GE Vivid T8 Basic Service Manual page 117

Hide thumbs Also See for Vivid T8:
Table of Contents

Advertisement

Setup connection to a DICOM server (continued)
NOTE:
Vivid T8/ Vivid T8 Pro – Basic Service Manual
5490863-100 English Rev. 03
1. Press Config (F2) on the control panel and log on as
administrator.
2. Select Connectivity and then select Dataflow tab.
3. Select the arrow to the right for the Name field to list all
dataflows in a pull-down menu and select the dataflow
DICOM Server.
Figure 3-25. Select Dataflow
If Direct Store is enabled, the image will go to the
DICOM server immediately after it is acquired.
Make sure the DICOM server is capable of keeping the
connection during an examination.
If Hidden is enabled, the dataflow will be invisible in the
Search/Create Patient screen. Uncheck Hidden to
display the dataflow in the Search/Create Patient
screen.
Connectivity setup
3-41