Endura AZ30 series integral probe and remote transmitter
Combustion oxygen monitor

6.4.2 Calibrate

Menu
Calibrate
Exit
Parameter
Sensor Cal.
1-Pt AutoCal
2-Pt AutoCal
Manual Cal – Offset
Manual Cal – Factor
Manual Cal – 2-Pt
Restore Defaults
Table 6.7 Calibrate Menu
40
Used to calibrate the sensor, select test gas types, set oxygen
2
hold action, enable automatic calibration hardware (if fitted)
and set calibration diagnostics options.
Note. Ensure reference air and test gas supplies are set up
before running a calibration – see Probe User Guide
OI/AZ30P–EN.
Select
Comment / [Range]
See Section 7.1, page 69 for calibration routines.
Note. Calibration options 1-Pt AutoCal to Manual Cal -
2-Pt are displayed only when the probe reaches 690 °C
(1274 °F) and all error messages are cleared.
Restore Defaults is displayed only when the probe
temperature is < 690 °C (1274 °F).
Performs a single-point automatic calibration.
Enabled only if calibration is currently inactive, Autocal is
fitted and temperature has stabilized.
Performs a two-point automatic calibration.
Enabled only if calibration is currently inactive, Autocal is
fitted and temperature has stabilized.
Performs a manual single-point offset calibration.
Enabled only if calibration is currently inactive and when
temperature has stabilized.
Performs a manual single-point factor calibration.
Enabled only if calibration is currently inactive and when
temperature has stabilized.
Performs a manual two-point calibration.
Enabled only if calibration is currently inactive and when
temperature has stabilized.
Restores the factory calibration offset and factor values
entered in Easy Setup – see page 39.
6 Programming
Default
N / A
COI/AZ30E–EN Rev. D