Power On / Boot Up

3-4-3-1

Scanner Power On

1.) Connect the Power Cable to the back of the system.
2.) Connect the Main Power Cable to a hospital grade power receptacle with the proper rated voltage.
Never use an adapter that would defeat the safety ground.
3.) Switch ON the Circuit Breaker and the Power Switch of peripherals at the rear of the system.
Circuit Breaker (F1)
Connector for
Mains Power Cable
Figure 3-8 Circuit Breaker (F1) and Power Switch of Peripherals (F2)
When AC power is applied to the scanner, the ON/OFF switch on the Control panel is illuminated,
NOTICE
indicating that the System (including the Back-end Processor) is in standby mode.
3-4-3-2

Back-end Processor Boot Up

Press the ON/OFF Stand-By switch left belowthe Control Panel once.
ON/OFF Stand By Switch
When the ON/OFF Stand-By switch left below the Control Panel is pressed once, the System (including
the Back-end Processor) starts and the software code is distributed to initiate the scanner.
The company logo appears on the Touch screen during boot up.
Depending on the BIOS-Version no status messages are displayed during this process.
Boot up time is about 3 minutes.
NOTE:
The mains outlet of the system for peripheral auxiliary equipment are commonly switched with
the Stand-By switch. The switch of printers has to be in ON position before starting the system.
So the auxiliary equipment need not to be switched ON/OFF separately if the F2 power switch
on the back of the system and the switches of the peripherals are always ON.
U
RETZTECHNIK
LTRASOUND
1
Figure 3-9 ON/OFF Stand-By Switch
Chapter 3 Installation
V
® 730 S
OLUSON
ERVICE
Power
Switch of
Peripherals
(F2)
M
ANUAL
3-9