Table of Contents

Advertisement

Innova
TM
Innova
TM
Pre-Installation Manual
IGS 520, Innova
IGS 540
IGS 530,
TM
OPERATING DOCUMENTATION
5499972-1-1EN
Revision 2

Advertisement

Table of Contents
loading

Summary of Contents for GE Innova IGS 520

  • Page 1 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual OPERATING DOCUMENTATION 5499972-1-1EN Revision 2...
  • Page 2: Direction 5499972-1-1En, Revision

    Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2...
  • Page 3 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Important Information LANGUAGE ПРЕДУПРЕЖДЕНИЕ Това упътване за работа е налично само на английски език. (BG) • Ако доставчикът на услугата на клиента изиска друг език, задължение на клиента...
  • Page 4 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 VÝSTRAHA Tento provozní návod existuje pouze v anglickém jazyce. • (CS) V případě, že externí služba zákazníkům potřebuje návod v jiném jazyce, je zajiště‐ ní překladu do odpovídajícího jazyka úkolem zákazníka. •...
  • Page 5 FIGYELMEZTETÉS Ezen karbantartási kézikönyv kizárólag angol nyelven érhető el. • (HU) Ha a vevő szolgáltatója angoltól eltérő nyelvre tart igényt, akkor a vevő felelőssége a fordítás elkészíttetése. • Ne próbálja elkezdeni használni a berendezést, amíg a karbantartási kézikönyvben leírtakat nem értelmezték.
  • Page 6 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 AÐVÖRUN Þessi þjónustuhandbók er aðeins fáanleg á ensku. • (IS) Ef að þjónustuveitandi viðskiptamanns þarfnast annas tungumáls en ensku, er það skylda viðskiptamanns að skaffa tungumálaþjónustu. •...
  • Page 7 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 ĮSPĖJIMAS Šis eksploatavimo vadovas yra tik anglų kalba. • (LT) Jei kliento paslaugų tiekėjas reikalauja vadovo kita kalba – ne anglų, suteikti vertimo paslaugas privalo klientas. •...
  • Page 8 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 ATENŢIE Acest manual de service este disponibil doar în limba engleză. • (RO) Dacă un furnizor de servicii pentru clienţi necesită o altă limbă decât cea engleză, este de datoria clientului să...
  • Page 9 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 VARNING Den här servicehandboken finns bara tillgänglig på engelska. • (SV) Om en kunds servicetekniker har behov av ett annat språk än engelska, ansvarar kunden för att tillhandahålla översättningstjänster. •...
  • Page 10 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 This page left intentionally blank. Important Information...
  • Page 11 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Revision History Part/Rev Date Reason for Change Pages 5499972-1-1EN rev 1 January 27, 2015 Initial release of direction 5499972-1-1EN 5499972-1-1EN rev 2 March 10, 2015 Second release of direction 5499972-1-1EN Revision History...
  • Page 12 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 This page left intentionally blank. Revision History...
  • Page 13: Table Of Contents

    Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Table of Contents Chapter 1 GENERAL REQUIREMENTS....................21 1 Objectives & Overview........................21 1.1 Object and Scope of this manual..................21 1.1.1 Object and Scope......................21 1.1.2 Quebec........................
  • Page 14 1.2 Description of Innova System....................49 1.2.1 Description of the Gantry..................... 49 1.2.2 Description of the Patient table..................50 1.2.2.1 Innova IGS 520 Case..................50 1.2.2.2 Innova IGS 530 and Innova IGS 540 Case............50 1.2.3 User Interfaces......................52 1.2.4 Accessories for Innova Table and OR Table............
  • Page 15 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 1.2.11 CENTRICITY CA1000 option..................63 1.2.12 Injectors........................63 1.2.13 Fluoro UPS........................ 64 1.2.14 Volcano IVUS s5i Option................... 66 1.2.14.1 Purpose......................66 1.2.14.2 Product Description..................66 1.2.15 ECG Acquisition kit Option..................
  • Page 16 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 2.2.2 Control Room Layout....................118 2.2.3 Technical Room Layout..................... 119 2.2.4 IVUS Room Layout....................120 2.2.4.1 IVUS Rev 2 and Rev 2+..................120 2.2.4.2 IVUS Rev 3....................... 120 2.2.5 ECG Device Room Configurations................
  • Page 17 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 3.3.3.3 2285050 – Assembly and separation select kit..........153 3.3.3.3.1 2285051 – Above grade, Floor Anchors without insert......153 3.3.3.3.2 2285646 – Through bolts without insert........... 153 3.3.3.3.3 2286398 –...
  • Page 18 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 2 EMI Consideration........................180 2.1 General Scope........................180 2.2 Electromagnetic Emission....................180 2.3 Electromagnetic Immunity....................180 2.3.1 Electromagnetic Immunity IEC 60601-1-2..............180 2.3.2 Recommended Separation Distances for Portable and Mobile RF Communications Equipment IEC 60601-1-2...........
  • Page 19 2.1.3 Physical Run - System Core Detail................217 2.2 MIS (Master Interconnect System)..................221 2.3 IVUS Wiring........................222 2.3.1 For IVUS rev 2 wiring....................222 2.3.2 For S5I GE rev 2+ wiring................... 222 2.3.3 For IVUS rev 3 wiring....................223 2.4 Cable Channeling....................... 224 2.4.1 General........................224 2.4.2 Conduit........................
  • Page 20 IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 2.4.4.2.2 For S5I GE Rev 2+ option................229 2.4.4.2.3 For IVUS Rev 3 option................230 2.4.5 Vivid E9 Ultrasound Option..................230 3 Lighting Specifications.........................231 3.1 Room Light Distribution...................... 231 3.1.1 Requirements for lighting...................231...
  • Page 21: Chapter 1 General Requirements

    1.1.1 Object and Scope This document is intended as a guide and information resource to properly plan and prepare a site for the installation of an Innova system (This includes the Innova IGS 520, Innova IGS 530 and Innova IGS 540).
  • Page 22: Pre-Installation Process

    Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 1.2 Pre-Installation Process Complete the checklists in ROOM LAYOUTS, ELECTRICAL REQUIREMENTS, and GENERAL REQUIREMENTS of this manual. They represent an important part of the pre–installation process.
  • Page 23: Customer Responsibilities

    Be sure to inform the customer that the cost of any alteration or modification not specified in the sales contract are liable to the customer. The following GE-supplied equipment must be installed by the Hospital's Contractors, per room drawings: •...
  • Page 24: Equipment Classifications

    Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 2.2 Equipment Classifications The following equipment classifications are applicable to the product: Classification category Equipment classification Protection against electric shock Class I WARNING TO AVOID THE RISK OF ELECTRIC SHOCK, THIS EQUIPMENT MUST ONLY BE CONNECTED TO A SUPPLY MAINS WITH PROTECTIVE EARTH.
  • Page 25 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 NOTICE The product is not classified as AP, APG (Equipment not suitable for use in the presence of a flammable anaesthetic mixture with air or with oxygen or nitrous oxide).
  • Page 26: Pre-Installation Checklist

    IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 2.3 Pre-Installation Checklist GE Healthcare Site Readiness Checklist GEHC Global Order # : Customer: GEHC On-site Representative : MI Supplier: Name of customer reviewed with :...
  • Page 27 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Floor is clean and prepared for final floor covering. Customer has verified floor lev‐ eling meets the equipment installation drawings and PIM specs and no visible defects are observed.
  • Page 28: Delivery Requirements

    Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 3 Delivery Requirements 3.1 Shipping Information 3.1.1 Product Shipping Information Refer to the table below. To obtain shipping information for components not specified in the table, refer to the appropriate component Pre-Installation Manual listed in Chapter 2, Basic Innova System...
  • Page 29: Detail Of Innova Shipping Information

    Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 DIMENSIONS MM (INCHES) PRODUCT OR COMPO‐ WEIGHT POUNDS (KILO‐ METHOD OF SHIPMENT NENT GRAMS) Height Width Depth Fluoro UPS CE (*) 1750 (68.9) 890 (35) 1000 (39.4) 1287 (585) On pallet...
  • Page 30: Gantry Air Shipment

    Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 3.1.2.2 Gantry Air Shipment Illustration 1-3: 3 Delivery Requirements...
  • Page 31: C1 And C2 Cabinet In Crate, On Pallet

    Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 3.1.2.3 C1 and C2 Cabinet in crate, on pallet Illustration 1-4: The shipping weight (crate and pallet) is 81kg (179 lbs). Chapter 1 General Requirements...
  • Page 32: Omega Shipment

    Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 3.1.2.4 Omega Shipment Illustration 1-5: 3 Delivery Requirements...
  • Page 33: Omega Using Positioner Dollies

    Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 3.1.2.5 Omega Using Positioner Dollies Illustration 1-6: Chapter 1 General Requirements...
  • Page 34: Innova Iq Table And Or Table Shipment

    Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 3.1.2.6 Innova Table and OR Table Shipment Illustration 1-7: Table Shipment 3 Delivery Requirements...
  • Page 35: Large Display Option

    Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 1-8: Covers Shipment 3.1.2.7 Large Display Option Illustration 1-9: Large Display Monitor on pallet Chapter 1 General Requirements...
  • Page 36: Other Elements Package

    Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 1-10: Large Display cabinet on pallet 3.1.2.8 Other Elements Package NOTE: All OEM parts are shipped inside there original boxes group as needed on pallets. Illustration 1-11: 3 Delivery Requirements...
  • Page 37: Tools And Test Equipment

    Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 3.2 Tools and Test Equipment Refer to Table 1-2. To obtain a list of tools and test equipment for components not specified in Table 1-2, refer to the appropriate component Pre-Installation Manual listed in Chapter 2, Basic Innova System Compatibility.
  • Page 38 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 PRODUCT OR COM‐ TOOL OR TEST EQUIP‐ USED FOR SOURCE RECEIVED (DATE) PONENT MENT DL User parts Monitor Suspension Chiller Phillips/Flathead screwdriver. Open chiller. Install wiring and hoses. Chiller Autotransformer Phillips/Flathead screwdriver.
  • Page 39: Door Size Requirements

    Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 3.3 Door Size Requirements Minimum door sizes also apply to hallways and elevators. For additional details, refer to Shipping Information. 3.3.1 Door Height The minimum door height (to accommodate Innova positioner on its dolly) is 1.980 m (78 in). If the height is limited to 2 m (79 in), you will need a fall over cabinet.
  • Page 40: Route Survey

    Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 3.4 Route Survey 3.4.1 Step One — Sketch Start preparing Route Survey by sketching a floor plan of the hospital or clinic which will receive the equipment.
  • Page 41 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Table 1-3: Ref. no. Area: loading dock, doorway, hallway, turn, elevator, Limitations: loading capacity, height, width, depth floor‐ obstruction, etc. ing material, radius, etc. Chapter 1 General Requirements...
  • Page 42: European Process Order Select

    Installation Specialist on Site. All cable and component selections must be sent at least 6 weeks before RSDD to your PMC. The selection tool is available on the website: http://supportcentral.ge.com/products/ sup_products.asp?prod_id=99484 3 Delivery Requirements...
  • Page 43: Product Storage And Handling Requirements

    Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 4 Product Storage and Handling Requirements 4.1 Product Storage and Handling limits NOTICE Avoid extremes in temperatures To obtain Temperature, Humidity or Pressure requirements for components not specified in Table 1-4, refer to the appropriate component Pre–Installation Manual listed in Chapter 2, Basic...
  • Page 44: Fluoro & 3Kva Ups Shipment Duration

    Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 NOTE: (2) Special Humidity Instructions: The following parts can be can be shipped in standard shipment conditions with the requirement that on arrival to installation site, and before supplying power to these parts, they shall be kept in an environmental relative humidity equal or lower than their specified capability, and that’s for a minimum of 48 hours.
  • Page 45: Handling Instructions

    Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 4.3 Handling instructions The packaging of the following components must be marked with special handling instructions for transport and storage. • C1 Cabinet and C2 Cabinet: Illustration 1-13: •...
  • Page 46 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 This page left intentionally blank. 4 Product Storage and Handling Requirements...
  • Page 47: Chapter 2 Equipment Requirements

    Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Chapter 2 Equipment Requirements 1 System Components 1.1 Presentation of the 3 Rooms 1.1.1 Examination room • Gantry • Omega / Tilting table • Performix X–Ray Tube Assembly •...
  • Page 48 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 • DL Keyboard • Monitors 1 System Components...
  • Page 49: Description Of Innova System

    TSSC, Smart Box or Smart handle (options for remote installation). Illustration 2-1: • L-arm • Pivot • X-Ray LED • C-arc • Motorized elevator for the Revolution Digital Detector, • Revolution Digital Detector ○ 21 cm Revolution Digital Detector for Innova IGS 520. Chapter 2 Equipment Requirements...
  • Page 50: Description Of The Patient Table

    1.2.2 Description of the Patient table 1.2.2.1 Innova IGS 520 Case Innova IGS 520 can be supplied with an OMEGA IV Compact patient table, an OMEGA V long patient table or a Tilting table. 1.2.2.2 Innova IGS 530 and Innova IGS 540 Case Innova IGS 530 and Innova IGS 540 can be supplied with an OMEGA V Long patient table or Tilting table.
  • Page 51 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-3: Innova Table Illustration 2-4: OR Table Chapter 2 Equipment Requirements...
  • Page 52: User Interfaces

    Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 1.2.3 User Interfaces Illustration 2-5: User interfaces with Omega table 1 System Components...
  • Page 53 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-6: User interfaces with Innova Table Chapter 2 Equipment Requirements...
  • Page 54 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-7: User Interfaces with OR Table 1 System Components...
  • Page 55: Accessories For Innova Table And Or Table

    Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 1.2.4 Accessories for Innova Table and OR Table Illustration 2-8: Chapter 2 Equipment Requirements...
  • Page 56: Description Of The X-Ray Generator

    Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-9: 1.2.5 Description of the X-Ray generator The Innova System uses a High Voltage System with component parts as follows according to the power requested (100 kW): •...
  • Page 57: Description Of The X-Ray Head

    Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 • Ingrid H.V. tank mounted on the Performix 160A X-Ray tube housing. 1.2.6 Description of the X-Ray head The Innova System uses a Performix 160A X-Ray tube housing: •...
  • Page 58 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 • LCD 19” monitors. • LCD 58’’ EIZO GmbH LS580W Color Flat Panel (LDM option) Illustration 2-11: C1 Cabinet 1 System Components...
  • Page 59: Description Of The Image Monitors Suspension

    Illustration 2-13: Thermo-con Digital detector chiller 1.2.8 Description of the image monitors suspension GE provides as option several types of suspensions; alternatively, the customer can install the suspension of his choice (“open suspension”), provided all requirements in the paragraph Section 1.2.8.3 are met.
  • Page 60: Lcd Monitors Suspension (Without Ldm)

    In the OR configuration of the system, it is mandatory to use a suspension that is compatible with the OR environmental constraints. The suspension provided by GE are not compatible with the OR environment and shall not be used. 1.2.8.1 LCD monitors suspension (without LDM) The system can be equipped with a suspension for 4 LCD monitors or 6 LCD monitors.
  • Page 61 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 4. KVM–local 5. Video Splitter • The user area is made of: 1. I/R Receiver / Emitter 2. DL color LCD monitor (19”) 3. Keyboard, 4.
  • Page 62 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-14: Innova User Area devices 1 System Components...
  • Page 63: Aw Workstation

    Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 1.2.10 AW workstation AW workstation option is composed of a workstation, 1 or 2 monitors 19’’ flat panel in the Control Room and 1 monitor (flat panel) fixed on suspension (option) and 1 video switcher. Check Innova Installation Service Manual for switcher installation.
  • Page 64: Fluoro Ups

    Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 NOTE: With Tilting Table: • ACIST CVI Table mount injector is not authorized. • Table Mounted Injectors are not allowed to mount on the OR Table as they are not compatible: ○...
  • Page 65 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-15: Fluoro UPS Chapter 2 Equipment Requirements...
  • Page 66: Volcano Ivus S5I Option

    Manual. 1.2.14.1 Purpose Volcano is a manufacturer of Intravascular Ultrasound Systems (IVUS), which are used in the cath lab. GE has a strategic agreement with Volcano. GE will be selling and servicing some of the Volcano products. WARNING FOR OR TABLE, IVUS ACCESSORIES MUST BE INSTALLED ON A CART 1.2.14.2 Product Description...
  • Page 67 IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 • Patient Interface Modules as the S5I GE PIM, the PIMr for rotational catheters and Pimette for the FFR catheter, located in the Procedures Room, • Various user interface units to control the system ○...
  • Page 68 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-17: IVUS Rev 2 components 1 System Components...
  • Page 69: Ecg Acquisition Kit Option

    Illustration 2-18: IVUS Rev 2+ components 1.2.15 ECG Acquisition kit Option The ECG Acquisition kit allows for the connection of an ECG device (such as GE ECG devices MacLab, CardioLab or ComboLab) to the Innova system (C2 cabinet). The ECG Acquisition kit hardware consists of a Hubican and Physio modules and their associated cables.
  • Page 70: Vivid E9 Diagnostic Ultrasound System Option

    IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 • ECG devices installed in Control Room (GE ECG devices like MacLab, CardioLab or ComboLab) • ECG devices installed in Exam Room (Non-GE ECG device) Illustration 2-19: ECG Acquisition kit layout 1.2.16 Vivid E9 Diagnostic Ultrasound System Option...
  • Page 71 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-20: VIVID E9 major components VIVID E9 major components, refer to Illustration 2-20: • 1 Main Monitor • 2 Alphanumeric Keyboard • 3 Upper panel with Stereo Loudspeakers and Touch Monitor •...
  • Page 72: Large Display Subsystem Option

    Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 1.2.17 Large Display Subsystem option The Innova system can integrate a Large Display solution to: • See images larger at full IQ with greater flexibility in monitor distance in the procedure room •...
  • Page 73: Dimension Drawings

    Patient Table side clearence (CPR access) Illustration 2-30 Table Head Extender Illustration 2-31 Gantry and Omega IV Compact Patient Table Relative Positions (Innova IGS 520) Illustration 2-32 Gantry and Omega V Long Patient Relative Positions Illustration 2-33 C2 Cabinet Dimensions...
  • Page 74 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-21: Gantry Dimensions: - Side View Illustration 2-22: Gantry Dimensions: - Top view 1 System Components...
  • Page 75 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-23: Gantry Dimensions: - Front view Chapter 2 Equipment Requirements...
  • Page 76 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-24: Omega IV Compact Patient Table Dimensions 1 System Components...
  • Page 77 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-25: Omega IV Compact Patient Table Interference Regions (Innova IGS 520) Chapter 2 Equipment Requirements...
  • Page 78 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-26: Omega V Long Patient Table Dimensions 1 System Components...
  • Page 79 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-27: Omega V Long Patient Table Interference Regions Chapter 2 Equipment Requirements...
  • Page 80 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-28: Tilting table Dimensions 1 System Components...
  • Page 81 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 NOTE: With a Tilting table, the minimum distance from table pivot to the medical Gas Box is 600 mm and the maximum dimensions of the medical Gas Box are : •...
  • Page 82 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-29: Tilting table Interference Regions 1 System Components...
  • Page 83 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-30: Patient Table side clearence (CPR access) Chapter 2 Equipment Requirements...
  • Page 84 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-31: Table Head Extender 1 System Components...
  • Page 85 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-32: Gantry and Omega IV Compact Patient Table Relative Positions (Innova IGS 520) - side view Chapter 2 Equipment Requirements...
  • Page 86 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-33: Gantry and Omega V Long Patient Table Relative Positions - side view 1 System Components...
  • Page 87 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-34: C2 Cabinet Dimensions Chapter 2 Equipment Requirements...
  • Page 88 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-35: C1 Cabinet Dimensions 1 System Components...
  • Page 89 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-36: Dimension Diagram for X-Ray Tube Chiller 4100 NOTE: Required floor space depends on ambient room temperatures. When in doubt, allow for maximum floor space. Chapter 2 Equipment Requirements...
  • Page 90 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-37: Floor Space Diagram for X-Ray Tube Chiller 4100 NOTICE The chiller cannot be located more than 3 m (10 ft) below the detector 1 System Components...
  • Page 91 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-38: Detector Chiller Thermo-Con Dimensions & Orientation Chapter 2 Equipment Requirements...
  • Page 92 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-39: Fluoro UPS UL Layout (Optional) The left, right or back side of the UPS cabinet can be positioned against the wall. The front side of the UPS cabinet must be accessible for maintenance operation. In front of the cabinet, the clear width of the service area to insure electrical safety shall be at least 0.9 m.
  • Page 93 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-40: Fluoro UPS CE Layout (Optional) To improve the ventilation and to accommodate the maintenance operations for UPS and battery, a minimum distance of 200 mm (8”) from the wall is required. The right and left side of the UPS cabinet must be accessible for maintenance operation.
  • Page 94 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-41: UPS IF Box (Optional) 1 System Components...
  • Page 95 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-42: 3 kVA UPS NOTE: The height of the 3 kVA UPS is 325 mm (12.79 in) without feet. Chapter 2 Equipment Requirements...
  • Page 96 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-43: EMI Filter Box (CE Only) 1 System Components...
  • Page 97 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-44: Gas box outlets Omega IV NOTE: With a Tilting table, the minimum distance from table pivot to the medical Gas Box is 600 mm and the maximum dimensions of the medical Gas Box are : •...
  • Page 98 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-45: Gas box outlets Omega V NOTE: With a Tilting table, the minimum distance from table pivot to the medical Gas Box is 600 mm and the maximum dimensions of the medical Gas Box are : •...
  • Page 99 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-46: DL Keypad Dimensions Illustration 2-47: DL Image Monitor Dimensions Chapter 2 Equipment Requirements...
  • Page 100 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-48: VCR Mounting Holes Location feet on the bottom. Use these four holes to fix the VCR. The NOTE: Unscrew the four mounting holes accommodate M3 x 10 mm screws. The holes are 0.4 inches (10 mm) deep.
  • Page 101 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-49: ECG Acquisition Device Modules Chapter 2 Equipment Requirements...
  • Page 102 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-50: Large Display cabinet dimensions (Optional) 1 System Components...
  • Page 103 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-51: Large Display suspension dimensions (Optional) Chapter 2 Equipment Requirements...
  • Page 104: Basic Innova System Compatibility

    Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 1.4 Basic Innova System Compatibility 1.4.1 System Compatibility Cross-Reference - Gantry PRODUCT NAME MODEL NUMBER PRE-INSTALLATION DOCU‐ NOTES MENT NUMBER LC ANGIO POSITIONER 2341727 This document HV cable 24 m LC COMBO POSITIONER 2396420...
  • Page 105: Control

    VCIM 2340519 DL Liberty user interface 2345576 This document C1 cabinet 2335129-8 This document Innova IGS 520, 21 cm Revolu‐ tion Digital detector 5127984 This document Innova IGS 530, 31 cm Revolu‐ tion Digital detector 2359035 This document Innova IGS 540, 41 cm Revolu‐...
  • Page 106 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Eizo 19” LCD HB color monitor sm 5499528-1-8EN RX150 GE 1 System Components...
  • Page 107: Optional Innova System Components

    Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 1.5 Optional Innova System Components 1.5.1 System Compatibilities Cross-Reference - Omega Table Sub-System PRODUCT NAME MODEL NUMBER PRE-INSTALLATION DOCU‐ CATALOG NUMBER MENT NUMBER SMART HANDLE 5408842 This document SMART BOX...
  • Page 108: Room Layouts

    Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 2 Room Layouts 2.1 Room Dimension Requirements WARNING LOCATION IN TECHNICAL ROOM FOR ELECTRICAL CABINETS IS MANDATORY. THE ELECTRONIC CABINETS (C1, C2, OPTIONAL LD CABINET, COOLIX 4100 CHILLER, DETECTOR CHILLER AND FLUORO UPS WHEN INSTALLED ) INCLUDE FANS THAT ARE CREATING AIR–CIRCULATION OF PULSED–AIR.
  • Page 109 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Table 2-4: Omega V Table Configuration Minimum Room length in mm (inches) without Head Extender with Head Extender 6036 (237.6) 6036 (237.6) + 150 (6) Angio 1278 mm (50.3 in) 6036 (237.6) 6036 (237.6) + 150 (6)
  • Page 110 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-52: Minimum Room Length Dimension 2 Room Layouts...
  • Page 111: Room Layout Drawings

    Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 2.2 Room Layout Drawings 2.2.1 Patient Room Layout 2.2.1.1 Room Layout for Innova System (examples of room layouts) WARNING LOCATION IN TECHNICAL ROOM FOR ELECTRICAL CABINETS IS MANDATORY.
  • Page 112 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-53: Monitor suspensions rails parallel to patient table (to side of table) WARNING ILLUSTRATION 2-53 IS MANDATORY WITH HOSPITAL HAVING AIR FLOW INSTALLATION 2 Room Layouts...
  • Page 113 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-54: Monitor suspensions rails parallel to patient table (over table) Chapter 2 Equipment Requirements...
  • Page 114 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-55: Monitor suspensions rails perpendicular to patient table (Gantry side) 2 Room Layouts...
  • Page 115 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-56: Monitor suspensions rails perpendicular to patient table (table side) Chapter 2 Equipment Requirements...
  • Page 116: Room Layout For Innova System (Example With Mobile System 9800)

    Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 2.2.1.2 Room Layout for Innova System (example with Mobile system 9800) Illustration 2-57: Room Layout for Innova System 2 Room Layouts...
  • Page 117 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Legend of Illustration 2-57 1. Gantry 2. Patient table 3. XT monitor suspension 4. 21” (53 cm) FFD Monitors put on frame with monitoring screen 5.
  • Page 118 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 2.2.2 Control Room Layout Illustration 2-58: 2 Room Layouts...
  • Page 119 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 2.2.3 Technical Room Layout For the service access and ventilation restraint dimensions below, see Dimension Drawings Room Layout Considerations. NOTE: The phone outlet must be located less than 1 meter (3 feet) from the C1 Cabinet. Illustration 2-59: Chapter 2 Equipment Requirements...
  • Page 120 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 2.2.4 IVUS Room Layout 2.2.4.1 IVUS Rev 2 and Rev 2+ Illustration 2-60: NOTE: (For OR Table) the CONTROL STATION shall be installed on a cart. 2.2.4.2 IVUS Rev 3 Refer to Volcano s5i Imaging System Integration in IGS Systems - Service...
  • Page 121 Innova system: 2.2.5.1 Installation Configuration #1: For ECG devices installed in Control Room (such as GE ECG like MacLab, CardioLab or ComboLab) Both the Physio & Hubican modules are installed in the Control Room as shown below: Illustration 2-61: 2.2.5.2 Installation Configuration #2:...
  • Page 122 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-62: 2 Room Layouts...
  • Page 123 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 2.2.6 Vivid E9 Ultrasound Layout On the application defined Vivid E9 positions of the Structural Heart procedures, the 3.5 m power connection cord limits the location of the power access point. For the minimal perturbation of the layout, the 3 Vivid E9 connection cables (power, network and the LDM image link ) shall run close to each other, and shall be directed to a connection area on the head end wall of the patient room as close to the floor as it is possible, in order to keep the cables on the...
  • Page 124 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 2.3 Room Layout Considerations 2.3.1 Radiation Protection Because X–Ray equipment produces radiation, you may need to take special precautions or make special site modifications. The General Electric Company does not make recommendations regarding radiation protection.
  • Page 125 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 During the X-Ray operation, the Vascular gantry movements also limit the Vivid E9 positions, and a dedicated back out path has to be cleared for Vivid E9 park position. 2.3.4 Peripheral Equipment Consult hospital personnel regarding additional space requirements for the following types of hospital equipment:...
  • Page 126 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 ○ Patient Interface Module(s). • Table Side User Interfaces (TSUI) ○ ISmart Box ○ Table Side Status Control ○ Table Panning Handle ○ NOTE: No IPX4 components shall be installed on a cart. •...
  • Page 127 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 WARNING THE HEAD HOLDER SHALL NOT BE USED TO RESTRAIN THE PATIENT'S HEAD WHEN THE TABLE IS TILTED, ONLY WHEN IT IS HORIZONTAL. IT IS NOT DESIGNED TO SUPPORT THE HEAD FROM MOVING AND MAY CAUSE INJURY TO THE PATIENT.
  • Page 128 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-64: CAUTION The service area dimensions shown above are minimum requirements. Service areas must comply with local regulations if more stringent. 2.3.8 Layout Constraints NOTICE The X-Ray tube Chiller elevation cannot exceed 3 meters (9.84 feet) in height.
  • Page 129 GE Healthcare’s Customer is responsible for the structural analysis and mounting of the base plates. If GE Healthcare is forced to mount the base plate, the LCT must hire a structural engineer to design and approve the mounting method and provide GE Healthcare with an engineering report.
  • Page 130 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 NOTICE The gap between the Table Foot bottom end the Gantry Baseplate bottom end shall be lower than 20 mm (0.97 in). Any bigger gap would make the system incompatible with the Innova Vision Applications.
  • Page 131 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 3.2 Floor Requirements 3.2.1 Floor requirements when using provided floor anchors The maximum pullout force per provided anchor was calculated assuming: • A concrete compression strength of 17.24 MPa at 28 days (which is the minimum required compression strength).
  • Page 132 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-69: Table floor mounting layout NOTICE Prepare the floor such that the table base plate will be flush with the floor finish surface, taking into account the thickness of the floor finish material. Innova For alternative table bolts or seismic area, refer to template drawing shown in Illustration LC Positioner And Table Floor Mounting Template contained in...
  • Page 133 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-70: Hole location in concrete floor Chapter 2 Equipment Requirements...
  • Page 134 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Table 2-7: D distance ANGIO CARDIO NEURO Omega IV Compact 1395 mm (54.9 in) Omega V Long 1278 mm (50.3 in) 1395 mm (54.9 in) 1395 mm (54,9 in) Omega V non motorized Long 1278 mm (50.3 in)
  • Page 135 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-71: Water Conduit location with “Above Grade” anchor kits NOTICE In some countries, depending on local regulations, it may be forbidden to run electrical cables and water pipes in the same conduit. In this case, two separate conduits are required.
  • Page 136 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-72: Water Conduit location with “On Grade” anchor kits Illustration 2-73: Cable Curvature with “Above Grade” anchor kits 3 Room Structural Requirements...
  • Page 137 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-74: Cable Curvature with “On Grade” anchor kits NOTE: In case of on grade cabling, because of the minimum curvature constraint of 150 mm (6”), the cable will have to come from the side between Innova LC Positioner and patient table.
  • Page 138 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 3.3 Mounting Requirements 3.3.1 Floor Loading and Recommended Mounting Methods Table 2-8. To obtain floor loading and recommended mounting methods for components not specified in Table 2-8, refer to the appropriate component Pre-Installation Manual listed in Basic Innova System...
  • Page 139 Large Display See illustration sion dimensions (Optional) in monitor suspen‐ 356 (784) Dimen‐ Ceiling-mounted sion (See NOTE 5) sion Drawings IVUS S5I GE Rev 2 Option PIM (Ultrasound 2 (4.4) probe) Control Station 4 (8.8) Joystick 1 (2.2) Isolation trans‐...
  • Page 140 176 x 522 S5I GE CPU 18 (39.7) 522 (20.55) 176 (6.93) 430 (16.93) 2.0 kg/m On floor (6.93 x 20.55) S5I GE Keyb. & 1 (2.2) On desk mouse pad 330 x 247 S5I GE monitor 9 (19.8) 330 (12.99) 247 (9.72) 596 (23.46) 1.1 kg/m...
  • Page 141 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Table 2-9: Title Illustration Illustration 2-75 Illus‐ Gantry Floor Mounting Methods tration 2-76 Inner Base Plate (Water Electric Separator) For Above Grade Floor Anchor Kit Illustration 2-77 Cable Conduit For On-Grade Floor Anchor Kit Illustration 2-78...
  • Page 142 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-75: Gantry Floor Mounting Methods (1/2) 3 Room Structural Requirements...
  • Page 143 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-76: Gantry Floor Mounting Methods (2/2) Chapter 2 Equipment Requirements...
  • Page 144 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-77: Inner Base Plate (Water Electric Separator) For Above Grade Floor Anchor Kit 3 Room Structural Requirements...
  • Page 145 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-78: Cable Conduit For On-Grade Floor Anchor Kit Chapter 2 Equipment Requirements...
  • Page 146 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-79: Fixing Bolt Overview NOTE: For more details on Floor plate or Base plates, refer to Illustration 2-80 and illustration Hole location in concrete floor in Floor Requirements.
