System Calibration Menu For Oxygen Cell - GE Moisture Image Series 1 Startup Manual

Hygrometer
Hide thumbs Also See for Moisture Image Series 1:
Table of Contents

Advertisement

Entering Delta F Cell
Calibration Data (cont.)
Initial Setup
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
5. Move the pointer to the Zero
µ
6. Enter the microamp (
7. Move the pointer to the Zero ppm (or %) line and press [YES].
8. Enter the zero range value, and press [YES]. The range units will
be either ppm, ppb or %, depending on the cell type/model
selected in the Probe Configuration Menu.
9. Repeat Steps 5 through 8 to enter the microamp (
value for the span line.
IMPORTANT: Leave the O2 Current Multiplier at 1.00 unless you
are using a background gas other than nitrogen. See
the section Background Gas Correction Factors for
the Delta F Oxygen Cell in Chapter 2,
Troubleshooting and Maintenance, of the Service
Manual.
Repeat Steps 1 through 9 to enter oxygen calibration data for the
remaining channels.
To enter calibration data for moisture probes, refer to A. Entering
Moisture Probe Calibration Data, page 2-12. (For other types of
probes, see Chapter 2 of the Programming Manual.) To exit, press the
DONE menu key until Main Menu appears on the message line. Then
refer to Entering High and Low Reference Values on page 2-16.
Oxygen Probe Calibration
S/N
µA
Zero:
___
Span:
___
O2 Current Multiplier: 1:00
System Calibration Menu
HELP
CHANNEL
Figure 2-7: System Calibration Menu for Oxygen Cell
µ
A line and press [YES].
A) value and press [YES].
µ
ppm O
2
___
___
PROBE
April 2004
A) and range
Ch
1
DONE
2-15