screwdriver. 8-21-3 Removal Procedure 1) Remove 1 screw and remove the Link Arm Top Cover as following figure.
Page 450
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-22 Replacement of Caster 8-22-1 Manpower 2 persons, 15 minutes. 8-22-2 Tools 24mm Socket or adjustable Wrench, Standard Phillips Screwdriver NOTE: “Short” type driver preferable for Foot Rest Cover removal/installation.
Page 451
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL NOTE: This does not require any tools. Be careful not to damage Hand during pulling hook of the cap Figure 8-148 Caster Cap and hook points 4) Remove Caster Cap by slightly pulling hook of the cap to outside and then move upward.
Page 452
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 6) Remove Nut and replace Caster. Figure 8-150 Removable Nut of Caster Proceed with extreme caution and avoid system tipping over. The system weighs CAUTION approximately 60kg.
Page 453
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-23 Replacement of the COM Express 8-23-1 Manpower 1 Person, 5 minutes. 8-23-2 Tools Standard Phillips Screwdriver. 8-23-3 Removal Procedure 1) Separate MCB Assy and MDC Assy. (Refer to section 8-17-10) 2) Remove 8 screws around MCB assy and carefully separate MCB assy from MPI assy.
Page 454
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 3) Unscrew 5 screws on bottom side of MCB board assy. Refer to following figure. Figure 8-152 Screw points for removal of COM Express. 4) Disconnect fan cable and carefully detach COM Express module. Refer to following figure.
Page 455
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-23-4 Installation Procedure Install components in reverse order of removal. When replacing with new SOM, Push the SOM in both side of MCB Assy like as below Figure CAUTION 8-130.
Page 456
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Right bottom key of trackball Left bottom key of trackball to for ‘ESC’ in BIOS enter BIOS and select menu and set or change the value in BIOS 2) Type “maru”...
Page 457
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL NOTE: There are various types of VPD EEPROM integrated in the system. Board Name Main VPD Contents Replaceable Physical VPD location FuncRev Integrated with MCB FuncRev...
Page 458
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Device Grid Data Grid Device Grid The list of devices in the system that support VPD. Click on a row in this grid to select a device and view the VPD fields.
Page 459
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 1) Select “COMEXP” from device grid. (right pane). COMEXP Load Template 2) Select “Load Template” from right pane. 3) Select “Comexp4amaster.vpd”. The file can be found under c:\Scanner\target\resources\Acqusition\Housekeeping 4) Click “Open”...
Page 460
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL NOTE: This procedure can be applied to other components with VPD, for example, for initializing MPI VPD. However, it is unlikely needed for such event arises, as these components are hardware-integrated into boards.
Page 461
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-24 Replacement of the HDD/SSD 8-24-1 Manpower 1 Person, 5 minutes. 8-24-2 Tools Standard Phillips Screwdriver. 8-24-3 Removal Procedure 1) Open the front cover. (refer to section 8-6) 2) Remove 4 screws around HDD/SSD Bracket Refer to following figure.
Page 462
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 4) Unscrew 4 screws to remove HDD/SSD bracket. Refer to following figure. Figure 8-159 Removal of HDD/SSD bracket 8-24-4 Installation Procedure Install components in reverse order of removal.
Page 463
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-25 Replacement of Battery 8-25-1 Manpower 1 person, 1 hour NOTE: LP9/P7 BUM 5504583 provides the information, "Refreshing the battery" in Chapter 3 Preparing to System for Use.
Page 464
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 2) Unscrew 4 screws and disconnect 2 connectors from Battery. Refer to following figure. Figure 8-161 Screw points and Connectors on battery assy 3) Unscrew 4 screws to remove Battery bracket...
Page 465
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-26 Replacement of 4D 8-26-1 Manpower 1 Person, 1 hour. 8-26-2 Tools Standard Phillips Screwdriver. 8-26-3 Removal Procedure 1) Unscrew 12 screws to remove the IO BOX Cover.
Page 466
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 2) Unscrew five screws to remove the DC4D from the IO BOX. Refer to following figure. Figure 8-164 Screw points of DC4D Assy 3) Carefully lift up the DC4D unit.
