The last step in setup is entering high and low reference values. The
Series 1 requires reference values for its measurement circuitry for
moisture and oxygen inputs. The references are factory calibration
values that are specific to each channel card. Reference values are
located on a label placed on the side or front of the unit.
Note: If you receive a channel card replacement or change the
instrument program, you need to re-enter the reference data
for that channel.
Table 2-4 shows the key sequences for entering the Reference Menu.
(See Figure 2-14 on page 2-26 for a menu map.)
Table 2-4: Entering the Reference Menuu
Press the Main Menu key:
SETTINGS*
SERVICE
REFS
* The Settings Menu will prompt for a passcode (see page 2-5).
Use the appropriate section that follows to verify or enter reference
values for the probes. Once you verify and/or enter reference data for
each input, refer to Displaying Measurements on page 2-21.
IMPORTANT: You should record this data on the Program
Information List in Appendix A, or on a separate
sheet of paper, and keep it in a safe place.