Custody Transfer; Operating Conditions; Verification - Siemens SITRANS FC430 Operating Instructions Manual

Coriolis flowmeters with hart
Hide thumbs Also See for SITRANS FC430:
Table of Contents

Advertisement

Custody Transfer

The SITRANS FC430 flowmeter is suitable for custody transfer measurement for liquids
other than water according to OIML R 117-1 with accuracy class 0.3.
Note
Ordering
Only flowmeters with local display ordered with Z-option "B31" (Custody Transfer) can be set
to this mode.
When "B31" is specified, the transmitter specification nameplate shows "SW Function" as
"CT standard".
Available versions:
• Standard: 7ME4613-XXXXX-XXX3-Z AXX+B31+EXX+FXX
• Hygienic: 7ME4623-XXXXX-XXX3-Z AXX+B31+EXX+FXX
• NAMUR: 7ME4713-XXXXX-XXX3-Z AXX+B31+EXX+FXX
9.1

Operating conditions

The operating conditions stated in the evaluation certificate may be reduced compared to the
operating conditions stated on the product nameplates. A copy of the certificate is included
on the SensorFlash and can be downloaded from www.siemens.com/FC430
(www.siemens.com/FC430).
Note
Operating conditions
Only the operating conditions stated in the evaluation certificate are valid.
For identification of custody transfer evaluated devices, see Figure 1-5 FCT030 approval
nameplate example (Page 18).
9.2

Verification

Custody transfer requirements
All custody transfer devices are verified on site using reference measurements.
FC430 with HART
Operating Instructions, 05/2015, A5E03361511-AF
9
161