Page 467
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-27 Replacement of Wireless LAN 8-27-1 Manpower 1 Person, 5 minutes. 8-27-2 Tools None Required. 8-27-3 Removal Procedure 1) Remove Top Cover (refer to section 8-7) 2) Attach loop sheet on Wireless LAN dongle.
Page 468
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-28 Replacement of the Harness Cable and OPIO Cable Assy 8-28-1 Manpower 1 person, 30 minutes 8-28-2 Tools Standard Phillips Screwdriver, Nipper, 4mm Hex Wrench, 3 Cable ties (GEPN: 2160955).
Page 470
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Figure 8-167 Removal of Harness Cable Assy 8-28-5 Removal Procedure (OPIO Cable Assy) 1) Unscrew 1 screw, remove cable clamp, and then loosen cable through hole.
Page 471
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-28-7 Functional Check Test Refer to Note System Exterior Visual Check Control panel visual check only section 4-3-1 Power On/Off section 4-3-3 Visual check on LCD monitor to see if image displays without color distortion Check OSD Control function operates.
Page 472
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-29 Replacement of Peripheral Cable Assy 8-29-1 DVD Cables 8-29-1-1 Manpower 1 person, 5 minutes. 8-29-1-2 Tools Nipper, 2 Cable ties (GEPN: 2160955). 8-29-1-3 Pre work 1) Remove Rear Cover.
Page 473
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 3) Disconnect 2 DVD cables from the rear panel. Refer to following figure. Figure 8-171 DVD cable connectors 4) Disconnect 2 DVD cables from the DVD.
Page 474
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-29-1-5 Installation Procedure Parts to be installed in reverse order of removal. But, refer to below procedure for cable ties. 2 holes for fixing cable tie...
Page 475
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-29-2 B/W Printer Cables 8-29-2-1 Manpower 1 person, 5 minutes. 8-29-2-2 Tools Standard Phillips Screwdriver. 8-29-2-3 Pre work 1) Remove Rear Cover. (refer to section 8-9) 2) Remove Top Cover (refer to section 8-7) 3) Remove Cable Cover Bracket.
Page 476
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 3) Unlock cable clamp. Refer to following figure. Figure 8-176 Cable clamp 8-150 Section 8-29 - Replacement of Peripheral Cable Assy...
Page 477
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 4) Disconnect 2 BW printer cables from the rear panel. Refer to following figure. USB signal connector AC Power connector Figure 8-177 BW printer connectors 5) Disconnect 2 BW printer cables from printer and carefully pull out the cables.
Page 478
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-29-2-5 Installation Procedure Parts to be installed in reverse order of removal. But, refer to below procedure for cable ties. 2 holes for fixing cable tie...
Page 479
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-30 Replacement of the Probe Holder 8-30-1 Manpower 1 Person, 5 minutes. 8-30-2 Tools No tool is required. 8-30-3 Removal Procedure 1) Simply pull out Probe Holder from its location.
Page 480
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-31 Replacement of the Cable Hook 8-31-1 Manpower 1 Person, 5 minutes. 8-31-2 Tools Standard Phillips Screwdriver. 8-31-3 Removal Procedure 1) Remove screws as following figure and carefully remove cable hook.
Page 481
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-32 Replacement of the OPIO Tray 8-32-1 Manpower 1 Person, 5 minutes. 8-32-2 Tools Standard Phillips Screwdriver. 8-32-3 Removal Procedure 1) Remove screw as following figure and carefully remove OPIO Tray.
Page 482
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-33 Replacement of the Paper tray 8-33-1 Manpower 1 Person, 1 minutes. 8-33-2 Tools None Required. 8-33-3 Removal Procedure 1) Remove Paper tray from Top Cover.
Page 483
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-34 Replacement of AC Cable Holder 8-34-1 Manpower 1 Person, 5 minutes. 8-34-2 Tools Standard Phillips Screwdriver. 8-34-3 Removal Procedure 1) Remove 2 Screws from AC Cable Holder...
Page 484
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-35 Replacement of the BW printer 8-35-1 Manpower 1 person, 10 minutes 8-35-2 Tools Standard Phillips Screwdriver. 8-35-3 Pre work 1) Remove Rear Cover. (refer to section 8-9) 2) Remove Top Cover (refer to section 8-7) 3) Remove Cable Cover Bracket.
