Advertisement

Alarms

Alarm unit default settings

Telemetry alarm control defaults

5-12
You can set Telemetry Unit Defaults for Parameter Limits And Alarm Levels, as
well as for Arrhythmia Alarm Levels. These defaults are in effect for all telemetry
patients admitted to your unit, unless they are modified in an individual patient's
Alarm Control tab sheet.
To set Telemetry Alarm Control Defaults, follow the steps below:
NOTE
These settings are service-level defaults and are password protected.
1.
Click CIC Setup. A set of tabs displays.
2.
Click on the Service Password tab to bring it to the front.
3.
Use the keyboard to enter the service password, then click the Apply button. The
Current Permission entry changes from User to Service.
CAUTION
The service mode is intended for use only by qualified personnel
with training and experience in its use. The consequences of misuse
include loss of patient data, corruption of the CIC Pro center
operating system software, or disruption of the entire Unity network.
4.
Click on the Telemetry Alarm Control Defaults tab to bring it to the front.