  • Page 147 Pull out efforts and recommendations about chemical anchors not provided by GE. The following table provides the recommended chemical anchors for Table/Gantry base plates and for the floor plate ordered locally that they could be used instead of bolts provided by GE. Table 2-10: Chemical anchors Pull out efforts and recommendations...
  • Page 148 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-80: Gantry and table mounting holes 3 Room Structural Requirements...
  • Page 149 3.3.3 Gantry and Omega table/Tilting table Floor Preparation Kits (GEMS supplied) Illustration 2-81: All GE supplied vascular system floor preparation are contained in catalog. There are some additional gantry/table mounting kits based on each mounting method: through bolts or floor anchors.
  • Page 150 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 • Assembly and separation (Select kit) 2285050 - Refer to tables, Section 3.3.3.3 ○ Above grade, Floor anchors without insert – 2285051. ○ Through bolts without insert – 2285646. ○...
  • Page 151 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 3.3.3.1 2285048 - Base Plate Assembly Table 2-12: Item Name Part # Description Quan. Notes Base plate 2285059 12 Hole Floor Base plate 1 pc Lift Plate 2290939 Lifting Straps...
  • Page 152 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 3.3.3.2 2360538 – Tilting table Add–On Kit Table 2-13: Item Name Part # Description Quan. Notes Omega ta‐ 2361993 Plate to be anchored under the ta‐ 1 pc ble/ Tilting table com‐...
  • Page 153 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 3.3.3.3 2285050 – Assembly and separation select kit 3.3.3.3.1 2285051 – Above grade, Floor Anchors without insert Table 2-14: Item Name Part # Description Quan. Notes Floor An‐...
  • Page 154 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 3.3.3.3.3 2286398 – On grade, Floor anchor Table 2-16: Item Name Part # Description Quan. Notes Floor An‐ 46-302265P1 Floor Anchor Bolt; 5/8-10 x 6 in. 16 pc On Grade anchor mounting chor...
  • Page 155 Screw M16x40x40 Inox A4-80 10 pc 4 only are used for Omega Pass NOTE: The kit 2285632 Floor plate / Base plate assembly is needed when the floor plate is installed. This part isn’t GE part. It’s ordered locally. Chapter 2 Equipment Requirements...
  • Page 156 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 3.3.3.4 2285054 – Templates 3.3.3.4.1 2285056 - A/C/N Templates Table 2-20: Item Name Part # Description Quan. Notes A/C/N Tem‐ 2185979 LC Positioner and Omega C/N Pa‐ 1 pc Illustration 2-80 plates...
  • Page 157 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-82: Base plate plan 2359194 - Angio and Cardio configurations Chapter 2 Equipment Requirements...
  • Page 158 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 3.3.4 Injector Mounting Requirements CAUTION Table accessory rail load consideration: The maximum load per table accessory rail is 40 kg (88 lbs) at 150 mm (0.49 ft) (60 N.m or 44.25 ft/lbs).
  • Page 159 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 3.3.5.3 Specific Prerequisites for Installation Configuration #2 Verify the following items only in the case of Installation Configuration #2 with ECG Device in Exam Room: 1.
  • Page 160 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 3.4 Ceiling Requirements Aluminum rails support the In-Room Monitor bridge used in Innova system X-Ray rooms. Reference: For additional details on ceiling requirements for stationary rails, refer to: - Direction 46-019639, Advantx (VHLA) XT Stationary Rails Installation and Adjustment .
  • Page 161 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-83: SPECIFICATIONS FOR A TYPICAL 17’-10” (5.44 M) INBOARD STATIONARY RAIL MOUNTING INTERFACE (BOTH RAILS CEILING MOUNTED), FOR MAVIG SUSPENSION Table 2-22: Stationary rail in different length Rail length mm (ft) INBOARD RAILS 4,724 (15’6”)
  • Page 162 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 3.4.4 Boom Mounted for Monitor Suspension The purchaser/customer should prepare a Unistrut/ Halfen or equivalent ceiling structure. The distance between center lines of the two longitudinal rails shall be 660.4 mm (26"). One type of boom mounted is supplied as counterpoised monitor suspension not adjusted on site as follows: - for 2 x 21”...
  • Page 163 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 3.5 Wall Requirements The C1 Cabinet, the C2 Cabinet, the PDB Cabinet and LD cabinet (optional) must be securely fastened to the wall to prevent them from tipping. The optional LD secondary monitor outside the patient room is mounted on wall.
  • Page 164 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 4 Mounting Data, Including Seismic 4.1 Seismic areas 4.1.1 Floors Chiller: The seismic kit for the Chiller 4100 (SMC) is included with the chiller when shipped. LD cabinet: the seismic kit to fasten the LD cabinet to the wall is included with the cabinet when shipped.
  • Page 165 Seismic attachment hardware shown on seismic calculations may differ from hardware supplied with system. Any additional hardware that is required will be the responsibility of the institution and/or their contractor. Contact your local GE Installation Program Manager to obtain seismic calculations.
  • Page 166 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-87: C1 cabinet Illustration 2-88: C2 cabinet 4 Mounting Data, Including Seismic...
  • Page 167 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-89: Fluoro UPS UL Illustration 2-90: Fluoro UPS CE Chapter 2 Equipment Requirements...
  • Page 168 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-91: PDB UL Illustration 2-92: PDB CE 4 Mounting Data, Including Seismic...
  • Page 169 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-93: Tube Chiller Coolix 4100 Illustration 2-94: Detector Conditioner Chapter 2 Equipment Requirements...
  • Page 170 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-95: Omega IV Table Illustration 2-96: Omega V Long Table 4 Mounting Data, Including Seismic...
  • Page 171 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-97: Tilting table Chapter 2 Equipment Requirements...
  • Page 172 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-98: Gantry 4 Mounting Data, Including Seismic...
  • Page 173 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-99: 3 kVa UPS (LD option) Illustration 2-100: 19” Desk Mounted Monitor Chapter 2 Equipment Requirements...
  • Page 174 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-101: Mavig Overhead Flat Panel Suspension 4 Mounting Data, Including Seismic...
  • Page 175 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-102: Large Display Cabinet (LD option) Chapter 2 Equipment Requirements...
  • Page 176 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-103: Large Display Monitor suspension (LD option) 4 Mounting Data, Including Seismic...
  • Page 177 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 2-104: Large Display secondary monitor Swing out arm (LD option) Chapter 2 Equipment Requirements...
  • Page 178 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 This page left intentionally blank. 4 Mounting Data, Including Seismic...
  • Page 179 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Chapter 3 Special Construction Requirements 1 Radiation Protection Because x-ray equipment produces radiation, special precautions may be needed or special site modifications may be required. The General Electric Company does not make recommendations regarding radiation protection.
  • Page 180 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 2 EMI Consideration IEC60601-1-2 Electromagnetic Standard Compliance & Documentation The information contained in this section is also found in the Innova system Operator Manual. 2.1 General Scope This equipment complies with IEC-60601-1-2: Edition 2.1 and Edition 3 standard for medical devices The Innova system is suitable to be used in the electromagnetic environment, as per the limits &...
  • Page 181 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Table 3-2: Immunity Test IEC 60601-1-2 Test Compliance Level Electromagnetic Environment Level Electrostatic dis‐ +/-6 kV contact +/-6 kV contact Floors are wood, concrete or ceramic tile or floors are covered charge (ESD) +/-8 kV air +/-8 kV air...
