AreaRAE Plus & AreaRAE Pro User's Guide
These are the screens that are shown on an AreaRAE Plus/Pro after startup if "Can Bypass" is selected:
If "Can't Bypass" is selected, the display looks like this, and only allows the options of performing the
test or shutting down:
16. Once you have made your selections in ProRAE Studio II, you must upload the changes to the
instrument. Click the icon labeled "Upload all settings to the instrument."
17. A confirmation screen is shown. Click "Yes" to perform the upload, or "No" to abort.
Uploading takes a few seconds, and a progress bar is shown. You can abort the upload by clicking
"Cancel."
18. Exit ProRAE Studio II.
19. Press [Y/+] on the AreaRAE Plus/Pro to exit Communication Mode.

14.2 Deactivating Policy Enforcement

If the AreaRAE Plus/Pro screen displays the message that it must be bump tested or calibrated, and if the
option to bypass bump testing or calibration is not available, you should shut off the instrument and
follow the procedure outlined here if you want to change the Policy Enforcement settings:
1. Use a USB cable to connect the AreaRAE to a computer running ProRAE Studio II.
2. Enter Diagnostic Mode on the AreaRAE Plus/Pro (with the instrument turned off, press and hold
[Y/+] and [MODE] until it starts up.
3. After startup, enter the password when prompted (default is "0000") and press [MODE].
4. Press [N/-] repeatedly until you see the "Enter Communications Mode?" screen.
5. Press [Y/+] to enter Communications Mode.
6. Start ProRAE Studio II.
7. Select "Administrator."
65