2. If the model for the serial number to be
entered is shown, click on it to select it
and click OK.
To add a new model, click the "New" button
and refer to section 6.4 for instructions on
adding a new model.
3. Enter the serial number and other
specific information about the apparatus.
Some information shown on the serial
number screen can be affected by the
options chosen in the Setup / Options
screen. The options can be displayed and
changed by using the Actions button at the
bottom of the screen.
Required fields are marked in red. The
Serial Number field is always mandatory.
The list of other required fields can be
changed in the options menu.
The display of auxiliary IDs and overhaul
dates is also controlled in the options screen.
See the Posi 3 USB Help File for further
details.
4. To enter another serial number that is
identical in configuration to the first one,
click on "Retain entries when adding
another serial number" at the bottom,
then click on "Add Another". The
software will create a new Enter Serial
Number Information screen while
retaining the other information that has
been entered.
5. To enter another apparatus that has a
different configuration from the first one,
click "Add Another" and leave "Retain
entries...." unchecked.
6. Once all of the apparatus have been
entered, click "Accept" at the lower right.

6.6 Import Serial Numbers

For those who have been using the
PosiChek3 Titanium Edition 32-Bit Software,
it is possible to import apparatus information
from the existing MS Access database into
the new PostgreSQL database. To start the
import, click on the File Menu followed by
Import / Serial Numbers from Access
Database.
The Access Import Wizard screen will be
shown.
23