Purpose of this section
Customer provided prerequisite
Data Management - moving all images
CAUTION
Vivid T8/ Vivid T8 Pro – Basic Service Manual
5490863-100 English Rev. 03
Loading the software
This section describes how to reinstall software on Vivid T8/
Vivid T8 Pro.
•
Formatted and labelled media for Images storage.
•
Formatted and labelled media for Patient Archive and
Presets (User Defined Settings).
•
Password for the user ADM.
An error, or a power loss may occur.
Always backup the Patient Archive and the Presets (System
Configurations) before loading the software!
In order to complete a successful restore of the Patient
Database, as needed after a hard disk replacement, or if all the
content on the hard disk has been erased, the images must be
moved away from Vivid T8/ Vivid T8 Pro before doing backup
of the Patient Database.
Depending on the location set-up, either move the images to a
remote server or to removable media like DVD or CD discs.
•
Move the images to a remote server or to removable media.
For instructions, please see "Disk management" in the User
Manual/User Guide.
Loading the software
8-13