  • Page 182 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 The Innova system is intended for use in the electromagnetic environment specified below. The customer or the user of the Innova system should assure that it is used in such an environment.
  • Page 183 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 ISM Frequency (MHz) Field Level Modulation 433.920 (ISM) 915 (ISM) 1440 3 V/m 80 % AM at 1 kHz rate 1750 1920 2450 (ISM) NOTE: : Industrial, Scientific and Medical (ISM) radio bands.
  • Page 184 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 However, when installed and operated as specified herein, the system will maintain its essential performance by continuing to acquire, display, and store diagnostic quality images safely. For example, a 1W mobile phone (800 MHz to 2.5 GHz carrier frequency) shall be put 2.3 meters (see (*) Table...
  • Page 185 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 2.6 Installations Requirements & Environment Control NOTICE In order to minimize interference risks, the following requirements shall apply: • Cables shielding & grounding: All interconnect cables to peripheral devices must be shielded and properly grounded.
  • Page 186 The maximum distance between the EMI filter and the PDB is 3 m. NOTE: The Fluoro UPS (CE) option requires an EMI filter box placed upstream the PDB. The EMI filter box will be delivered as a single product by GE Consumer & Industrial. 2 EMI Consideration...
  • Page 187 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Chapter 4 Environmental Requirements 1 Relative Humidity and Temperature NOTICE Avoid extremes in temperatures Innova system room climate requirements – relative humidity and temperature (individual products or components are classed by their installation area for In use see -Table 4-1)
  • Page 188 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 IVUS Option : 1. IVUS Rev 2 and Rev 2+ Table 4-2: Humidity (non condensing) : 30 - 80 % Temperature (at 20% Relative Humidity ) : 5 - 35 C (41 - 95 F) 2.
  • Page 189 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 2 Altitude and Atmospheric Pressure Refer to Table 4-3. To obtain altitude and atmospheric pressure requirements for components not specified in Table 4-3, refer to the appropriate component Pre–Installation Manual listed in Chapter 2, Basic Innova System Compatibility.
  • Page 190 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 3 Heat Output 3.1 Equipment Heat Output tables Refer to Table 4-4. To obtain heat output information for components not specified in Table 4-4, refer to the appropriate component Pre–Installation Manual listed in Chapter 2, Basic Innova System Compatibility.
  • Page 191 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 NOTE: (2) For more details, consult appropriate pre–installation manual NOTE: (3) For UPS 20 kVA option refer to Section 3.2 NOTE: (4) Moderate use: 8 cases / 10 hours, typical use: 11 cases / 10 hours, maximum use: during case.
  • Page 192 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 The size of the air inlet has therefore to be dimensioned accordingly. Contact your Local Distributor or one of the Service Centre, which will help you to find valuable solutions.
  • Page 193 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 NOTE: Refer to the UPS vendor Service Operating Manual for more details: • UL Model: UPS Option - Operating Manual SG-UL Series - 20 kVA •...
  • Page 194 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 these filters can cause reduced speed at the air inlet. The size of the air inlet has therefore to be dimensioned accordingly. NOTICE If installed in the technical room, the Large Display Cabinet may impact its layout. It may also be installed in a separate technical room.
  • Page 195 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 4 Acoustic Specifications • Less than 50 dB (A) at 1 meter for Gantry. • Limited to 50 dB (A) at 1 meter for Omega IV, Omega V tables. •...
  • Page 196 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 This page left intentionally blank. 4 Acoustic Specifications...
  • Page 197 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Chapter 5 Electrical Requirements 1 Power Requirements 1.1 System Electrical Characteristics NOTICE It is the customer’s responsibility to ensure that electrical installation is compliant to local regulations.
  • Page 198 Single phase (± 3 Hz) 100-127 V / 200-240 V 50 Hz or 60 Hz 11 A / 5.5 A Single phase S5I GE 100-120 V / 230 V 50 Hz or 60 Hz 400 VA Single phase WARNING PRIOR TO EACH INSTALLATION, ENSURE THAT THE ECG POWER CABLE...
  • Page 199 The power cords between the peripherals in the control room and the power strip are not shown on the following diagram. All are single phase power cords (with phase, neutral line and protective Earth) and are provided by GE. 1.2.1 Schematic...
  • Page 200 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 1.2.2.1 Core system Table 5-3: Core system cables Cable # From Type Characteristics Comments Diameter in conformity with 3 phases the Max Line Impedance ta‐ Diameter in conformity with Mains Neutral...
  • Page 201 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 1.2.2.2 For the Fluoro UPS option Table 5-5: Fluoro UPS option cables Cable # From Type Characteristics Comments Diameter in conformity with 3 phases the Max Line Impedance ta‐ Diameter in conformity with Only with UL Fluoro UPS op‐...
  • Page 202 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Cable # From Type Characteristics Comments UL: AWG10 (5.26 mm²) Phase, neutral and Mains LDM UPS ground CE: AWG12 (2.5 mm²) 1.2.3 LOTO devices provided by the installer 1.2.3.1 Core system A wall circuit breaker or equivalent device with LOTO capability must be installed on the mains line to the PBD.
  • Page 203 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 CE system: The terminals 2 & 3 from the Positioner Bulkhead of the C2 cabinet shall be connected to PDB terminals BNC 12 & 13 (cable provided by the installer, maximum length 24m, minimum diameter 1.5 mm²).
  • Page 204 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 • The creepage distances and air clearances shall comply with the following table: Table 5-7: System Between phase and ground and be‐ Between phase, Neutral and external Between phase and Neutral type tween neutral and ground...
  • Page 205 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 1.3 Other Options 1.3.1 For USA only A purchasable option I-sense (catalog number E4504B) allows the monitoring of the hospital main power line. It is recommended to install this option everywhere RMS and waveform variation events can impact the standard behavior of the system.
  • Page 206 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 2 System Cable Information 2.1 Physical Runs 2.1.1 Physical Run Synoptic Illustration 5-3: Mini / Maxi Interconnection Length (system with LCD monitors) NOTE: A 24 m Emergency UPS stop cable 27561A is available in FRU. 2 System Cable Information...
  • Page 207 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 5-4: Mini / Maxi Interconnection Length (system with Large Display Monitor option) NOTICE Radius of curvature of the cables that connect the 8MP Monitor: min 35 mm. Chapter 5 Electrical Requirements...
  • Page 208 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 2.1.2 System Core Matrix NOTICE All lengths of cable are: • in useable meter when you look at group level, or • in meters (connector to connector) when you look at the cable level. For a description of how to use the following cable group schematics, see below: Illustration 5-5: Example of cable group schematic Cable length data is as follows:...
  • Page 209 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 5-6: Cables Group 1 and Ground Cables Group – From Technical Area to Exam Room Chapter 5 Electrical Requirements...
  • Page 210 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 5-7: Cables Group 2 – From Technical Area to Control Room 2 System Cable Information...
  • Page 211 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 5-8: Fast Link Cable Group - Option Chapter 5 Electrical Requirements...
  • Page 212 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 5-9: Cables Group 3 – From Technical Area to Technical Area 2 System Cable Information...
