Service and maintenance
11.3 Recalibration
Service information parameters
The basic service information parameters are:
● Driver Current
● Pickup 1 Amplitude
● Pickup 2 Amplitude
● Sensor Frequency
● Frame Temperature
● Process Media Temperature
● Zero Point Adjustment Auto/Manual
● Zero Point Offset Value
● Manual Zero Point
● Zero Point Standard Deviation
11.3
Recalibration
Siemens A/S, Flow Instruments offers to recalibrate the sensor at our works in Denmark.
The following calibration types are offered as standard according to configuration (standard,
density, °Brix/°Plato, fraction):
● Standard calibration
● Customer specified calibration
● Accredited Siemens ISO/IEC 17025 calibration
● Density calibration (incl. fraction setup if requested)
● Witness calibration
Note
SensorFlash
For sensor recalibration the SensorFlash memory unit must always be returned with the
sensor.
182
Operating Instructions, 05/2015, A5E03361511-AF
FC430 with HART