Accessing the Full System Diagnostic Options
In 2D-Mode, press <ALT +Config> on the alphanumeric keyboard. The Diagnostic Test window is
displayed, as shown in
7-3-1-1
Full System Test (FE and BEP)
NOTE:
For a description of this test, refer to the
NOTE:
The Full System Test (FE and BEP) performs a full sequence of complete system tests (automatically)
and requires approximately 20 minutes for completion.
1.) In the Diagnostic Test window, trackball to Full System Test (FE and BEP) in the list of tests
available for selection from the Diagnostic Test Tree (see
Test" on page
Execute, and all system components are highlighted blue in the Schematic View, as shown below:
2.) As indicated in the Special Instructions area, make sure the system is calibrated and disconnect
any probe currently connected to the system.
7-22
2
EVISION
Figure 7-1 "Diagnostic Test Window" on page
7-22) and press Set. The Full System Test (FE and BEP) name is displayed under
Figure 7-6 Full System (FE and BEP) Test
Section 7-3 - Performing Full System Diagnostics
7-2-4-1-1 "Full System Test (FE and BEP)" on page
Figure 7-6 "Full System (FE and BEP)
VS5 N
VS6 N
S
AND
ERVICE
7-3.
M
ANUAL
7-11.