7 Calibration

Caution. Set up the test gas and reference air supplies before running a calibration – see OI/AZ30P-EN.

7.1 Overview of Calibration Procedures

Procedure
1-Pt AutoCal
2-Pt AutoCal
Manual Cal – Offset
Manual Cal – Factor
Manual Cal – 2-Pt
Table 7.1 Overview of Calibration Procedures
COI/AZ30E–EN Rev. D
Calibration Type and Description
Single-point automatic calibration of Calibration Offset
The values of the two test gases are compared and an offset
calibration is performed using whichever test gas is closest to air
(20.95 %).
If the values of the two test gases are equal, Test Gas 1 is used.
Two-point automatic calibration of Calibration Offset and Calibration Factor
Performs slope and offset calibration using both test gases.
Single-point manual calibration of Calibration Offset
The values of the two test gases are compared and an offset
calibration is performed using whichever test gas is closest to air
(20.95 %).
If the values of the two test gases are equal, Test Gas 1 is used.
Single-point manual calibration of Calibration Factor
The values of the two test gases are compared and a factor
calibration is performed using whichever test gas is furthest from air.
If the values of the two test gases are equal, Test Gas 2 is used.
Two-point manual calibration of Calibration Offset & Calibration Factor
Performs slope and offset calibration using both test gases.