GE ApexPro Operator's Manual page 27

Telemetry system
Hide thumbs Also See for ApexPro:
Table of Contents

Advertisement

Introduction
CAUTION
RESTRICTED SALE — U.S. federal law restricts this device to sale
by or on the order of a physician.
CAUTION
SECURITY — The web browser which runs in conjunction with the
telemetry server is intended for hospital INTRANET use only. If
confidential patient information is made available from the hospital
intranet, the security of the data is the responsibility of the hospital.
CAUTION
SUPERVISED USE — This equipment is intended for use under the
direct supervision of a licensed health care practitioner.
CAUTION
UNINTENTIONAL RADIO FREQUENCY (RF)
INTERFERENCE — Unintentional RF interference could degrade
the reliability and performance of the wireless data link. The facility
must maintain an RF environment free from unintentional
interference. Refer to the service manuals for more information.
CAUTION
VENTILATION REQUIREMENTS — Set up the device in a
location which affords sufficient ventilation. The ventilation
openings of the device must not be obstructed. The ambient
conditions specified in the technical specifications must be ensured
at all times.
CAUTION
When using the ApexPro transmitter with a 3-leadwire set, an RF
signal continues to be transmitted upon removal of ECG leadwires.
A Leads Fail alarm appears at the central station however the
system does not transition to a No TELEM alarm.
Please take appropriate measures as explained in this manual for the
leads fail condition.
If an SpO2 and/or NIBP device is connected to the transmitter, an
NIBP or SpO2 limit alarm with warning or advisory as configured
by the user will supersede the existing Leads Fail system warning
alarm.
2001989-200E
ApexPro™
1-15