GE LOGIQ E9 Service Manual page 407

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

D
5180483-100, R
IRECTION
Table 7-51
Amber light ON but no green light
(or briefly ON) when pressing ON/
OFF button
7 - 204
PROPRIETARY TO GE
. 7
EV
System not booting up
Symptom
When there is a failure that forces the Main Power Supply to shutdown, you will
only see the amber light trying to turn green once, until the rear panel Main
Power Switch is reset.
If there is no green light at all,
even after resetting the main
Section 7-9 - Troubleshooting Tips
breaker
If the green light appears
momentarily
LOGIQ E9 P
ROPRIETARY
Task / Conclusion
Measure test connector, pin 12 and ground
5Vstb.
Test connector pin (24V) to see if system turns
ON.
- Use the Power ON/Off button on the
Motherboard.
If system does not power up, replace:
1. BEP PS
2. EPS
3. BEP
If system powers up, troubleshoot connection
to OP Panel.
- Main Power Supply Fan failure may cause
this problem. Use a mirror to look at the fans;
check if they try to rotate at boot up (two
people may be needed to test this).
Mirror Viewing fans under Main Power
Supply
If the fans do not move, tip the Main PS and
gently move each fan with a screw driver to
move them from their current position.
Sometimes removing the PS is enough
movement to move the fans.
If the Main PS powers up after the fans are
moved, you will see a failure message in the
Log Files. See:
7-9-4-6 "Main Power Supply
Logs" on page
7-214.
If the fans do not mo ve after this step, the Main
PS needs replacement.
NOTE: This is a rare occurrence. Monitor site
for future occurrences.