Troubleshooting

6
Your in-house biomedical technician or technically qualified
personnel can perform basic troubleshooting. Advanced
troubleshooting may be found in the T2100-ST1/T2100-ST2 Service
Manual shipped with your system, or contact GE Medical Systems
Information Technologies, Inc., and open a customer-billable service
call.

Power Switch Will Not Illuminate

This procedure requires the use of a meter that is capable of
reading at least 250 VAC.
1.
2.
3.
4.
NOTE:
2097937-001 Rev E
12 June 2017
Turn the treadmill's main power switch to OFF and unplug the
treadmill's power cable from the wall receptacle.
Measure the voltage at the wall outlet for the correct voltage
level.
If the voltage is not correct, reset the circuit breaker. If the
voltage is still not correct, refer the problem to your local
electrician.
If the voltage is correct, proceed to step 3.
Plug the treadmill's power cable into the wall receptacle and
turn the treadmill's main power switch to ON.
Observe whether the power switch LED lights up, indicating
the treadmill is receiving power.
Contact GE Medical Systems Information Technologies, Inc., for
additional troubleshooting if the power switch will not
illuminate.