Summary of Contents for GE IV-908

  • Page 1 GE Healthcare Valves IV-908, PV-908 Instructions...
  • Page 2 – connected to other CE-marked GE Healthcare instruments, or – connected to other products recommended or described in this manual, and – used in the same state as it was delivered from GE Healthcare except for alterations described in this manual. Recycling...
  • Page 3 The valves have 8 positions and each position connects the central port with one of the peripheral ports. The valve IV-908 is used on the low pressure side (before the pump) and PV-908 on the high pressure side (after the pump).
  • Page 4 1 Introduction Instructions 56-8101-02 Edition AE...
  • Page 5: Installation

    Installation 2 2 Installation CAUTION! Before connecting Valves IV-908 or PV-908 ensure the power is switched OFF at the system pump or for the complete system. Unpacking Unpack the instrument and check the items against the packing list. Inspect the items for obvious damage which may have occured during transportation.
  • Page 6 2 Installation Instructions 56-8101-02 Edition AE...
  • Page 7: Operation

    Operation 3 3 Operation The valves are controlled from UNICORN. Storage Overnight: The valves can be left filled with a buffer. Overnight and Long time storage: Flush the valves with water and then fill with 20% ethanol. Instructions 56-8101-02 Edition AE...
  • Page 8 3 Operation Instructions 56-8101-02 Edition AE...
  • Page 9: Maintenance

    Maintenance 4 4 Maintenance CAUTION! Only spare parts approved or supplied by GE Healthcare may be used for maintaining and servicing the valves. Period Action Every 12 months Change channel plate and or when required distribution plate Cleaning-in-place Pump a cleaning or sanitizing agent through the valve. The standard recommendation is to pump 1 M NaOH for 30 minutes and wash out with buffer.
  • Page 10 4 Maintenance Recycling This symbol indicates that the waste of electrical and electronic equipment must not be disposed as unsorted municipal waste and must be collected separately. Please contact an authorized representative of the manufacturer for information concerning the decommissioning of equipment. Instructions 56-8101-02 Edition AE...
  • Page 11: Troubleshooting

    Trouble shooting 5 5 Trouble shooting If the suggested actions do not correct the fault, call GE Healthcare. Fault Action The valve is not switching Check the connection to the pump. The valve should be connected to the UniNet 2 socket, not the UniNet 1 socket.
  • Page 12 5 Trouble shooting Recycling This symbol indicates that the waste of electrical and electronic equipment must not be disposed as unsorted municipal waste and must be collected separately. Please contact an authorized representative of the manufacturer for information concerning the decommissioning of equipment. Instructions 56-8101-02 Edition AE...
  • Page 13: Reference Information

    The main difference between the IV-908 and PV-908 is the diameter of the flow channels and the maximum pressure ratings. PV-908 has 0.8 mm channels while IV-908 has 1.2 mm channels which allow higher flow rates at lower back pressure.
  • Page 14 • used in laboratory locations • used in the same state as it was delivered from GE Healthcare except for alterations described in the user manual • connected to other CE labelled GE Healthcare instruments or other products as recommended. Instructions 56-8101-02 Edition AE...
  • Page 15 Reference information 6 Accessories and spare parts Item Quantity Code no. per pack Valve IV-908 including one UniNet cable 18-1108-42 Valve PV-908 including one UniNet cable 18-1108-41 Valve kit, including channel plate and distribution plate IV-908 18-1109-07 PV-908 18-1109-06 Cable UniNet, 0.7 m...
  • Page 16 Contact your local GE Healthcare representative for the most current information. UNICORN Any use of this software is subject to GE Healthcare Standard Software End-User License Agreement for Bio-Sciences Software Products. © 2006 General Electric Company – All rights reserved.