Performance Log; Log Entries - ABB Endura AZ30 series Programming Manual

Integral probe and remote transmitter combustion oxygen monitor
Hide thumbs Also See for Endura AZ30 series:
Table of Contents

Advertisement

Endura AZ30 series integral probe and remote transmitter
Combustion oxygen monitor

9.3 Performance Log

9.3.1 Log Entries

Information displayed in the Performance Log is derived from the values obtained during calibration
routines.
When the Performance Log is selected, an icon (
successful or failed calibration or accuracy check; or if a new probe / cell has been fitted. Icon types are
shown in Table 9.6 below:
Fig. 9.2 Typical Performance Log and Performance Code Structure
Item
1
Calibration check
successful
Table 9.6 Performance Log Icons
Item
Description
hh:mm:ss
YYYY–MM–DD
2
Time and date of entry
Off
3
Calibration offset
Fct
4
Calibration factor
Table 9.7 Performance Log Codes
COI/AZ30E–EN Rev. D
1
hh:mm:ss
YYYY–MM–DD
Off
Fct
RR
Dev
t
Back
Calibration check
Accuracy check
failed
successful
in Fig. 9.2) at the left side of the display indicates a
x.xx mV
x.xxx
xx.xx %O2/s
xx.xx %O2
xxh xxm xxs
Icon
Accuracy check
failed
Item
Description
RR
5
Recovery rate
Dev
6
Deviation
t
7
Time duration
9 Troubleshooting
New probe / cell
93

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents