Assembly and Setup
Item
3
4

Using the test plug

1.
2.
3.
4.
5.
6.
32
Description
Incoming Power Cord
Main Power Switch
Turn the power "OFF" at the treadmill.
Disconnect the RS232 or USB interface cable from the
treadmill and plug in the test connector.
Press and hold the button on the test connector and turn the
treadmill power "ON". Continue holding until the treadmill
begins to elevate.
When the treadmill begins to raise, repeatedly press the
button to cycle through the different phases of the test.
Once the treadmill begins to rise, each press of the
o
button should raise it in 5% increments.
Once the treadmill reaches a 20% elevation, the next
o
press of the button should start the treadmill running
belt.
Once the treadmill running belt starts, each press of the
o
button should increase the speed by 2.5 mph (4 km/h).
Once the treadmill reaches 10 mph (16 km/h), each
o
press of the button should decrease the speed by 2.5
mph (4 km/h) and lower the elevation in 5% increments.
Once the treadmill reaches the minimum speed and
o
elevation, the next press of the button should stop the
treadmill running belt.
NOTE:
Successful completion of the preceding testing
procedure ensures that the treadmill is fully functional
and responsive to command signals.
NOTE:
Unsuccessful completion of the preceding testing
procedure indicates a problem with the setup. Call GE
Medical Systems Information Technologies to trouble
shoot failure of test plug procedure.
Remove the test connector.
Reconnect the RS232 or USB interface cable from the host
computer. You are ready to begin the set-up procedure
prescribed by your medical test equipment supplier.