Maintenance - GE SE 250 Mighty Small II Operating Instructions Manual

Table of Contents

Advertisement

6 Maintenance

6
Maintenance
About this chapter
This chapter provides information to enable users and service personnel to clean and
maintain the product.
Precautions
General procedures
Immediately after each use, rinse the unit with water and then rinse thoroughly with
distilled water. Do not use organic solvents, abrasives, strong cleaning solutions, or
strong acids or bases to clean the chambers. Handle the upper buffer chamber core
with care to prevent damage to the banana plugs. Allow to air dry.
Clean glass and alumina plates and spacers with a dilute solution of a laboratory cleanser,
then rinse thoroughly with tap and distilled water. Glass plates can also be treated with
(but not stored in) acid cleaning solutions. Do not autoclave or heat any part above 45°C.
42
WARNING
Decontaminate before maintenance. To avoid personnel being
exposed to potentially hazardous substances, make sure that the
SE 250/260 Mighty Small II is properly decontaminated and sani-
tized before maintenance or service.
WARNING
Use only approved parts. Only spare parts and accessories that
are approved or supplied by GE may be used for maintaining or
servicing the product.
WARNING
Disconnect power. Always disconnect power from the instrument
before performing any maintenance task.
SE 250/260 Mighty Small II Operating Instructions 29281629 AA