Page 485
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 2) Pull out BW printer as following figure. Figure 8-187 Pulling out the BW printer 3) Remove 4 screws on BW printer bottom part. Refer to following figure.
Page 486
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-36 Replacement of the Speaker 8-36-1 Manpower 1 Person, 10 minutes. 8-36-2 Tools Standard Phillips Screwdriver, Hex Wrench, Nipper, 1 Cable ties (GEPN: 2160955). 8-36-3 Pre work 1) Remove Rear Cover.
Page 487
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 2) Remove the cable tie. Figure 8-190 Cable tie point 3) Unlock cable clamp. Refer to following figure. Figure 8-191 Cable clamp 4) Remove 4 screws around speaker assy. Refer to following figure.
Page 488
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 5) Remove 4 screws, and then remove speaker brkt. Refer to following figure. Figure 8-193 4 screw points to remove speaker brkt 6) Remove 8 screws and carefully pull out speaker assy. Refer to following figure.
Page 489
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-36-5 Installation Parts to be installed in reverse order of removal. But, refer to below procedure for cable ties. 2 holes for fixing cable tie Cable tie...
Page 490
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-37 Replacement of the Monitor Arm 8-37-1 Manpower 1 person, 15 minutes. 8-37-2 Tools Standard Phillips Screwdriver, Nipper, 4 mm Hex Wrench, 3 Cable ties (GEPN: 2160955).
Page 491
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-38 Replacement of Front Fan Filter 8-38-1 Manpower 1 Person, 2 minutes. 8-38-2 Tools None Required. 8-38-3 Removal Procedure 1) Remove Front Cover. (refer to section 8-6) 2) Pull out Filter upward.
Page 492
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-39 Replacement of Bottom Fan Filter 8-39-1 Manpower 1 Person, 3 minutes. 8-39-2 Tools None Required. 8-39-3 Removal Procedure 1) Pull out bottom fan filter as following figure.
Page 493
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-40 Replacement of SW CD set (R1, R2, R2.5 only) and SW USB (R3 only) 8-40-1 Manpower 1 Person, 3 minutes. 8-40-2 Tools None Required.
Page 494
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-41 Replacement of Drawer 8-41-1 Manpower 1 Person, 5 minutes. 8-41-2 Tools Standard Phillips screwdriver. 8-41-3 Removal Procedure 1) Pull out the Top cover. Figure 8-198 Drawing out Top cover.
Page 495
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-41-4 Installation Procedure Parts to be installed in reverse order on removal. 8-41-5 Functional Check Visual Check Only Chapter 8 - Replacement Procedures 8-169...
Page 496
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-42 Replacement of ECG 8-42-1 Manpower 1 Person, 5 minutes. 8-42-2 Tools Standard Phillips screwdriver. 8-42-3 Removal Procedure 1) Remove OPIO covers (refer to section 8-14) 2) Remove cables on ECG Assy (square) and unscrew 4 screws (circle).
Page 497
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 5) Unscrew 4 screws and remove ECG BRKT Figure 8-202 ECG screw points. 8-42-4 Installation Procedure Parts to be installed in reverse order on removal. 8-42-5...
Page 498
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-43 Replacement of DCWD ASSY 8-43-1 Manpower 1 person, 15 minutes 8-43-2 Tools Standard Phillips screwdriver. 8-43-3 Removal Procedure 1) Separate MPI assy from MCB and MDC. Refer to the section 8-17-8 Removal Procedure - Separating MPI and MCB.
Page 499
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-43-4 Installation Procedure Parts to be installed in reverse order on removal. 8-43-5 Functional Check Refer to section 4-4 Functional Diag Test List Chapter 8 - Replacement Procedures...
Page 500
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-44 Replacement of MCWD ASSY 8-44-1 Manpower 1 person, 15 minutes 8-44-2 Tools Standard Phillips Screwdriver. 8-44-3 Removal Procedure 1) Separate MCB Assy and MDC Assy. (Refer to section 8-17-7)
Page 501
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-45 Replacement of battery pack for LP7-P9 R2 ext battery pack 8-45-1 Manpower 1 person, 30 minutes 8-45-2 Tools Standard Phillips Screwdriver, Hex-Wrench 8-45-3 Removal Procedure 1) Unscrew 2 screws and then pull out the BATTERY PACK COVER BRKT LP7_P9 from 6 points of LATCH.
