Create new dataflow with Worklist and DICOM Storage
Vivid T8/ Vivid T8 Pro – Basic Service Manual
5490863-100 English Rev. 03
1. Press F2 on the control panel.
2. Select Connectivity->Dataflow to display the dataflow
screen. Select Add to start the creation of a new dataflow.
Figure 3-47. Connectivity-> Dataflow
3. It brings up the Enter new name dialog. Enter a name, e.g.
Worklist-DICOM Storage, and then select OK.
Figure 3-48. Enter New Name
Connectivity setup
3-61