Front Cover Replacement Procedures
8-2-6-1
Tools
Use the appropriate Phillips-type screwdriver, as indicated in the front cover replacement procedures.
8-2-6-2
Preparations
Shut down the Vivid S5 N or Vivid S6 N ultrasound unit, as described in
page
4-7.
8-2-6-3
Front Cover Removal Procedure
1) Remove the left and right side covers, as described in
page 8-6
2) Remove all probes and ECG cables currently connected to the system.
3) Place each of the three RS connector locking mechanisms in the open position (refer to
Figure
Front Cover
RS Connector
Locking Mechanisms
NOTE:
The front cover is secured to the base of the main assembly frame by a screw through a
securing tab on each side of the cover (left and right inner side), as illustrated above.
4) Loosen and remove the two screws from the securing tabs (one each side of the cover - left and
right - inner sides) - see
2
EVISION
and the
"Right Side Cover Removal Procedure" on page
8-18).
Screw
(not visible
in picture)
Figure 8-18 Position of Front Cover and Securing Screws
Figure
Chapter 8 - Replacement Procedures
"Left Side Cover Removal Procedure" on
(through securing tab)
Screw
8-19.
VS5 N
VS6 N
S
AND
ERVICE
4-2-3 "Power Shut Down" on
8-5.
Screw
M
ANUAL
8-17