Page 503
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 5) Disassemble LP7-P9 R2 EXT BATTERY PACK from extrude shape of BATTERY PACK FRAME BRKT. Figure 8-210 Removing LP7-P9 R2 EXT BATTERY PACK 6) Unscrew 8 screws to remove BATTERY PACK TOP BRKT.
Page 504
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 7) Lift up POLARIS BATTERY PACK and disconnect BATTERY Harness from PWA. Then remove POLARIS BATTERY PACK. NOTE: Once replacing battery pack, all of battery packs (4 pcs) need to be replaced in same time.
Page 505
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 2) Boot-up the console using power button on OPIO ASSY. Figure 8-214 Booting-up using power button 3) Check that battery icon is displayed on left bottom side.
Page 506
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-46 Replacement of battery pack for LP7-P9 R2 Battery option 8-46-1 Manpower 1 person, 30 minutes 8-46-2 Tools Standard Phillips Screwdriver, Hex-Wrench, Socket driver 8-46-3 Removal Procedure 1) Unscrew 2 screws to remove POWER CORD FIX BRKT.
Page 507
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 3) Unscrew 7 screws to remove the NEST EMI BRKT and pull out MIO ASSY from system. Figure 8-218 Unscrewing 7 screws to remove NEST EMI BRKT and MIO ASSY 4) Unscrew 12 screws to remove the MIO BOX BRKT.
Page 508
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 5) Disconnect 2 connectors of BATTERY and unscrew 4 screws to remove BATTERY ASSY. Figure 8-220 Removing BATTERY ASSY 6) Unscrew 4 screws to remove BATTERY TOP BRKT. Then remove POLARIS BATTERY PACK.
Page 509
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 8-46-5 Functional Check 1) Turn on the circuit breaker. Figure 8-222 Turning on circuit breaker 2) Boot-up the console using power button on OPIO ASSY. Figure 8-223 Booting-up using power button 3) Check that battery icon is displayed on left bottom side.
Page 510
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-47 Replacement of MBOB 8-47-1 Manpower 1 person, 20 minutes 8-47-2 Tools Standard Phillips Screwdriver 8-47-3 Removal Procedure 1) Remove POWER CORD FIX BRKT and POWER CORD.
Page 511
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 3) Unscrew 5 screws and remove REAR CABLE COVER BRKT. REAR CABLE COVER BRKT Figure 8-227 Unscrewing 5 screws 4) Remove CABLES (RED CIRCLE) and unscrew 7 screws, then remove NEST EMI BACK BKRT.
Page 513
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 7) Disconnect PWR and SIG connectors. Figure 8-231 Disconnecting PWR and SIG connectors 8) Disassemble BATTERY ASSY. Figure 8-232 Disassembling BATTERY ASSY 9) Disassemble MBOB PWR HARNESS (5728317) and MBOB SIG HARNESS (5728318)
Page 514
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 10) Disassemble MBOB PWA from MIO-B PWA Figure 8-234 Disassembling MBOB PWA 8-47-4 Installation Procedure All components to be installed in reverse order of removal. 8-47-5 Functional Check Refer to section 4-4 for Functional Diag Test List.
Page 516
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 2) Disassemble ACFL BOARD FIX BRKT(5538682). Figure 8-236 Disassembling ACFL BOARD FIX BRKT(5538682) 3) Disassemble Nut from the IOBOX-B Chassis Bracket Figure 8-237 Disassembling Nut from the IOBOX-B Chassis Bracket...
Page 517
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 4) Disassemble the following screws Figure 8-238 Disassembling the following screws 8-48-4 Installation Procedure All components to be installed in reverse order of removal. 8-48-5 Functional Check Refer to section 4-4 for Functional Diag Test List.
Page 518
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-49 Replacement of BBC 8-49-1 Manpower 1 person, 30 minutes 8-49-2 Tools Standard Phillips Screwdriver, Hex-Wrench 8-49-3 Removal Procedure 1) Unscrew 2 screws to remove COVER BRKT.
Page 519
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 3) Disassemble CONNECTOR BRKT using 2 screws. (2327596, SCREW WH M3X6) Figure 8-241 Disassembling CONNECTOR BRKT using 2 screws 4) Disconnect BATTERY SIG HARNESS and PWR HARNESS.
