ABB Endura AZ30 series Programming Manual page 74

Integral probe and remote transmitter combustion oxygen monitor
Hide thumbs Also See for Endura AZ30 series:
Table of Contents

Advertisement

Endura AZ30 series integral probe and remote transmitter
Combustion oxygen monitor
...Calibration Procedures
<Cal Type>
O2
1.00 %O2
Cal Offset
0.00 mV
Cal Factor
1.000
Reject
Reject
• New calibration
• New calibration
coefficients
discarded
Save successful
Accuracy
check entry added
to Performance
Log
• Return to Sensor
• Return to Sensor
Cal.
Save unsuccessful
• Display Failure
<Cal Type>
O2
20.95 %O2
Calibration Failed
Signal Not Stable
Exit
Exit
• Return to Sensor
Cal.
72
Calibration completed successfully
Accept
Accept
coefficients saved
Calibration
entry added to
Performance Log
Cal.
Accuracy
check entry added to
Performance Log
Calibration failed to complete
The reason for failure is displayed:
Signal Not Stable
No Response
test gas detected
NV Error
new calibration factor unsuccessful
Result Out of Bou
Result Out Of Bounds
calibration factor is outside the limits defined by
Factor Limit and / or the newly calculated
calibration offset is outside the Offset Limit
parameter – see page 46.
– stability criteria not met
– no response to the application of a
NV Err
– calibration successful but Save of
7 Calibration
– newly calculated
COI/AZ30E–EN Rev. D