Honeywell Genesis DM Monitor Programming, Troubleshooting Manual page 27

Hide thumbs Also See for Genesis DM Monitor:
Table of Contents

Advertisement

#
VFD Display
441
ERROR_ECG_COMM
442
ERROR_ECG_OFF
450
PEAKFLOW_TIMEOUT
451
PEAKFLOW_COMM
452
PEAKFLOW_OFF
601
Relay URL Rejected
602
Relay URL Rejected
603
Call Customer Service
604
Call Customer Service
605
Call Interrupted
606
CONNECTION CLOSED
607
PIN NUMBER DENIED
P4520EN.06
GENESIS DM MONITOR
Programming/Troubleshooting Guide
Meaning
ECG cable not connected
ECG disconnected
Loose connection
ECG not assigned a COM
port
User timeout – peak flow
Patient noncompliance
Peak flow cable not connected
Peak flow discontinued
Loose connection
Peak flow not assigned a
COM port
The HomMed Relay is having a
problem and refused to accept a
connection to the monitor.
Remote TCP Server refused
The monitor is not properly
connection
configured for what HomMed
relay servers to use (not a
clinician accessible configuration
item)
The HomMed Relay is having a
problem and fails to respond to
monitor in a timely manner.
Connection failed for
The monitor is not properly
unknown reason
configured for what HomMed
relay servers to use (not a
clinician accessible configuration
item)
The monitor has not been
activated with the ISP by
Honeywell HomMed staff.
User name or password
The ISP configuration has been
was incorrect
modified and is no longer correct.
Internal and external serial
number do not match.
Failed communications with
modem
Communications with
Call waiting is active and an
server failed after
incoming call has interrupted the
connection established
transmission.
Socket was closed by
remote host
The pin number is not correctly
configured on the monitor.
Pin number not recognized
The pin number is not yet
by server
activated on the HomMed relay
server.
6/26/2014
Possible Cause
Resolution/Action
Connect cable.
Make sure all connections are
secure.
In Options menu, set ECG to On.
Retrain the patient.
Connect the cable.
Make sure all connections are
secure
In Options menu, set peak flow
to On.
Add a DSL filter.
If problem persists, call
Honeywell.
Add a DSL filter.
If problem persists, call
Honeywell.
Add a DSL filter. If problem
persists, call Honeywell.
Add a DSL filter. Restart monitor
and watch for any errors during
the power up routine. If problem
persists, call Honeywell.
Add a DSL filter. If problem
persists, call Honeywell.
Add a DSL filter. If problem
persists, call Honeywell.
Verify the pin number is correct.
Add a DSL filter.
If problem persists, call
Honeywell.
27