Assigned And Unassigned Boxes; Example Of A Displayed Measurement - GE Moisture Image Series 1 Startup Manual

Hygrometer
Hide thumbs Also See for Moisture Image Series 1:
Table of Contents

Advertisement

June 2003
Displaying
Measurements (cont.)
2-22
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Channel
Measurement
Number
5 Pressure
27.10 PSIG
X Unassigned
Value
Figure 2-11: Assigned and Unassigned Boxes
The channel number is from 1 to 6, depending on how many channel
cards are installed in the unit. Additional channel cards can be
purchased separately and installed later (see Installing a Channel
Card on Chapter 2, page 2-22 of the Service Manual).
The measurement mode is the parameter currently being measured
by the Series 1. There are a variety of units for each measurement
mode. Table 2-5 on page 2-23 shows a list of the measurement modes
and a description of the units for each. The table also shows the
measurement modes and units as they are displayed on the screen. An
example of a measurement is shown in Figure 2-12.
The shaded area in Figure 2-11 represents the indicator area. When
you program the meter with constants or use Enhanced Response, a
symbol appears in this area. A "K" indicates a constant is being used
in the measurement and an "E" indicates the Series 1 is using
Enhanced Response to determine the measurement. An example of a
measurement using a constant and Enhanced Response activated is
shown in Figure 2-12.
The value is expressed in the units selected for a desired measurement
mode.
Measurement
Channel
Number
5
H
O
2
123.45
Value
Figure 2-12: Example of a Displayed Measurement
Units
Mode
Unassigned
Mode
E K
PPMw
Units
Indicators
Box
Indicators
Initial Setup