Page 521
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 6) Pull out LP7-P9 R2 EXT BATTERY PACK from BATTERY PACK FRAME BRKT LP7-P9 R2 EXT BATTERY PACK Figure 8-244 Pulling out LP7-P9 R2 EXT BATTERY PACK from BATTERY PACK FRAME BRKT...
Page 522
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 7) Unscrew 8 screws to remove BATTERY PACK TOP BRKT. BATTERY PACK TOP BRKT Figure 8-245 Unscrewing 8 screws to remove BATTERY PACK TOP BRKT 8) Remove 4 BATTERY PACKS.
Page 523
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 9) Disassemble BBC PWA Figure 8-247 Disassembling BBC PWA 8-49-4 Installation Procedure All components to be installed in reverse order of removal. 8-49-5 Functional Check Refer to section 4-4 for Functional Diag Test List.
Page 524
2.) Unlock cable clamp and carefully pull down 3 cables. 3.) Unscrew 1 screw and remove the cable clamp, and then extract out the 5 cables through the hole of OPIO BASE. OPIO BASE 5 CABLES 8-198 Section 8-50 - Replacement of LOGIQ P9 ROTATION MODULE KIT...
Page 525
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 4.) Unscrew nut and unfix the cable wire from the UP DOWN LINK BRKT. 5.) Unscrew nut and unfix the cable wire from the CABLE WIRE FIX BRKT.
Page 526
10.)Unscrew 2 screws and remove the ROTATION BUTTON from RATATE LEVER BRKT. ROTATION BUTTON ROTATE LEVER BRKT 11.)Unscrew 3 screws and remove the ROTATION MODULE ASSY from the OPIO base. 8-200 Section 8-50 - Replacement of LOGIQ P9 ROTATION MODULE KIT...
Page 527
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL When reassemble the ROTATION MODULE ASSY, HSH screw should be tightened proper CAUTION torque (60kgf-cm) When reassemble the screws, please use SCREW HSH M5X16 NYLOK(5729013). CAUTION 12.) Unscrew 6 screws and carefully remove the OPIO base.
Page 528
When reassemble the ROTATION MODULE ASSY, HSH screw should be tightened proper CAUTION torque (60kgf-cm). When reassemble the screws, please use SCREW FSH M5X16 NYLOK(5729014). Carefully unscrew the each screw to prevent wear of screw head. CAUTION ROTATION PLATE 8-202 Section 8-50 - Replacement of LOGIQ P9 ROTATION MODULE KIT...
Page 529
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 16.) Remove the THRUST NEEDLE BEARING WASHER 5578, THRUST NEEDLE BEARING 5578, OPIO SWIVEL SHAFT from the UP DOWN LINK ARM in order. Do not wipe out grease from part.
Page 530
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-51 Replacement of ODD Dummy Cover R3 8-51-1 Manpower 1 person, 5 minutes 8-51-2 Tools Standard Phillips Screwdriver 8-51-3 Pre work 1) Remove Rear Cover (refer to section 8-9)
Page 531
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 8-52 Replacement of Probe Light assy 8-52-1 Manpower 1 person, 20minutes 8-52-2 Tools Standard Phillips Screw driver 8-52-3 Pre work 1) Remove OPIO Panel assy (refer to section 8-14)
Page 532
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 2) Untie clamp for release Probe light cable Figure 8-250 Un-tie of Probe light cable from OPIO base 8-206 Section 8-52 - Replacement of Probe Light assy...
Page 533
ANUAL Chapter 9 Renewal Parts Section 9-1 Overview 9-1-1 Purpose of Chapter 9 This chapter gives an overview of replacement parts available for the LOGIQ P9/P7. Table 9-1 Contents in Chapter 9 Section Description Page Number Overview List of Abbreviations...
Page 534
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 9-2 List of Abbreviations Alternating Current Analog to Digital Converter A/N Key Alpha Numeric Keys ASIC Application Specific Integrated Circuit Assy Assembly MARU Core board(Merged Radio frequency module and Beaforming board)
Page 535
5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 9-3 LOGIQ P9/P7 Software Table 9-2 LOGIQ P9/P7 software (R1 to R2.5) Application SW Item Version Base SW Part Number Part Number R2.5 R1.0.0 5507301 5507300 R1.0.1 5507301-2 R1.0.2...
