Zero Point Adjustment - Siemens SITRANS FC430 Operating Instructions Manual

Coriolis flowmeters with hart
Hide thumbs Also See for SITRANS FC430:
Table of Contents

Advertisement

Zero point adjustment

In the following the automatic zero point adjustment function is described in detail.
Note
Preconditions
Before a zero point adjustment is initiated, the pipe must be flushed, filled and at an absolute
flowrate of zero preferably also at operating pressure and temperature. Refer to "Zero point
adjustment" via HMI (Page 76) or PDM (Page 105) for instructions.
Note
Change of parameters during zero point adjustment
Do not change any other parameter during the zero point adjustment procedure.
Automatic zero point adjustment
SITRANS FC430 measures and calculates the correct zero point automatically.
The automatic zero point adjustment of the flowmeter is set by the following parameters:
● Zero Point Adjustment Period
● Start Zero Point Adjustment
When zero adjust is initiated by selecting "Start Zero Point Adjustment", the massflow values
are acquired and totalized for the configured period (Zero Point Adjustment Period). The
default zero point adjustment period (30 s.) is normally sufficient for a stable zero point
measurement.
Note
Extremely low flow quantity
If the flow quantity is extremely small, extremely precise measurement is necessary. In this
case, a long zero point adjustment period can be selected for improved zero point
adjustment.
FC430 with HART
Operating Instructions, 05/2015, A5E03361511-AF
D
379