Chapter 5: Calibration and Verification
When the Analyzer measures the reagent water, it also measures the TOC in the reagents and dilution water
used in the measurement. Because the TOC concentration in the reagent water is very low, this measurement
approximates the TOC concentration in the reagents and dilution water for the particular volumes of reagents
and dilution water used in the calibration. The Analyzer corrects the calibration for the TOC in the reagents and
dilution water by setting the measurement measured response (i.e., Mass Response) to equal 0 ppm.
The operator enters the concentration of point 2 into the calibration protocol. The concentration of the standard
used for point 2 is referred to as C
100 ppm, so C
= 100 ppm. C
2
The calibration constant, b, is calculated from the difference in the response to the two points:
where,
R
= Response of Analyzer to point 2
2
R
= Response of Analyzer to reagent water
RW
C
= TOC concentration of point 1 (= 0 for RW)
1
C
= TOC concentration of point 2 (mg/L)
2
GE Analytical Instruments ©2013
Figure 16: Two-Point Calibration (Point 1 is Reagent Water)
. In the example in Figure 16, the operator set the concentration of point 2 to
2
(or C
) for reagent water is zero.
1
RW
b =
94 of 238
C
- C
2
RW
R
- R
2
RW
DLM 68100-06 EN Rev. A