Entering Calibration Data; Entering The System Calibration Menu - GE Moisture Image Series 1 Startup Manual

Hygrometer
Hide thumbs Also See for Moisture Image Series 1:
Table of Contents

Advertisement

Entering Calibration
Data
Initial Setup
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The Series 1 needs calibration data for moisture and oxygen probes.
GE Panametrics has already entered calibration data for probes;
however, you should verify this data before startup. You will need to
enter this data if you:
sent the probes back to the factory for calibration
use a different probe
use a non-GE Panametrics input
To enter, verify, or change calibration data for any probe type, you
must enter the System Calibration Menu as described in Table 2-3.
(See Figure 2-14 on page 2-26 for a menu map.)
.
Table 2-3: Entering the System Calibration Menu
Press the Main Menu key:
SETTINGS*
SYSTEM
CALIB
*The Settings Menu will prompt for a passcode (see page 2-5).
Refer to the appropriate section that follows to verify or enter
calibration data for:
Moisture Probes, page 2-12
Delta F Oxygen Cell, page 2-14
Be sure you have the Calibration Data Sheets that are supplied with
each GE Panametrics probe. Each Calibration Data Sheet consists of
a list of data points that you will need to enter or verify. Each
Calibration Data Sheet lists its corresponding probe serial number, as
well as the preassigned channel number. Calibration Data Sheets are
usually packed inside the probe cases.
IMPORTANT: Staple Calibration Data Sheets to the Program
Information List (Appendix A) and store them in a
safe place. If they are lost, contact GE Panametrics
for a duplicate. If you alter any of the calibration
data, make sure you make the change on the
Calibration Data Sheet.
April 2004
To enter the:
Settings Menu
System Settings Menu
System Calibration Menu
2-11