Vivid T8/ Vivid T8 Pro – Basic Service Manual
5490863-100 English Rev. 03

General procedures

Ultrasound system requires all covers.
Operate this Ultrasound system only when all board covers and
frame panels are securely in place. The covers are required for
safe operation, good Ultrasound system performance and
cooling purposes.
Energy Control and Power Lockout for Vivid T8/ Vivid T8 Pro.
When servicing parts of the Ultrasound system where there is
exposure to voltage greater than 30 volts:
1. Follow LOCK OUT/TAG OUT procedures.
2. Turn off the breaker.
3. Unplug the Ultrasound system.
4. Maintain control of the Ultrasound system power plug.
5. Wait for at least 30 seconds for capacitors to discharge as
there are no test points to verify isolation.
6. Remove/disconnect the battery, if present.
Ultrasound System components may be energized.