Introduction
WARNING
EMC— Magnetic and electrical fields are capable of interfering
with the proper performance of the device. For this reason make sure
that all external devices operated in the vicinity of the monitoring
system comply with the relevant EMC requirements. X-ray
equipment or MRI devices are a possible source of interference as
they may emit higher levels of electromagnetic radiation.
Changes or modifications to this device/system not expressly
approved by GE may cause EMC issues with this or other
equipment. This device/system is designed and tested to comply
with applicable standards and regulations regarding EMC and needs
to be installed and put into service according to the EMC
information stated as follows:
Use of known RF sources, such as cell/portable phones, or other
radio frequency (RF) emitting equipment near the system may cause
unexpected or adverse operation of this device/system. Consult
qualified personnel regarding device/system configuration.
The device/system should not be used adjacent to, or stacked with,
other equipment. If adjacent or stacked use is necessary, the device/
system should be tested to verify normal operation in the
configuration in which it is being used. Consult qualified personnel
regarding device/system configuration.
The use of accessories, transducers and cables other than those
specified may result in increased emissions or decreased immunity
performance of the device/system.
This device/system is suitable for use in all establishments other
than domestic and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic
purposes. Mains power should be that of a typical commercial or
hospital environment. Device is compliant to Class A.
WARNING
EXPLOSION HAZARD — Do not use this equipment in the
presence of flammable anesthetics, vapors or liquids.
WARNING
IMPROPER TRANSMITTER/LEADWIRE APPLICATION —
Applying a transmitter and/or leadwire that is not thoroughly dry to
a patient can result in an electrically conductive path being
established and a Leads Fail alarm not being provided if leadwires
come off the patient.
2001989-200E
ApexPro™
1-9