Step 13: Calibrate The Analyzer - GE Sievers InnovOx Operation And Maintenance Manual

On-line toc analyzer
Hide thumbs Also See for Sievers InnovOx:
Table of Contents

Advertisement

Chapter 4: Installation
8.
Ensure that the other settings in the Protocol Setup/Advanced Setup screen are:
• Sparge:
• Blank Correction:
If any of the settings need to be changed, press the button that is to be changed, and the choices will
show up on the right side of the screen, as described above.
9.
Press Back then select the Maintenance tab.
10. Press the Flush Reagents button, then press the Flush All button. The reagent, dilution water, and
check standard lines will be flushed. All bubbles, except for very small ones, should be flushed out of the
lines when the Flush is complete. If not, press Flush All again. (If one or more of the reagent, dilution
water, or check standard lines are not flushed, see Chapter 9, "Troubleshooting")
11. When the flush is completed, press the Back button to return to the Menu screen.
12. Press the Back button.
13. When ready, press the Start Analysis button. Press Main to return to the Main screen. A graph of the
CO
peaks will be displayed. Each peak appears after its peak height is measured, and the measured
2
NPOC concentration appears at the top of the screen.
14. When the analysis is complete, the CO
the Mass Response axis is only 1.0 µg full scale, then refer to the Troubleshooting section of this manual.
Otherwise, proceed to the next step.

Step 13: Calibrate the Analyzer

Before placing the Analyzer into normal operation, calibrate the Analyzer in a manner suited to your application.
Consult Chapter 5, "Calibration and Verification" for details on calibration options.
GE Analytical Instruments ©2012
0.8 min.
Off
peaks will be displayed on the screen. If there are no peaks and
2
88 of 238
DLM 68100-06 EN Rev. A