Entering Moisture Probe
Calibration Data
2-12
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
You only need to enter calibration data for M and TF Series probes. It
is not necessary to enter calibration data for the Moisture Image
Series Probe unless you send the probe back to GE Panametrics for
calibration without its electronics module. If this is the case, enter the
calibration data for the recalibrated probe as described in this section.
The Series 1 will automatically download the new calibration data
into the Moisture Image Series Probe electronics module.
Note: You do not need to enter calibration data for the Moisture
Image Series Probe because it is stored in the probe's
electronics module. The Moisture Image Series Probe uploads
the calibration data into the Series 1 memory when needed.
Use the procedure below to enter the following data:
the probe serial number
the number of data points
the dew point and MH (or FH, depending on the probe) reading for
each data point
Procedure for Entering Moisture Probe Calibration Data
Referring to the Calibration Data Sheets, enter calibration data
separately for each probe on the designated channel as described
below. See Figure 2-14 on page 2-26 for a menu map.
Note: If you want to enter a constant value, refer to Entering
Constants and User Functions in Chapter 3 of the
Programming Manual. It is not necessary to enter the
calibration data if you are using a constant.
Initial Setup