Reactor Outlet Valve Leak Check - GE Sievers InnovOx Operation And Maintenance Manual

On-line toc analyzer
Hide thumbs Also See for Sievers InnovOx:
Table of Contents

Advertisement

Chapter 9: Troubleshooting

Reactor Outlet Valve Leak Check

Refer to see Figure 74 on page 205, as needed.
Press the Reactor Outlet Valve button to start the test.
The following valves are closed: V1-R, V3-R, V5-R, V6-R and V7-SH.
Valves V2-R, V4-R and V7-R are opened.
The Gas Control Module pressurizes the lines using the optional air pump P1-GS or the pressurized gas
source for 10 sec., then valve V2-GS closes to trap that pressure in the area being tested. During this
10-sec. pressurization period, the timer remains at 00:00.
Monitor the pressure appearing on the right side of the display. Note the pressure at time = 00:00, then
note the pressure 30 sec. later (time = 00:30).
Press the Reactor Outlet Valve button to stop the test.
The pressure should not have dropped more than 3 psia. If the pressure did drop more than that
amount, but it passed the previous leak check, then seal in V5-R may need to be replaced.
Remove the plug on the NDIR end cap outlet gas line.
GE Analytical Instruments ©2013
204 of 238
DLM 68100-06 EN Rev. A