GE ApexPro Operator's Manual page 171

Telemetry system
Hide thumbs Also See for ApexPro:
Table of Contents

Advertisement

Graph all patients
Initiating a graph all patients request
2001989-200E
manual graph will be saved in alarm history. Additional data maybe available with
Full Disclosure.
Clicking on the Print button on the CIC Pro center display sends a Graph All Patients
request to all beds displayed on the CIC Pro center, initiating a 10-second graph for
all telemetry and Combo patients and a 20-second graph for all bedside patients.
When this option is selected for telemetry patients, graph requests always print at a
speed of 25 millimeters per second.
NOTE
This Graph All Patients function is only available when the single patient viewer
is closed. If a single patient viewer is open, selecting Print from the main menu
initiates a printout of whichever tab sheet is in front.
The print process stops automatically. If the Graph Stop control key is pressed on the
external DDW, the current patient's graph stops and the writer begins to print a graph
for the next patient.
If a patient's data is currently printing or is being saved to print when a Graph All
Patients request is initiated, this patient's data will not be included in the Graph All
Patients graph. This patient's data will print independently of the Graph All Patients
graph.
Clicking on the ECG parameter window for a patient whose data is saving will cancel
the Graph All Patients request for that patient.
If, while a Graph All Patients request is in progress, an arrhythmia alarm occurs for a
patient, the alarm data will replace the data that was saved for the Graph All Patients
request. The alarm data will then appear on the graph printout.
If, while a Graph All Patients request is running, a telemetry patient initiates a graph
from his or her transmitter, the saved data for the Graph All Patients graph will be
replaced by data from the patient's transmitter. The data from the transmitter will then
appear on the graph printout.
To initiate a Graph All Patients request, follow these steps.
1.
Click on the Print button on the CIC Pro center display. The Graph All Patients
window opens.
2.
Click on Limits or Waveforms.
Selecting Limits graphs all patient limits.
Selecting Waveforms graphs all patient waveforms.
3.
Click on the OK button to complete the Graph All Patients request.
ApexPro™
Printing
9-5

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents