Endura AZ30 series integral probe and remote transmitter
Combustion oxygen monitor

6.4.7 Input/Output

Menu
Input/Output
Exit
Parameter
Relay 1 (2)
Relay Assignment
Alarm 1 (2, 3, 4)
Cal. In Progress
Cal. Failed
Gas 1 Not Present
Gas 2 Not Present
Test Gas 1 Valve
Test Gas 2 Valve
Diag. – Failure
Diag. – Off Spec.
Diag. – Maint. Reqd. Activates the relay if a Maintenance Required
Diag. – Chk.
Function
Table 6.12 Input/Output Menu
58
Used to assign relays (1 and 2), digital I/O (1 and 2) and
7
current outputs (1 and 2).
Menus displayed at this level are dependent on the system
configuration / options fitted.
Select
Comment / [Range]
Selects the alarm and diagnostic states required to
activate the relay.
Each Relay Assignment parameter can be Assigned
or Not Assigned.
Note. Assigned parameters are OR'd together.
See Section 6.4.6, page 57 for alarm types and
triggers.
Relay is activated when a calibration is in progress.
Relay is activated if a calibration fails.
Relay is activated if Test Gas 1 is not detected.
Relay is activated if Test Gas 2 is not detected.
Relay is activated if Test Gas 1 valve is open.
Relay is activated if Test Gas 2 valve is open.
Activates the relay if a Failure diagnostic status is
generated – see Section 9.2.1, page 84.
Activates the relay if an Out of Specification
diagnostic status is generated – see Section 9.2.2,
page 87.
diagnostic status is generated – see Section 9.2.3,
page 89.
Activates the relay if a Check Function diagnostic
status is generated – see Section 9.2.4, page 91.
6 Programming
Default
Not
Assigned
(all
parameters)
COI/AZ30E–EN Rev. D