P R E L I M I N A R Y
GE
P
N
FN091065, R
ART
UMBER
Table 8-3
Step
19.
Navigate to the TCPIP tab and press the Set button on
the control panel.
20.
Navigate to the Save Settings button and press the Set
button on the control panel.
21.
Click OK and re-boot the system.
22.
If this is a new software version, an adhesive Rating Plate
is supplied with the software CD.
Adhere the plate in the appropriate position at the rear of
the system.
23.
Proceed to perform the following functionality tests:
"Software Configuration Checks" on page 4-45
• "Calibration Tests" on page 7-27
• "Full System Test" on page 7-11
2
EVISION
Vivid™ S5 N and Vivid™ S6 N Software Re-Installation (cont'd)
Instruction
Chapter 8 - Replacement Procedures
Expected Result
Connectivity dialog box - TCPIP tab activated:
A message is displayed prompting for system re-boot:
VS5 N
VS6 N