  • Page 213 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 5-10: Cable Group 4-2 - From Technical Area to Exam room ceiling Illustration 5-11: Cable Group 4-3 - From Technical Area to Exam room ceiling Chapter 5 Electrical Requirements...
  • Page 214 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 5-12: Cables Group 4-4 with LCD Mavig Suspension For suspension cabling, two length options are available : 24 meter or 36 meter (78 ft 9 in or 118 ft 1 in).
  • Page 215 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 5-13: Cable Group 4-5 - From Technical Area to Exam Room ceiling Chapter 5 Electrical Requirements...
  • Page 216 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 5-14: Fluoro UPS Cable Set - Option NOTE: (1): provided in Group 3 (27678A). 2 System Cable Information...
  • Page 217 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 2.1.3 Physical Run - System Core Detail Table 5-8: MIS num‐ Cable As‐ UL Style Voltage rating Max Voltage Cable di‐ Connector type Bigger Plug sembly carried (V) ameter...
  • Page 218 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 MIS num‐ Cable As‐ UL Style Voltage rating Max Voltage Cable di‐ Connector type Bigger Plug sembly carried (V) ameter size (mm) (mm) 27495A 5305381 1354 BNC 75 ohms 14.3...
  • Page 219 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 MIS num‐ Cable As‐ UL Style Voltage rating Max Voltage Cable di‐ Connector type Bigger Plug sembly carried (V) ameter size (mm) (mm) 28047 2385191 28048 2385192 28049...
  • Page 220 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 MIS num‐ Cable As‐ UL Style Voltage rating Max Voltage Cable di‐ Connector type Bigger Plug sembly carried (V) ameter size (mm) (mm) 28187 2385198 Power 28189 2385200...
  • Page 221 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 2.2 MIS (Master Interconnect System) Innova system interconnect cables are described in MIS (Master Interconnect System) documents. These documents specify all interconnections between components within the system.
  • Page 222 Control Booth cables between Equipment Room and Control Room. Illustration 5-15: IVUS rev 2 wiring 2.3.2 For S5I GE rev 2+ wiring The connection cables between the S5I GE Control Room and Procedures Room components run in a dedicated under floor conduit (see Cable Channeling).
  • Page 223 IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 A GND wire between the C1 cabinet and the S5I GE Rev 2+ Isolation Transformer runs in the underfloor conduit prepared for the Control Booth cables between Equipment Room and Control Room..
  • Page 224 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 2.4 Cable Channeling 2.4.1 General High voltage and power cables must be separated from other cables. Use a separate trough in the duct system, or use a separate conduit. Minimize cable length between the line disconnect and the System Cabinet power unit to reduce voltage regulation problems and wiring costs.
  • Page 225 & PC are installed • If no GE Maclab cable 2016134-106 installed between the TRAM (Exam Room) and the Control Room, need to route it so that installation/connection of Physio module can be made in Control Room.
  • Page 226 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 5-19: Cable Channeling Layout NOTE: 18 meters (59 ft) is the only cable length available for the Remote TSUI box data cable connecting remote TSUIs in the control booth and the patient table. 2 System Cable Information...
  • Page 227 Locate the position of the S5I GE components in the Control Room, according to the workflow requirement of the operators. • Establish a 76 mm (3”) conduit below floor from the S5I GE CPU location in the Control Room, to the foot end area of the Patient Table in the Procedures Room: ○...
  • Page 228 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 2.4.4.2 Run Interconnection cables through the Conduit 2.4.4.2.1 For IVUS Rev 2 option Use the IVUS Rev 2 PIM (#806365002) as reference to the procedures below. •...
  • Page 229 Make sure, that the out of conduit parts of the cables with the connectors are properly protected during the rest of the installation 2.4.4.2.2 For S5I GE Rev 2+ option Use the S5I GE Rev 2+ PIM (#806365026) as reference to the procedures below. • Unpack , and check the completeness of the Preinstallation kit: ○...
  • Page 230 Ensure sufficient cable length on the Procedures Room end of the cables according to the table below Cable Free length from the conduit exit S5I GE PIM cable * 2.4 m (8’) FFR cable for Pimette ** 0 m (0’) Control Station cable 2.1 m (7’)
  • Page 231 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 3 Lighting Specifications 3.1 Room Light Distribution 3.1.1 Requirements for lighting Requirement for lighting concern the following, general, light–technique characteristics: • Illuminator level. • Lighting distribution. •...
  • Page 232 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 3.1.3 Room Lighting for System Illustration 5-20: Room Lighting Zones 3.1.4 Windows and curtains When the examination room has a window with an aperture outside of the controlled light area (day light, other...) a curtain has to maintain the light intensity under a limit fixed to 150 lux.
  • Page 233 Volcano s5i Imaging System Integration in IGS Systems - Service Manual. The S5I GE CPU located in the control room shall be connected to the hospital Ethernet network. A wall Ethernet outlet shall be available in the control room to connect the device. DICOM Image Storage: Saving patient cases to DVD: The archived images are stored in DICOM format with Volcano s5i system acting as a File Set Creator (FSC), following the guidelines in the 2004 DICOM 3.0...
  • Page 234 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 • 10/100/1000 Mbit Interface 1.2.2.3 VIVID E9 Purpose of the DICOM network function: DICOM services provide the operator with clinically useful features for moving images and patient information over a hospital network.
  • Page 235 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 2 IP Addressing Process To obtain an IP address, contact the following for your pole: • GEMSAM: Contact: OnLine Center–Americas, Network Products and Services (NP&S) Telephone: 1–800–321–7937 NOTE: Press [1] for the Online Center.
  • Page 236 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 This page left intentionally blank. 2 IP Addressing Process...
  • Page 237 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Chapter 7 Appendix 1 PIST0021 - US Fluoro UPS and PDB connection 1.1 Personnel Requirements Personnel Preliminary Reqs Procedure Finalization Requirements Not Applicable 1 hour Not Applicable 1.2 Preliminary Requirements 1.2.1 Tools and Test Equipment...
  • Page 238 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 1.3 Procedure 1.3.1 Connection of the Power Distribution Box NOTICE Hospital main power must enter from the Top. Correct phase sequence during installation is very important. Phase failure relays are provided to monitor voltages less than 85%, loss of a phase, or incorrect phase sequence.
  • Page 239 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 7-3: 1.3.2 Connection of the Fluoro UPS Option Cables 1. Connect the Fluoro UPS power cables. Remove the protection panel Chapter 7 Appendix...
  • Page 240 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 7-4: Remove the protection panel ( Illustration 7-4). Remove the metallic window from the protection panel (Illustration 7-4). d. The power cable shall enter the UPS by the bottom right hand side entry: Remove RHS UPS cover (remove two top screws make two bottom screws loose), ii.
  • Page 241 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 Illustration 7-5: iii. Drill a hole in the bottom cover plate for cable access. WARNING RISK OF ELECTRIC SHOCK HIGH VOLTAGE PRESENT THE FLUORO UPS REQUIRES COMMON POWER INPUT FOR RECTIFIER AND BYPASS .
  • Page 242 Innova IGS 520, Innova IGS 530, Innova IGS 540 Pre-Installation Manual Direction 5499972-1-1EN, Revision 2 • Rectifier neutral input to X1-N-rectifer terminal, • Rectifier ground to X1-PE-rectifier terminal. g. Connect the Fluoro UPS output cable from PDB power block 2 as follows (Illustration 7-6): •...
  • Page 244 © 2015 General Electric Company. All Rights Reserved. General Electric Company 283, rue de la Minière 78530, Buc FRANCE www.gehealthcare.com...

This manual is also suitable for:

Innova igs 530Innova igs 540

Table of Contents