Note: BDH depends on threaded or flanged. Rods are either 30 or 50 cm. See
source not found.
Figure 2-10 Minimum blocking distance, steam application for a flanged HTHP process
connector
Process
connector type
Threaded
Flanged
Minimum BDH and distance from reference plane to top inlet depends on transmitter
configuration.
Note: these distances also apply to coax probes as we turn off the dynamic calculation when
the surface is closer than this value.
Page 26

Table 2-1: Blocking Distance High

Saturated Steam
Ref Length
30 cm
50 cm
30 cm
50 cm
SLG 700 SmartLine Level Transmitter User's Manual
Minimum BDH
Min dist to inlet or surface
with DC corrected
measurement
47.0 cm
58.0 cm
67.0 cm
78.0 cm
44.5 cm
55.5 cm
64.5 cm
75.5 cm
Error! Reference
Revision 8