Gfi Configuration - 9L-D Probe - GE LOGIQ E9 Service Manual

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

D
5180483-100, R
IRECTION
7-9-21
How to Recognize missing Transmit and Receive Channels for a 9L-D Probe in a
LOGIQ E9 GFI Configuration (cont'd)
7-9-21-3
9L-D Probe - Missing Transmit and Receive Channels
The dark areas or artifacts, within the near field (in the viewing area), indicate missing transmit
and receive channels associated with:
• GTX2
• DRX2
• GRX128
Approximate distance from A (0 cm) to the following points:
B = 0.7 cm
C = 1.5 cm
D = 2.9 cm
E = 3.8 cm
7 - 300
PROPRIETARY TO GE
. 7
EV
Table 7-63
A
B
GTX2
DRX2
GRX128
Section 7-9 - Troubleshooting Tips

GFI Configuration - 9L-D Probe

C
D
LOGIQ E9 P
ROPRIETARY
E
GTX2
DRX2
GRX128