Page 536
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 9-4 Parts List Groups Figure 9-1 Console Views (R1 to R2.5) Figure 9-2 Console Views(R3) Table 9-4 Mechanical and user accessible parts Item Part Group Name...
Page 537
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Table 9-4 Mechanical and user accessible parts Item Part Group Name Table Number Description 501- Table 9-10 on page 9-22 Mechanical Parts and speaker Mechanical Part...
Page 538
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 9-5 Plastics Covers (Front/Sides/Rear) 9-5-1 Plastics Covers (Front/Sides/Rear) (R1 ~ R2.5) Figure 9-3 Plastics Covers (Front/Side/Rear) Table 9-5 Plastic Covers Item Part Name Part Number Description R2.5...
Page 539
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Table 9-5 Plastic Covers Item Part Name Part Number Description R2.5 5507028 Plastic Link Arm Top Cover OPIO Top Cover L, OPIO Top 5582886 OPIO Top Cover Set...
Page 543
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Table 9-7 User Interface Item Part Name Part Number Description R2.5 5501560-121 Full 21.5" LCD Monitor set with hinge 21.5" LCD 21.5 INCH LCD MONITOR 5501560-14 Full 21.5"...
Page 547
OPIO ASSY LOGIQ P9 SET Full OPIO set, inclduing Main and Sub OPIO 7603848 OPIO ASSY LOGIQ P9 R3 sets for LOGIQ P9 R3 (Different Emblem) 5503847 OP Main PWA Assembly OPIO Main PWA Assy LOGIQ P7/P9 OPIO 10.4 inch 5561038 Maru OPIO 10.4 LCD Set...
Page 548
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Table 9-8 OPIO Part Item Part Name Number Description R2.5 LP7P9 OPIO MAIN BOARD S5781465 LP7P9 OPIO MAIN BOARD PWA SET PWA SET S7503830 Probe Light Assy...
Page 549
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 9-8 Nest Box Parts MCB-R3 MDC-R3 402-1 403-1 Figure 9-9 Nest Box Parts Chapter 9 - Renewal Parts 9-17...
Page 551
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Table 9-9 Nest Box Parts Item Part Name Part Number Description R2.5 5503745 MBP Assy MBP Assy 5494666-4 MCB PWA MCB PWA 6494666-2 MCB-H Assy MCB-H...
Page 552
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Table 9-9 Nest Box Parts Item Part Name Part Number Description R2.5 Lcd Power Cable LCD Montior Cable 5582890 Lcd Hdmi Cable Assy MARU Lcd USB Cable...
Page 553
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 9-9 Mechanical Part Figure 9-11 Mechanical Parts 1 Chapter 9 - Renewal Parts 9-21...
Page 554
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Figure 9-12 Mechanical Parts 2 Table 9-10 Mechanical Parts Part Name Part Number Description R2.5 5400391-2 except Swivel Lock Caster (3 per console) Total Lock Caster...
Page 555
Option Tag Label Option Tag Label 5794163 Option Tag Label Option Tag Label R3 LOGIQ P9 ROTATION 5728965 LOGIQ P9 ROTATION MODULE KIT MODULE KIT S5794028 RS KNOB SET R3 RS KNOB SET R3 Chapter 9 - Renewal Parts 9-23...
Page 559
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Table 9-11 Options Part Item Part Name Number Description R2.5 5582894 Multi Purpose Holder and Gel Cup Multi Purpose Holder set 5380960-2 3 button switchs operated by the feet...
Page 560
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Part Item Part Name Number Description R2.5 5450061-2 Battery Battery 5341188 ECG Patient cable only for EU, ASIA IEC Type ECG CABLE Universal Video Converter S300,...
Page 561
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 9-11 Optional Peripherals and Accessories 9-11-1 Printers Figure 9-16 Optional Peripherals and Accessories - Printers Table 9-12 Optional Peripherals and Accessories - Printers Item Part Name...
Page 563
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 9-13 Power Cord Figure 9-17 Power Cord Chapter 9 - Renewal Parts 9-31...
