General Procedures and Functional Checks

Back End Processor checks

Operator Panel Test

Peripheral checks

Printer checks

Turn OFF Power to Vivid T8/ Vivid T8 Pro

Mechanical Functions Checks

Operator Panel Movement
4-34
If all the previous tests have been passed successfully, the
Back End Processor is most likely OK.
If the system seems to be operating erratically, please refer
to
'Diagnostics/Troubleshooting' on page
The Operator Panel is tested when the is powered up as
part of the start-up scripts, run at every start-up.
For more info, please refer to
on page
7-1.
The internal printer is controlled from the P1 and P2 keys on the
Vivid T8/ Vivid T8 Pro's Operator Panel.
The factory default is:
P1 for the UP-D711MD printer
P2 for the whole screen secondary capture to clipboard
See 'Power off' on page 4-8 for more information.
Please refer to:
'LCD Monitor position adjustment' on page 4-9
Vivid T8/ Vivid T8 Pro – Basic Service Manual
7-1.
'Diagnostics/Troubleshooting'
5490863-100 English Rev. 03