Step
1
2
3
4
SE 250/260 Mighty Small II Operating Instructions 29281629 AA
Action
Fill the lower buffer chamber with running buffer.
The SE 250/260 Mighty Small II holds about 250 mL. Make sure that the
lower electrode (running along the bottom of the the upper buffer chamber
core) is completely submerged.
Note:
If using precast gels, check that the lower gel/buffer contact surface is exposed
(the colored plastic tape must be removed).
Place the safety lid on the unit.
Plug the color-coded leads into the jacks of an approved power supply. The
red lead plugs into the red output jack, and the black lead plugs into the
black output jack.
Optional cooling: Begin circulating cold water or a chilled 50/50 water/ethy-
lene glycol solution.
5 Operation
5.2 Assembly
39