Page 564
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Table 9-13 Power Cord Item Part Name Part Number Description R2.5 Power Cord for 5182611 Power Cord for India India Power Cord for 5182235 Power Cord for Switzerland...
Page 565
10-1-1 Preventive Maintenance Inspections It has been determined by engineering that your LOGIQ P9/P7 system does not have any high wear components that fail with use, therefore no Preventive Maintenance Inspections are mandatory. However, some Customers Quality Assurance Programs may require additional tasks and/or inspections at a different frequency than listed in this manual.
Page 566
The service procedures and recommended intervals shown in the Customer Care Schedule assumes that you use your LOGIQ P9/P7 for an average patient load (10-12 per day) and not use it as a primary mobile unit which is transported between diagnostic facilities.
Page 567
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 10-3-1 How often should care & maintenance tasks be performed? (cont?) Table 10-2 Customer Care Schedule Item Service at Indicated Time Notes • Recommend to clean filter at Air Filter Grid Remove the filter grid and clean the air filter.
Page 568
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Table 10-2 Customer Care Schedule Item Service at Indicated Time Notes Also after corrective Surface Probe Leakage • maintenance or as required by Current Checks your facilities QA program.
Page 569
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 10-4 Tools Required 10-4-1 Special Tools, Supplies and Equipment 10-4-1-1 Specific Requirements for Care & Maintenance Table 10-3 Overview of Requirements for Care & Maintenance...
Page 570
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 10-5 System Maintenance 10-5-1 Preliminary Checks The preliminary checks take about 15 minutes to perform. Refer to the system user documentation whenever necessary. Table 10-4...
Page 571
IRECTION EVISION ANUARY ERVICE ANUAL 10-5-2 Functional Checks The functional checks take about 60 minutes to perform. Refer to the LOGIQ P9/P7 Basic User Manual whenever necessary. 10-5-2-1 System Checks Table 10-5 System Functional Checks Step Item Description Verify basic B Mode (2D) operation.
Page 572
Clean LCD Monitor surface with a fluid detergent in warm water on a soft, damp cloth. LCD Monitor Caution: DO NOT spray any liquid directly onto the LOGIQ P9/P7 covers, LCD Monitor, keyboard, etc. 10-8 Section 10-5 - System Maintenance...
Page 573
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 10-5-4-2 Air filter cleaning Clean the system's air filters to ensure that a clogged filter does not cause the system to overheat and reduce system performance and reliability. It is recommended the filters be cleaned every two weeks, but the requirements will vary with environment.
Page 574
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 10-5-5 Physical Inspection Table 10-9 Physical Checks Step Item Description Labeling Verify that all system labeling is present and in readable condition. Scratches & Dents Inspect the console for dents, scratches or cracks.
Page 575
Failure to follow the prescribed cleaning or disinfection procedures will void the probe’s CAUTION warranty. DO NOT soak or wipe the lens with any product not listed in the LOGIQ P9/P7 Basic !! CAUTION: User Manual and/or care card. Doing so could result in irreparable damage to the probe and/or LOGIQ P9/P7 system.
Page 576
After you press a software download button, it may take up to three (3) seconds for the system to respond. NOTE: Software Updates through the GE service platform may not be available in all countries. 10-12 Section 10-6 - System Software Updates(Software Download) (R1, R2, R2.5 only)
Page 577
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 10-6-1 Software download and installation You receive notification of available software updates via the envelope icon that appears in the status bar on the bottom of the monitor.
Page 578
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL NOTE: If you are not logged in as an Administrator and there is a software package to download, you will see the following dialog when shutting down the system: Figure 10-5 Software Download Dialog If Not an Administrator If you get a dialog with “Unknown status,”...
Page 579
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 2) Press Download. Figure 10-7 System - Exit Dialog Window The download process starts. The progression of the download process is displayed. NOTE: The download step can be paused. While paused, you can return to normal operation. However, once the software installation has begun, the system is not useable until the software installation is done.
Page 580
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Press Download. Figure 10-8 Download in Progress Options • Stop: Stops the download process; no software upgrade is performed. If you select Stop, you WILL NOT be given an opportunity to download this package again.
Page 581
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Figure 10-10 Paused Software Download • Resume: Press Resume to continue the software download. 3) If auto-install was not checked, you will see the following dialog when the software download has finished (the message, “SW is ready to be installed,”...
