Chapter 2 Equipment Requirements; System Components; Presentation Of The 3 Rooms; Examination Room - GE Optima IGS 320 Preinstallation Manual

Table of Contents

Advertisement

Chapter 2 Equipment Requirements

1 System Components

1.1 Presentation of the 3 Rooms

1.1.1 Examination room

Gantry
Omega Patient Table
Performix X–Ray Tube Assembly
Collimator
Digital Detector
TSUI
ITU

1.1.2 Technical room

System Control Cabinet
Coolix 4100 chiller
Detector Conditioner
Fluoro UPS (option) (1 Cabinet)
Fluoro UPS IF box (option)
PDB
EMC Filter (Option)
Chapter 2 Equipment Requirements
WARNING
THE ELECTRONIC CABINETS (SYSTEM CONTROL CABINET, OPTIONAL LD
CABINET, COOLIX 4100 CHILLER, DETECTOR CONDITIONER AND FLUORO
UPS WHEN INSTALLED) INCLUDE FANS THAT ARE CREATING AIR-
CIRCULATION OF PULSED-AIR. WHEN THIS PULSED AIR IS IN AN
ENVIRONMENT THAT MAY CONTAIN AIRBORNE PATHOGENS LIKE AN
EXAM ROOM/CONTROL ROOM, THERE IS A RISK OF TRANSMISSION OF
THESE AIRBORNE PATHOGENS FROM PATIENTS TO OTHER PATIENTS
OR CLINICAL PERSONNEL (NOSOCOMIAL DISEASES).
TO REDUCE THIS RISK, THE ELECTRONIC CABINETS MUST BE
INSTALLED IN A ROOM SEPARATED FROM EXAM ROOM/CONTROL ROOM,
I.E., TECHNICAL ROOM.
Optima IGS 320, Optima IGS 330 Pre-Installation Manual
Direction 5537562-1-1EN, Revision 3
43