P R E L I M I N A R Y
GE
P
N
FN091065, R
ART
UMBER
8-3-8
Rear Handle Replacement Procedure
8-3-8-1
Tools
Appropriate cutting knife for cutting perspex as shown in
RTV 102 Silicone Sealant, P/N USFS-43
8-3-8-2
Preparation
Shut down the Vivid S5 N or Vivid S6 N ultrasound unit, as described in
page
4-7.
8-3-8-3
Rear Handle Removal Procedure
There are two types of rear handle removal procedure, as follows:
8-56
Section 8-3 - Control Console Components Replacement Procedures
2
EVISION
Figure 8-70 Cutting Knife with Pointed Protruding Edge
On older-type Vivid S5 N or Vivid S6 N systems, the rear handles are glued to the aluminium
bars and have to be cut and peeled off the aluminium bars to remove them.
Refer to the instructions in
On newer Vivid S5 N or Vivid S6 N systems, the rear handles are secured with a mounting
screw (see
Figure
8-72). In this instance, the screw is released and the handle slides off the
aluminium bar.
Refer to the instructions in
Removing Handles Glued in
Removing Handles Secured with Mounting
VS5 N
VS6 N
AND
Figure
8-70.
4-2-3 "Power Shut Down" on
Position.
Screws.
S
M
ERVICE
ANUAL