Page 582
New Software Verification !! CAUTION: checklist. These verification results are tracked for regulatory purposes, sent back to GE for tracking, and approved with your signature. 10-18 Section 10-6 - System Software Updates(Software Download) (R1, R2, R2.5 only)
Page 583
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL NOTE: Press the Question mark to get information on how to check each feature. Figure 10-13 New Software Verification Checklist NOTE: You can move this dialog, but DO NOT close it. You can also minimize this dialog by clicking on the “<<“...
Page 584
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL When there is a failed feature noted on the New Software Verification Checklist, the following Warning dialog displays: Figure 10-15 Re-Install Previous Software Version Dialog When the New Software Verification Checklist contains a ‘Failed’ feature, you have two options available: 1) Re-Install the previous version of software (“OK”), or 2) Revert to the Checklist (“Cancel”)
Page 585
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 10-7 Using a Phantom Refer to the User Manual of the Phantom for information on using a phantom and quality assurance tests. For measurement accuracy of the system review chapter 13.5 of the Basic User Manual of LOGIQ P9/P7.
Page 586
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 10-8-2 GEHC Leakage Current Limits The following limits are summarized for ANSI/AAMI ES60601-1(For USA), IEC 60601-1 Medical Equipment Safety Standards and IEC 62353 Medical Electrical Equipment - Recurrent test and test after repair of medical electrical equipment.
Page 587
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Table 10-14 Type BF Patient Auxiliary Limits Country Conditions Others Close Protective earth and close neutral with normal polarity 0.1mA* 0.1mA*** Close Protective earth and close neutral with reverse polarity 0.1mA*...
Page 588
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 10-8-3 Outlet Test - Wiring Arrangement - USA & Canada Test all outlets in the area for proper grounding and wiring arrangement by plugging in the neon outlet tester and noting the combination of lights that are illuminated.
Page 589
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 10-8-4 Grounding Continuity Electric Shock Hazard! CAUTION The patient or operator MUST NOT come into contact with the equipment during this test !! CAUTION: Measure the resistance from the third pin of the attachment plug to the exposed protectively - earthed metal parts of the case.
Page 590
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 10-8-5 Enclosure Leakage Current Test 10-8-5-1 Definition This test measures the current that would flow in a grounded person who touched accessible non- protectively-earthed parts of equipment if the ground wire were to break. The test verifies the isolation of the power line from the chassis.
Page 591
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL When using Dale 600/601 or a similar test instrument, its power plug may be inserted into the wall outlet and the equipment under test is plugged into the receptacle on the panel of the meter. This places the meter in the grounding conductor and the current flowing from the case to ground will be indicated in any of the current ranges.
Page 592
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 10-8-6-2 Generic Procedure The testing meter is connected from accessible protectively-earthed parts of the equipment to ground. Measurements should be made with power line polarity normal and reversed, and with the neutral open and closed.
Page 593
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 10-8-7 Type BF Patient Leakage Current Test - Probe 10-8-7-1 Definition This test measures the current that would flow to ground from any of the probes through a patient who is being scanned and becomes grounded by touching some other grounded surface.
Page 594
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Table 10-18 Typical Data Sheet for Type BF Patient Leakage Current Test Tester Neutral Tester Polarity Switch Ground Switch Measured leakage current Open Normal Closed Open...
Page 595
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 10-8-8 Type BF Patient Leakage current - Mains to Probe Reference the procedure in the IEC 60601-1. Measure leakage current flow from mains to Probe. Electric Shock Hazard.
Page 596
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Section 10-9 When There’s Too Much Leakage Current... 10-9-1 Earth and/or Enclosure Fails Check the ground on the power cord and plug for continuity. Ensure the ground is not broken, frayed, or intermittent.
Page 597
GEDRAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL 10-9-5 New Unit If the leakage current measurement tests fail on a new unit and if situation cannot be corrected, submit a Safety Failure Report to document the system problem. Remove unit from operation.
Page 598
GERAFT LOGIQ P9/P7 5604324, R DRAFT (J 24, 2019) IRECTION EVISION ANUARY ERVICE ANUAL Type BF Patient Leakage Current Test - Probes (This test should be done for all probe) Table 10-23 Tester Tester Polarity Ground Max Value Value Comments...
Need help?